The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00382903851621), received between Aug 7, 2018 and Nov 24, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Connection Problem1
  • Disconnection1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 24, 2025Connection Problem; Connection Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23630802 ↗
  • MalfunctionNov 3, 2025Disconnection; Disconnection

    ADDITIONAL INFORMATION: ADDITIONALLY, WE HAVE INTERNALLY CONDUCTED A COMPARISON BETWEEN THE OLD AND THE NEW RECEIVED EXTENSION AND IT SEEMS THAT THERE IS A DIFFERENCE IN THE TIP OF THE CONNECTION TUBING, WHICH HAS LED TO THE LOSS OF CONNECTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23454762 ↗
  • MalfunctionNov 16, 2021Leak/Splash; Leak/Splash

    IT WAS REPORTED BD Q-SYTE TRI-EXTENSION SET HAD LEAKAGE ISSUES.THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM GERMAN: ¿THE ANESTHESIA DEPARTMENT HAS DETERMINED THAT WHEN USING THE [TRI-EXTENSION SET] CONNECTED THE 3-WAY-STOPCOK... PROPOFOL LEAKS DROP BY DROP, IT DOES NOT MATTER HOW TIGHT THE CONNECTION IS. IT HAPPENS WITH OTHER FLUIDS AS WELL, BUT BECAUSE PROPOFOL IS KIND OF MILKY IT IS ALSO MORE NOTICEABLE. MOREOVER, THE SAME ISSUE WAS EXPERIENCED TWICE WHEN TWO 3-WAY-STOPCOKS WERE CONNECTED TO EACH OTHER.¿

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12824087 ↗
  • MalfunctionAug 7, 2018Break

    IT WAS REPORTED THAT BD Q-SYTE¿ TRI-EXTENSION SET 15 CM (6 IN) 2.25 ML HAD BROKEN VALVES. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7757240 ↗