BD QUINCKE Spinal Needles 408360Recalls & reported problems
Becton, Dickinson And Company · NEEDLE SPINAL S/SU 18GA 6IN QUINCKE · Class 2
All models of BD QUINCKE Spinal Needles (9)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00382904083601), received between Jul 16, 2019 and Feb 12, 2026 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Contamination1
- Device Contamination with Chemical or Other Material1
- Device Markings/Labelling Problem1
- Leak/Splash1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Known Impact Or Consequence To Patient2
- Injury1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 12, 2026Contamination
NO ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24322466 ↗MalfunctionAug 23, 2022Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT THE BD¿ LONG LENGTH SPINAL NEEDLE EXPERIENCED INCONSISTENT LABELLING INFORMATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DIFFERING EXPIRY DATE STATED ON THE CERTIFICATE OF COMPLIANCE AND THE PACKAGING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15277750 ↗MalfunctionMar 9, 2020Break; Break
IT WAS REPORTED THAT DURING QUALITY CONTROL TESTING A NEEDLE SPINAL S/SU 18GA 6 IN QUINCKE WAS FOUND TO NOT ALLOW REINSERTION OF THE STYLET INTO THE NEEDLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "HERE IS A PHOTOGRAPH OF THE BD PACKAGES AND THE LOT NUMBER OF THE NEEDLE THAT FAILED QC. THEY RECEIVED 2 LOTS AND THE FIRST LOT OF 50 NEEDLES WAS USED AND PASSED OUR INSPECTION TEST. FOR THE SECOND LOT, THERE WERE 20 UNITS RECEIVED IN TWO BOXES. WE OPENED ONE AND DID THE INSPECTION. THE STYLET WOULDN'T REINSERT INTO THE NEEDLE. SO THERE ARE 19 UNOPENED NEEDLES AND THE ONE THAT WAS USED TO INSPECT. AS YOU CAN SEE, THE RANDOMLY SELECTED NEEDLE FROM THAT LOT WOULD NOT ALLOW REINSERTION INTO THE NEEDLE. WE REJECT THE ENTIRE LOT BASED ON THIS RESULT"
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9804542 ↗InjuryFeb 4, 2020Leak/Splash; Leak/Splash
IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF NEEDLE SPINAL S/SU 18GA 6IN QUINCKE EXPERIENCED PRODUCT DAMAGE/DEFORMATION -DEVICE STILL OPERABLE, AND WERE INVOLVED IN A SERIOUS INJURY IN THE FORM OF MEDICAL INTERVENTION. DURING TREATMENT, A PORTION OF THE DEVICE BECAME EMBEDDED WITHIN THE PATIENT'S KIDNEY AND HAD TO BE REMOVED BY A HEALTH CARE PROFESSIONAL. NO ADDITIONAL INFORMATION REGARDING THE OUTCOME OF THIS MEDICAL INTERVENTION HAS BEEN SPECIFIED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE DEVICE IN QUESTION IS: AIG PL 150MM 18G 408360, REF: 408360 AND NUMBER OF SERIES: 8123805 AND 9088609. THE PROBLEM ENCOUNTERED IS: "DURING AN NLPC (NEPHROLITHOTOMY PER CUTANEOUS) USE OF THE SPINAL NEEDLE TO LOCATE THE KIDNEY. SMALLER NEEDLE DIAMETER AT THE END OF THE NEEDLE IMPEDING THE PASSAGE OF THE GUIDE THAT "SPINS" AND CAN'T REMOVE THE GUIDE." CLINICAL CONSEQUENCES: "RUPTURE OF PART OF THE GUIDE WIRE IN THE PATIENT'S KIDNEY. THIS PART HAD TO BE RECOVERED."
Patient problems: Injury
MAUDE report 9667274 ↗MalfunctionJul 16, 2019Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF NEEDLE SPINAL S/SU 18GA 6 IN QUINCKE EXPERIENCED FOREIGN MATTER CONTAMINATION WHICH WAS NOTED PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: 1 SPINAL NEEDLE MISSING IN 1 BOX (PRESENCE OF 9 NEEDLES INSTEAD OF 10). 1 SPINAL NEEDLE WITH AN IMPURITY OR INK STAIN INSIDE THE BLISTER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8795142 ↗
Other NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER) devices
- ARROW IPN922632Teleflex Incorporated · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)2 recalls0 reports
- ARROW IPN922603Teleflex Incorporated · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)1 recall0 reports
- ARROW IPN922604Teleflex Incorporated · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)1 recall0 reports
- ARROW IPN923597Teleflex Incorporated · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)1 recall0 reports
- ARROW IPN922631Teleflex Incorporated · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)1 recall0 reports
- ARROW IPN922605Teleflex Incorporated · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)1 recall0 reports
- 22G 1 1/2IN Regional Block Needle 408348Becton, Dickinson And Company · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)1 recall0 reports
- BD QUINCKE Spinal Needles 405180Becton, Dickinson And Company · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)23 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page