BD MAX™ StaphSR 443419Recalls & reported problems

Becton, Dickinson And Company · BD MAX™ StaphSR · Class 2

Quick summary

This BD MAX StaphSR is a laboratory system that quickly detects *Staph aureus* and methicillin resistance in patient samples. Most of its 44 reports describe test results that came back wrong — samples read as negative when the bug was actually present, or positive when it was not. There are 43 malfunctions, 1 death and no reported injuries. Two recalls, both started in July 2019 and now closed, involved foil bags of extraction tubes that may not have been sealed properly, allowing reagent contamination.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
44
Deaths reported
1
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 44 adverse event reports that list this exact device (DI 00382904434199), received between Dec 28, 2017 and Nov 5, 2025 — classified as 1 death, 43 malfunctions.

The most frequently reported specific device problem is “False Negative Result” (30 reports).

2 FDA recall records name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 3 reportsOct 2025: 15 reportsNov 2025: 10 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202615
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 17, 2019Terminated

    The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly.

    Z-2333-2019 · Root cause: Process control · FDA record ↗

  2. Jul 17, 2019Terminated

    The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly.

    Z-2332-2019 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • False Negative Result30
  • False Positive Result13
  • Incorrect, Inadequate or Imprecise Result or Readings10
  • Incorrect Or Inadequate Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions43
  • Death1

Recent adverse event reports

The 25 most recent of 44 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 5, 2025False Negative Result; False Negative Result

    REPORT 10 OF 10: IT WAS REPORTED THAT DURING USE OF BD MAX¿ STAPHSR, A MREJ FALSE NEGATIVE RESULT WAS OBTAINED. MREJ WAS DETECTED IN GENEXPERT. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23476210 ↗
  • MalfunctionNov 5, 2025False Negative Result; False Negative Result

    REPORT 8 OF 10: IT WAS REPORTED THAT DURING USE OF BD MAX¿ STAPHSR, A MREJ FALSE NEGATIVE RESULT WAS OBTAINED. MREJ WAS DETECTED IN GENEXPERT. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23476207 ↗
  • MalfunctionNov 5, 2025False Negative Result; False Negative Result

    REPORT 4 OF 10: IT WAS REPORTED THAT DURING USE OF BD MAX¿ STAPHSR, A MREJ FALSE NEGATIVE RESULT WAS OBTAINED. MREJ WAS DETECTED IN GENEXPERT. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23476206 ↗
  • MalfunctionNov 5, 2025False Negative Result; False Negative Result

    REPORT 7 OF 10: IT WAS REPORTED THAT DURING USE OF BD MAX¿ STAPHSR, A MREJ FALSE NEGATIVE RESULT WAS OBTAINED. MREJ WAS DETECTED IN GENEXPERT. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23476205 ↗
  • MalfunctionNov 5, 2025False Negative Result; False Negative Result

    REPORT 9 OF 10: IT WAS REPORTED THAT DURING USE OF BD MAX¿ STAPHSR, A MREJ FALSE NEGATIVE RESULT WAS OBTAINED. MREJ WAS DETECTED IN GENEXPERT. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23476203 ↗
  • MalfunctionNov 5, 2025False Negative Result; False Negative Result

    REPORT 6 OF 10: IT WAS REPORTED THAT DURING USE OF BD MAX¿ STAPHSR, A MREJ FALSE NEGATIVE RESULT WAS OBTAINED. MREJ WAS DETECTED IN GENEXPERT. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23476201 ↗
  • MalfunctionNov 5, 2025False Negative Result; False Negative Result

    REPORT 2 OF 10: IT WAS REPORTED THAT DURING USE OF BD MAX¿ STAPHSR, A MREJ FALSE NEGATIVE RESULT WAS OBTAINED. MREJ WAS DETECTED IN GENEXPERT. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23476194 ↗
  • MalfunctionNov 5, 2025False Negative Result; False Negative Result

    REPORT 3 OF 10: IT WAS REPORTED THAT DURING USE OF BD MAX¿ STAPHSR, A MREJ FALSE NEGATIVE RESULT WAS OBTAINED. MREJ WAS DETECTED IN GENEXPERT. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23476193 ↗
  • MalfunctionNov 5, 2025False Negative Result; False Negative Result

    REPORT 5 OF 10: IT WAS REPORTED THAT DURING USE OF BD MAX¿ STAPHSR, A MREJ FALSE NEGATIVE RESULT WAS OBTAINED. MREJ WAS DETECTED IN GENEXPERT. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23476192 ↗
  • MalfunctionNov 5, 2025False Negative Result; False Negative Result

    REPORT 1 OF 10: IT WAS REPORTED THAT DURING USE OF BD MAX¿ STAPHSR, A MREJ FALSE NEGATIVE RESULT WAS OBTAINED. MREJ WAS DETECTED IN GENEXPERT. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23476185 ↗
  • MalfunctionOct 28, 2025False Negative Result; False Negative Result

    REPORT 1 OF 3: IT WAS REPORTED THAT DURING USE OF BD MAX¿ STAPHSR, A MREJ FALSE NEGATIVE RESULT WAS OBTAINED. MREJ WAS DETECTED IN GENEXPERT. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23406165 ↗
  • MalfunctionOct 28, 2025False Negative Result; False Negative Result

    REPORT 3 OF 3: IT WAS REPORTED THAT DURING USE OF BD MAX¿ STAPHSR, A MREJ FALSE NEGATIVE RESULT WAS OBTAINED. MREJ WAS DETECTED IN GENEXPERT. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23405824 ↗
  • MalfunctionOct 28, 2025False Negative Result; False Negative Result

    REPORT 2 OF 3: IT WAS REPORTED THAT DURING USE OF BD MAX¿ STAPHSR, A MREJ FALSE NEGATIVE RESULT WAS OBTAINED. MREJ WAS DETECTED IN GENEXPERT. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23405822 ↗
  • MalfunctionOct 28, 2025False Negative Result; False Negative Result

    REPORT 2 OF 3: IT WAS REPORTED THAT DURING USE OF BD MAX¿ STAPHSR, A MREJ FALSE NEGATIVE RESULT WAS OBTAINED. MREJ WAS DETECTED IN GENEXPERT. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23405646 ↗
  • MalfunctionOct 28, 2025False Negative Result; False Negative Result

    REPORT 3 OF 3: IT WAS REPORTED THAT DURING USE OF BD MAX¿ STAPHSR, A MREJ FALSE NEGATIVE RESULT WAS OBTAINED. MREJ WAS DETECTED IN GENEXPERT. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23405643 ↗
  • MalfunctionOct 21, 2025False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings; False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings

    REPORT 7 OF 10: IT WAS REPORTED THAT DURING USE OF BD MAX¿ STAPHSR, A DISCREPANT PATIENT RESULT WAS OBTAINED. SAMPLE WAS TESTED ON GENEXPERT AND MAX RESULT WAS NOT CONFIRMED. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23351776 ↗
  • MalfunctionOct 21, 2025False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings; False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings

    REPORT 4 OF 10: IT WAS REPORTED THAT DURING USE OF BD MAX¿ STAPHSR, A DISCREPANT PATIENT RESULT WAS OBTAINED. SAMPLE WAS TESTED ON GENEXPERT AND MAX RESULT WAS NOT CONFIRMED. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23351770 ↗
  • MalfunctionOct 21, 2025False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings; False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings

    REPORT 10 OF 10: IT WAS REPORTED THAT DURING USE OF BD MAX¿ STAPHSR, A DISCREPANT PATIENT RESULT WAS OBTAINED. SAMPLE WAS TESTED ON GENEXPERT AND MAX RESULT WAS NOT CONFIRMED. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23350794 ↗
  • MalfunctionOct 21, 2025False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings; False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings

    REPORT 8 OF 10: IT WAS REPORTED THAT DURING USE OF BD MAX¿ STAPHSR, A DISCREPANT PATIENT RESULT WAS OBTAINED. SAMPLE WAS TESTED ON GENEXPERT AND MAX RESULT WAS NOT CONFIRMED. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23350779 ↗
  • MalfunctionOct 21, 2025False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings; False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings

    REPORT 9 OF 10: IT WAS REPORTED THAT DURING USE OF BD MAX¿ STAPHSR, A DISCREPANT PATIENT RESULT WAS OBTAINED. SAMPLE WAS TESTED ON GENEXPERT AND MAX RESULT WAS NOT CONFIRMED. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23350775 ↗
  • MalfunctionOct 21, 2025False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings; False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings

    REPORT 2 OF 10: IT WAS REPORTED THAT DURING USE OF BD MAX¿ STAPHSR, A DISCREPANT PATIENT RESULT WAS OBTAINED. SAMPLE WAS TESTED ON GENEXPERT AND MAX RESULT WAS NOT CONFIRMED. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23350496 ↗
  • MalfunctionOct 21, 2025False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings; False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings

    REPORT 5 OF 10: IT WAS REPORTED THAT DURING USE OF BD MAX¿ STAPHSR, A DISCREPANT PATIENT RESULT WAS OBTAINED. SAMPLE WAS TESTED ON GENEXPERT AND MAX RESULT WAS NOT CONFIRMED. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23350488 ↗
  • MalfunctionOct 21, 2025False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings; False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings

    REPORT 3 OF 10: IT WAS REPORTED THAT DURING USE OF BD MAX¿ STAPHSR, A DISCREPANT PATIENT RESULT WAS OBTAINED. SAMPLE WAS TESTED ON GENEXPERT AND MAX RESULT WAS NOT CONFIRMED. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23350483 ↗
  • MalfunctionOct 21, 2025False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings; False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings

    REPORT 1 OF 10: IT WAS REPORTED THAT DURING USE OF BD MAX¿ STAPHSR, A DISCREPANT PATIENT RESULT WAS OBTAINED. SAMPLE WAS TESTED ON GENEXPERT AND MAX RESULT WAS NOT CONFIRMED. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23350482 ↗
  • MalfunctionOct 21, 2025False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings; False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings

    REPORT 6 OF 10: IT WAS REPORTED THAT DURING USE OF BD MAX¿ STAPHSR, A DISCREPANT PATIENT RESULT WAS OBTAINED. SAMPLE WAS TESTED ON GENEXPERT AND MAX RESULT WAS NOT CONFIRMED. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23350481 ↗

Other real time Nucleic acid amplification system devices

All real time Nucleic acid amplification system devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.