BD MAX™ Enteric Viral Panel 443985Recalls & reported problems
Becton, Dickinson And Company · Kit BD MAX Enteric Viral Panel · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This BD MAX Enteric Viral Panel is a laboratory test kit for intestinal viruses. Most of the 135 reports describe wrong positive results, mainly for norovirus and rotavirus. There are no deaths or injuries. The 1 recall, started in July 2019 and now closed, concerns foil bags containing extraction tubes that may not have been sealed properly.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 135
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 135 adverse event reports that list this exact device (DI 00382904439859), received between Feb 2, 2021 and May 5, 2026 — classified as 135 malfunctions.
The most frequently reported specific device problem is “False Positive Result” (118 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Jul 17, 2019Class IITerminated
The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly.
Z-2339-2019 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- False Positive Result118
- Incorrect, Inadequate or Imprecise Result or Readings11
- False Negative Result7
- Non Reproducible Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions135
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 25 most recent of 135 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 5, 2026False Positive Result
IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC VIRAL PANEL, AN UNSPECIFIED NUMBER OF LOW FALSE POSITIVE NOROVIRUS AND ROTOVIRUS PATIENT RESULTS WERE OBTAINED. LOW POSITIVE SAMPLES WERE RETESTED AND WERE NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25085968 ↗MalfunctionMay 5, 2026False Positive Result
IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC VIRAL PANEL, AN UNSPECIFIED NUMBER OF LOW FALSE POSITIVE NOROVIRUS AND ROTOVIRUS PATIENT RESULTS WERE OBTAINED. LOW POSITIVE SAMPLES WERE RETESTED AND WERE NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25085967 ↗MalfunctionMay 5, 2026False Positive Result
IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC VIRAL PANEL, AN UNSPECIFIED NUMBER OF LOW FALSE POSITIVE NOROVIRUS AND ROTOVIRUS PATIENT RESULTS WERE OBTAINED. LOW POSITIVE SAMPLES WERE RETESTED AND WERE NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25085965 ↗MalfunctionMay 5, 2026False Positive Result
IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC VIRAL PANEL, AN UNSPECIFIED NUMBER OF LOW FALSE POSITIVE NOROVIRUS AND ROTOVIRUS PATIENT RESULTS WERE OBTAINED. LOW POSITIVE SAMPLES WERE RETESTED AND WERE NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25085963 ↗MalfunctionMay 5, 2026False Positive Result
IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC VIRAL PANEL, AN UNSPECIFIED NUMBER OF LOW FALSE POSITIVE NOROVIRUS AND ROTOVIRUS PATIENT RESULTS WERE OBTAINED. LOW POSITIVE SAMPLES WERE RETESTED AND WERE NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25085948 ↗MalfunctionApr 28, 2026False Positive Result
IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC VIRAL PANEL, A FALSE POSITIVE ASTROVIRUS PATIENT RESULT WAS OBTAINED. INITIAL TEST WAS BOTH SAPOVIRUS AND ASTROVIRUS POSITIVE. REPEAT OF THE SAMPLE GAVE SAPOVIRUS POSITIVE ONLY RESULT. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25022183 ↗MalfunctionApr 3, 2026False Positive Result
IT WAS REPORTED THAT DURING USE OF THE BD MAX¿ ENTERIC VIRAL PANEL , A DISCREPANT ASTROVIRUS PATIENT RESULT WAS OBTAINED. INITIAL TEST WAS ASTROVIRUS POSITIVE. TEST WAS REPEATED AND WAS ASTROVIRUS NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24778065 ↗MalfunctionApr 2, 2026False Positive Result
IT WAS REPORTED THAT DURING USE OF THE BD MAX¿ ENTERIC VIRAL PANEL , A DISCREPANT ASTROVIRUS PATIENT RESULT WAS OBTAINED. INITIAL TEST WAS ASTROVIRUS POSITIVE. TEST WAS REPEATED AND WAS ASTROVIRUS NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24765504 ↗MalfunctionApr 2, 2026False Positive Result
IT WAS REPORTED THAT DURING USE OF THE BD MAX¿ ENTERIC VIRAL PANEL , A DISCREPANT ASTROVIRUS PATIENT RESULT WAS OBTAINED. INITIAL TEST WAS ASTROVIRUS POSITIVE. TEST WAS REPEATED AND WAS ASTROVIRUS NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24765503 ↗MalfunctionFeb 17, 2026False Positive Result
IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC VIRAL PANEL, A FALSE POSITIVE NOROVIRUS PATIENT RESULT WAS OBTAINED. TEST WAS REPEATED ON MAX AND WAS NOROVIRUS NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24367787 ↗MalfunctionFeb 10, 2026False Positive Result
IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC VIRAL PANEL, AN UNSPECIFIED NUMBER OF FALSE POSITIVE ADENOVIRUS PATIENT RESULTS WERE OBTAINED. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24307416 ↗MalfunctionFeb 6, 2026False Positive Result
IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC VIRAL PANEL, A FALSE POSITIVE NOROVIRUS PATIENT RESULT WAS OBTAINED. TEST WAS REPEATED ON MAX AND WAS NOROVIRUS NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24282869 ↗MalfunctionFeb 4, 2026False Positive Result
REPORT 6 OF 6: IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC VIRAL PANEL, A FALSE POSITIVE PATIENT RESULT WAS OBTAINED. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24261926 ↗MalfunctionFeb 4, 2026False Positive Result
REPORT 2 OF 6: IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC VIRAL PANEL, A FALSE POSITIVE PATIENT RESULT WAS OBTAINED. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24261923 ↗MalfunctionFeb 4, 2026False Positive Result
REPORT 4 OF 6: IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC VIRAL PANEL, A FALSE POSITIVE PATIENT RESULT WAS OBTAINED. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24261922 ↗MalfunctionFeb 4, 2026False Positive Result
REPORT 3 OF 6: IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC VIRAL PANEL, A FALSE POSITIVE PATIENT RESULT WAS OBTAINED. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24261920 ↗MalfunctionFeb 4, 2026False Positive Result
REPORT 5 OF 6: IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC VIRAL PANEL, A FALSE POSITIVE PATIENT RESULT WAS OBTAINED. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24261905 ↗MalfunctionJan 21, 2026False Positive Result
IT WAS REPORTED THAT DURING USE OF THE BD MAX¿ ENTERIC VIRAL PANEL, A FALSE POSITIVE NOROVIRUS PATIENT RESULT WAS OBTAINED. TESTING OF THIS SPECIMEN WAS REPEATED AND WAS NEGATIVE FOR NOROVIRUS. SPECIMEN WAS ALSO SENT OUT FOR CONFIRMATORY TESTING AND WAS NEGATIVE FOR NOROVIRUS. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24133379 ↗MalfunctionJan 15, 2026False Positive Result
IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC VIRAL PANEL, A FALSE POSITIVE PATIENT RESULT WAS OBTAINED. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24081527 ↗MalfunctionJan 9, 2026False Positive Result
REPORT 2 OF 2. IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC VIRAL PANEL, A FALSE POSITIVE NOROVIRUS PATIENT RESULT WITH AN UNUSUAL PCR CURVE WAS OBTAINED. SAMPLE WAS TESTED ON BIOFIRE AND CEPHEID AND WAS NEGATIVE FOR NOROVIRUS. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24029035 ↗MalfunctionJan 9, 2026False Positive Result
REPORT 2 OF 2: IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC VIRAL PANEL, A FALSE POSITIVE NOROVIRUS PATIENT RESULT WITH AN UNUSUAL PCR CURVE WAS OBTAINED. SAMPLE WAS TESTED ON BIOFIRE AND CEPHEID AND WAS NEGATIVE FOR NOROVIRUS. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24029029 ↗MalfunctionDec 18, 2025False Negative Result; False Negative Result
IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC VIRAL PANEL, A FALSE NEGATIVE NOROVIRUS PATIENT RESULT WAS OBTAINED. SAMPLE WAS REPEATED ON GENEXPERT AND BIOFIRE AND WAS NOROVIRUS POSITIVE. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23842885 ↗MalfunctionDec 18, 2025False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings; False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC VIRAL PANEL, A DISCREPANT SAPOVIRUS PATIENT RESULT WAS OBTAINED. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23842488 ↗MalfunctionDec 18, 2025False Negative Result; False Negative Result
IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC VIRAL PANEL, A FALSE NEGATIVE NOROVIRUS PATIENT RESULT WAS OBTAINED. SAMPLE WAS REPEATED ON GENEXPERT AND BIOFIRE AND WAS NOROVIRUS POSITIVE. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23842473 ↗MalfunctionDec 17, 2025False Positive Result; False Positive Result
REPORT 2 OF 2: IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC VIRAL PANEL, A FALSE POSITIVE NOROVIRUS PATIENT RESULT WITH A WEEK PCR CURVE WAS OBTAINED. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23828881 ↗
FDA clearance and classification
Cleared through 510(k) K220607 on Sep 21, 2022 (decision: Substantially Equivalent). Submitted by Becton, Dickinson and Company as “BD MAX Enteric Viral Panel”. The FDA review took 203 days. FDA record ↗
The FDA regulates product code OOI (real time nucleic acid amplification system) under 21 CFR 862.2570 as a Class II device. 21 CFR 862.2570 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the BD MAX™ Enteric Viral Panel 443985
Has the BD MAX™ Enteric Viral Panel 443985 been recalled?
Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-2339-2019 (Class II), began on Jul 17, 2019: The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the BD MAX™ Enteric Viral Panel 443985?
The FDA's MAUDE database lists 135 adverse event reports for this device, received between Feb 2, 2021 and May 5, 2026. In those reports, the most common specific device problems are “False Positive Result” (118), “Incorrect, Inadequate or Imprecise Result or Readings” (11) and “False Negative Result” (7). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the BD MAX™ Enteric Viral Panel 443985?
None of the reports are classified as deaths, 0 as injuries and 135 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the BD MAX™ Enteric Viral Panel 443985 FDA approved?
It was cleared through 510(k) K220607 on Sep 21, 2022. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the BD MAX™ Enteric Viral Panel 443985, and is it still sold?
The FDA's device database lists Becton, Dickinson And Company as the company responsible for it (brand BD MAX™ Enteric Viral Panel, model 443985). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other real time Nucleic acid amplification system devices
- BD MAX™ System 441916Becton, Dickinson And Company · real time Nucleic acid amplification system2 recalls1,055 reports
- Alinity m 08N53-002Abbott Molecular Inc. · real time Nucleic acid amplification system4 recalls943 reports
- BD PCR Cartridge 437519Becton, Dickinson And Company · real time Nucleic acid amplification system1 recall140 reports
- Verigene® Processor SP 10-0000-07Nanosphere, Inc. · Real Time Nucleic Acid Amplification System3 recalls0 reports
- BD MAX™ StaphSR 443419Becton, Dickinson And Company · real time Nucleic acid amplification system2 recalls1 death44 reports
- Revogene 610210Meridian Bioscience, Inc. · Real Time Nucleic Acid Amplification System2 recalls10 reports
- FilmArray® BCID Panel RFIT-ASY-0126Biofire Diagnostics, LLC · Real Time Nucleic Acid Amplification System2 recalls2 deaths4 reports
- FilmArray® BCID Panel RFIT-ASY-0127Biofire Diagnostics, LLC · Real Time Nucleic Acid Amplification System2 recalls1 death1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page