BD PrepStain Slide Processor 490407Recalls & reported problems
Becton, Dickinson And Company · PrepStain® Slide Processor (Refurb) · Class 3
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00382904904074), received between Jan 26, 2022 and Aug 23, 2023 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Contamination” (2 reports).
11 other recalls cover devices cleared under the same FDA submission (P970018); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Contamination2
- Structural Problem1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 23, 2023Structural Problem; Structural Problem
IT WAS REPORTED THAT WHILE USING INSTRUMENT PEPSTATIN TECAN US II REFURB, THE LID WAS FALLING AND WON'T STAY OPEN. NO PATIENT IMPACT REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "TRIPATH US II - CENTRIFUGE TWO STRUT FAILURE THE AGENT ENGINEER, IS IN A HURRY TO GIVE FEEDBACK. THE CENTRIFUGE'S SUPPORT ROD FAILED, BUT DID NOT LOSE ITS SUPPORT COMPLETELY. THERE WAS NO FREE FALL, BUT A SLOW FALL WITHOUT INJURY."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17606959 ↗MalfunctionFeb 27, 2023Contamination; Contamination
IT WAS REPORTED THAT WHILE USING THE BD INSTRUMENT PREPSTAIN TECAN US II REFURB THAT THERE WAS CROSS-CONTAMINATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER STAINS OFFLINE SO CELLS OUTSIDE CELL CIRCLE ARE STAINED. CUSTOMER HAS DETERMINED THAT DRIPS ON TOP OF SLIDES ARE CROSS CONTAMINATION FROM OTHER SAMPLES. CUSTOMER DOES NOT SCREEN THESE SLIDES AND REPROCESSES SAMPLE ON A DIFFERENT SLIDE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16450239 ↗MalfunctionFeb 28, 2022Unintended Movement; Unintended Movement
IT WAS REPORTED THAT THE BD PREPSTAIN¿ SLIDE PROCESSOR'S CENTRIFUGE LID WOULD NOT CLOSE PROPERLY. THERE WAS NO ADVERSE USER OR PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CENTRIFUGE LID NOT CLOSING PROPERLY" "NOBODY WAS HURT DUE TO THIS ISSUE. YES, THE CENTRIFUGE NOW OPERATING NORMALLY."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13620831 ↗MalfunctionJan 26, 2022Contamination; Contamination
IT WAS REPORTED THAT THE BD PREPSTAIN¿ SLIDE PROCESSOR'S SPECIMEN DRIPPED OUT AND CREATED CROSS-CONTAMINATION. AS A RESULT, THE SPECIMENS HAD TO BE RERUN. THERE WAS NO INDICATION THAT THE ERRONEOUS RESULTS WERE REPORTED TO CLINICIANS, AND THERE WAS NO REPORT OF ADVERSE PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CUSTOMER REQUESTING DISPATCH DUE TO DRIPPING SPECIMEN. THERE WAS CONTAMINATION." "WERE ANY PTS TREATED BASED ON ERRONEOUS RESULTS? NO, WHEN THE CROSS CONTAMINATION WAS SEEN, THEY RERAN THE SPECIMENS. DID THE WRONG TREATMENT HARM THE PATIENT? NO."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13359371 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P970018. One submission often covers many models, so check the product and lot details in each record.
- Z-0877-2017 · Feb 12, 2016 · TerminatedBD SurePathTM Preservative Collection Vials, catalog numbers 490505, 490527, 491095, 491096, and 491097. SurePath Preservative Fluid is designed for use with the PrepStain System
- Z-1815-2012 · Mar 22, 2012 · TerminatedPrepStain GN 480 Disposable Kit Catalog/Model 490542 This is a complete kit, per the firm's explanation, that contains the PrepStain Syringing Pipettes 240, the product being recal
- Z-1811-2012 · Mar 22, 2012 · TerminatedPrepStain Syringing Pipettes 240 (Catalog #490517)
- Z-1814-2012 · Mar 22, 2012 · TerminatedManual Method 1: 480 Test Kit (PreCoat), Catalog/Model # 490529
- Z-1813-2012 · Mar 22, 2012 · TerminatedPrepStain Installation Kit (PreCoat), Catalog/Model # 490528
- Z-1812-2012 · Mar 22, 2012 · TerminatedPrepStain GYN 480 Disposable Test Kit with PreCoat Slides (Catalog #490523
- Z-2088-2012 · Dec 8, 2011 · TerminatedPrepStain PM Kit, T-AG (Catalog 490588); PrepStain PM Kit Europe (Catalog 490671); PrepStain Prep Only PM Kit, T-AG (Catalog 490954); PrepStain PM Kit XLP, T-US-II (Catalog 490424)
- Z-2085-2012 · Dec 8, 2011 · TerminatedPrepStain Kit: 1.2 DiTi Cone. The PrepStain System is a liquid-based thin layer cell preparation process. The PrepStain System produces SurePath Slides that are intended for use i
- Z-2087-2012 · Dec 8, 2011 · TerminatedPrepStain Slide Processor. The PrepStain System is a liquid-based thin layer cell preparation process. The PrepStain System produces SurePath Slides that are intended for use in t
- Z-2086-2012 · Dec 8, 2011 · TerminatedPrepStain Upgrade Kit: 1.3.0.3/1.2 DiTi Cone. The PrepStain System is a liquid-based thin layer cell preparation process. The PrepStain System produces SurePath Slides that are in
- Z-1090-2009 · Sep 26, 2008 · TerminatedBD PrepStain System, Model Number: 799-14000-00; Catalog Number: 05CR000021, 05CR00021R, 799-13000-00, 799-13000-00R, 799-14000-00 and 799-14000-00R; Catalog Number: 05CR000021, 05
Other Processor, Cervical Cytology Slide, Automated devices
- BD SurePath™ Collection Vial 491452Becton, Dickinson And Company · Processor, cervical cytology slide, automated82 reports
- BD Totalys SlidePrep 491346Becton, Dickinson And Company · Processor, Cervical Cytology Slide, Automated53 reports
- BD PrepStain Slide Processor 490100Becton, Dickinson And Company · Processor, Cervical Cytology Slide, Automated24 reports
- BD Totalys MultiProcessor 443327Becton, Dickinson And Company · Processor, Cervical Cytology Slide, Automated13 reports
- BD Totalys™ MultiProcessor Consumables Kit 491453Becton, Dickinson And Company · Processor, cervical cytology slide, automated4 reports
- BD PrepStain Slide Processor 490410Tripath Imaging, Inc. · Processor, Cervical Cytology Slide, Automated2 reports
- BD PrepStain™ Consumables Kit 491454Becton, Dickinson And Company · Processor, cervical cytology slide, automated1 report
- BD SurePath™ Preservative Fluid 491337Becton, Dickinson And Company · Processor, Cervical Cytology Slide, Automated1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page