Skip to content

BD FACSDuet 663128Recalls & reported problems

Becton, Dickinson And Company · BD FACSDuet Sample Prep US · Class 1

All models of BD FACSDuet (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00382906631282), received between Nov 15, 2021 and Nov 15, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Contamination” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Aug 12, 2024Class IIOpen, Classified

    Sample preparation system with software may cause: 1) Multidispense feature enabled with 2-tube assay: Specimen volume dispensed into tube1 at least 3% higher than volume in tube2, or 2) Absolute counts calculated using Trucount Tubes: increased volume in tube1 may result in 20% higher absolute counts in tube 1 than tube 2; could result in additional blood collection, delayed results/patient therapy.

    Z-0009-2025 · Root cause: Software change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Contamination1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 15, 2021Contamination; Contamination

    IT WAS REPORTED THAT WHILE RUNNING BD FACSDUET¿ SAMPLE PREP CARRYOVER OCCURRED. THERE WAS NO PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THERE WAS CARRY OVER AND CONTAMINATION OF THE ANTIBODIES FROM THE MACHINE. ARE YOU USING THIS PRODUCT FOR CLINICAL DIAGNOSTIC TEST? N. WERE ERRONEOUS RESULTS REPORTED AND USED TO TREAT A PATIENT? N. WAS THERE ANY INJURY OR POTENTIAL INJURY? N.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12815731 ↗

Other Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis devices

All Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page