Sample Preparation System PS-10Recalls & reported problems
Sysmex Corporation · The Sample Preparation System PS-10 is a fully automated, user-configurable, open system intended for use to prepare human specimens for subsequent analysis on flow cytometers. · Class 1
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04987562501908), received between Jan 9, 2025 and Feb 6, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Computer Software Problem” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Jan 6, 2025Open, Classified
Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.
Z-1113-2025 · Root cause: Software design · FDA record ↗
Most reported problems
Device problems
- Computer Software Problem3
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 6, 2025Computer Software Problem; Computer Software Problem
A SOFTWARE DEFECT WAS IDENTIFIED IN SAMPLE PREPARATION SYSTEM PS-10 SERIAL NUMBER (SN) (B)(6) SOFTWARE VERSION 1.5, WHERE OMISSION OF THE PROBE WASH STEP AND POTENTIAL ANTIBODY CARRY-OVER MAY LEAD TO POSSIBLE ERRONEOUS REPRESENTATION OF ANTIGEN MARKERS IN A SAMPLE AND THUS GENERATE ERRONEOUS RESULTS IN SUBSEQUENT FLOW CYTOMETRY ANALYSIS. THERE HAS BEEN NO REPORT OF INCORRECT TREATMENT, SERIOUS INJURY OR DEATH DUE TO THE PRODUCT PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21322948 ↗MalfunctionJan 10, 2025Computer Software Problem; Computer Software Problem
A SOFTWARE DEFECT WAS IDENTIFIED IN SAMPLE PREPARATION SYSTEM PS-10 SOFTWARE VERSION 1.5 AND SOFTWARE VERSION 1.6, WHERE OMISSION OF THE PROBE WASH STEP AND POTENTIAL ANTIBODY CARRY-OVER MAY LEAD TO POSSIBLE ERRONEOUS REPRESENTATION OF ANTIGEN MARKERS IN A SAMPLE AND THUS GENERATE ERRONEOUS RESULTS IN SUBSEQUENT FLOW CYTOMETRY ANALYSIS. THERE HAS BEEN NO REPORT OF INCORRECT TREATMENT, SERIOUS INJURY OR DEATH DUE TO THE PRODUCT PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21130098 ↗MalfunctionJan 9, 2025Computer Software Problem; Computer Software Problem
A SOFTWARE DEFECT WAS IDENTIFIED IN SAMPLE PREPARATION SYSTEM PS-10 SOFTWARE VERSION 1.5 AND SOFTWARE VERSION 1.6, WHERE OMISSION OF THE PROBE WASH STEP AND POTENTIAL ANTIBODY CARRY-OVER MAY LEAD TO POSSIBLE ERRONEOUS REPRESENTATION OF ANTIGEN MARKERS IN A SAMPLE AND THUS GENERATE ERRONEOUS RESULTS IN SUBSEQUENT FLOW CYTOMETRY ANALYSIS. THERE HAS BEEN NO REPORT OF INCORRECT TREATMENT, SERIOUS INJURY OR DEATH DUE TO THE PRODUCT PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21111764 ↗
Other Automated Pipetting, Diluting And Specimen Processing Workstations For Flow Cytometric Analysis devices
- FACS 647205Becton, Dickinson And Company · Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis123 reports
- BD FACSDuet 666340Becton, Dickinson And Company · Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis1 recall1 report
- BD FACSDuet 663128Becton, Dickinson And Company · Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis1 recall1 report
- Cellmek C44603Beckman Coulter, Inc. · Automated Pipetting, Diluting And Specimen Processing Workstations For Flow Cytometric Analysis1 recall0 reports
- PrepPlus 2 378600Beckman Coulter, Inc. · Automated Pipetting, Diluting And Specimen Processing Workstations For Flow Cytometric Analysis1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.