Sample Preparation System PS-10Recalls & reported problems

Sysmex Corporation · The Sample Preparation System PS-10 is a fully automated, user-configurable, open system intended for use to prepare human specimens for subsequent analysis on flow cytometers. · Class 1

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04987562501908), received between Jan 9, 2025 and Feb 6, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Computer Software Problem” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 6, 2025Open, Classified

    Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.

    Z-1113-2025 · Root cause: Software design · FDA record ↗

Most reported problems

Device problems

  • Computer Software Problem3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 6, 2025Computer Software Problem; Computer Software Problem

    A SOFTWARE DEFECT WAS IDENTIFIED IN SAMPLE PREPARATION SYSTEM PS-10 SERIAL NUMBER (SN) (B)(6) SOFTWARE VERSION 1.5, WHERE OMISSION OF THE PROBE WASH STEP AND POTENTIAL ANTIBODY CARRY-OVER MAY LEAD TO POSSIBLE ERRONEOUS REPRESENTATION OF ANTIGEN MARKERS IN A SAMPLE AND THUS GENERATE ERRONEOUS RESULTS IN SUBSEQUENT FLOW CYTOMETRY ANALYSIS. THERE HAS BEEN NO REPORT OF INCORRECT TREATMENT, SERIOUS INJURY OR DEATH DUE TO THE PRODUCT PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21322948 ↗
  • MalfunctionJan 10, 2025Computer Software Problem; Computer Software Problem

    A SOFTWARE DEFECT WAS IDENTIFIED IN SAMPLE PREPARATION SYSTEM PS-10 SOFTWARE VERSION 1.5 AND SOFTWARE VERSION 1.6, WHERE OMISSION OF THE PROBE WASH STEP AND POTENTIAL ANTIBODY CARRY-OVER MAY LEAD TO POSSIBLE ERRONEOUS REPRESENTATION OF ANTIGEN MARKERS IN A SAMPLE AND THUS GENERATE ERRONEOUS RESULTS IN SUBSEQUENT FLOW CYTOMETRY ANALYSIS. THERE HAS BEEN NO REPORT OF INCORRECT TREATMENT, SERIOUS INJURY OR DEATH DUE TO THE PRODUCT PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21130098 ↗
  • MalfunctionJan 9, 2025Computer Software Problem; Computer Software Problem

    A SOFTWARE DEFECT WAS IDENTIFIED IN SAMPLE PREPARATION SYSTEM PS-10 SOFTWARE VERSION 1.5 AND SOFTWARE VERSION 1.6, WHERE OMISSION OF THE PROBE WASH STEP AND POTENTIAL ANTIBODY CARRY-OVER MAY LEAD TO POSSIBLE ERRONEOUS REPRESENTATION OF ANTIGEN MARKERS IN A SAMPLE AND THUS GENERATE ERRONEOUS RESULTS IN SUBSEQUENT FLOW CYTOMETRY ANALYSIS. THERE HAS BEEN NO REPORT OF INCORRECT TREATMENT, SERIOUS INJURY OR DEATH DUE TO THE PRODUCT PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21111764 ↗

Other Automated Pipetting, Diluting And Specimen Processing Workstations For Flow Cytometric Analysis devices