InnoSpire Elegance 00 (REF 1102460)Recalls & reported problems
Respironics Respiratory Drug Delivery (Uk) Ltd · Innospire Elegance SSR 220v 60Hz Korea, 6-pack · Class 2
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00383730001753), received between Mar 23, 2021 and Mar 23, 2021 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Material Split, Cut or Torn4
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 23, 2021Material Split, Cut or Torn; Material Split, Cut or Torn
THE MANUFACTURER RECEIVED AN INNOSPIRE ELEGANCE DEVICE FOR EVALUATION WITH THE COMPLAINT THAT THE DEVICE WOULD NOT SWITCH ON. THERE WAS NO PATIENT HARM OR INJURY REPORTED. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, AN ISSUE RELATED TO THE DEVICE'S AC POWER CORD WAS OBSERVED. THERE WAS EVIDENCE OF EXPOSURE TO THE CONDUCTING WIRES. THE MANUFACTURER CONCLUDES THAT THIS EXPOSURE WAS CAUSED BY THE AC POWER CORD BEING EXPOSED TO FORCES BEYOND THE INTENDED DESIGN. THE INNOSPIRE ELEGANCE DEVICE IS INTENDED TO PROVIDE A SOURCE OF COMPRESSED AIR FOR MEDICAL PURPOSES. IT IS TO BE USED WITH A PNEUMATIC (JET) NEBULIZER TO PRODUCE AEROSOL PARTICLES OF MEDICATION FOR RESPIRATORY THERAPY FOR CHILDREN AND ADULTS. LABELING WARNS THE "IT IS RECOMMENDED TO HAVE A BACKUP DEVICE FOR RESPIRATORY DELIVERY IN CASE A SITUATION ARISES WHEN YOUR NEBULIZER CAN NOT BE USED". THIS DEVICE MEETS ALL RELEVANT UL AND IEC STANDARDS. THIS DEVICE IS NOT LABELED FOR LIFE SUPPORT. BASED ON THE INFORMATION AVAILABLE, THE MANUFACTURER CONCLUDES NO FURTHER ACTION IS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11554655 ↗MalfunctionMar 23, 2021Material Split, Cut or Torn; Material Split, Cut or Torn
THE MANUFACTURER RECEIVED AN INNOSPIRE ELEGANCE DEVICE FOR EVALUATION WITH THE COMPLAINT THAT THE DEVICE WOULD NOT SWITCH ON. THERE WAS NO PATIENT HARM OR INJURY REPORTED. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, AN ISSUE RELATED TO THE DEVICE'S AC POWER CORD WAS OBSERVED. THERE WAS EVIDENCE OF EXPOSURE TO THE CONDUCTING WIRES. THE MANUFACTURER CONCLUDES THAT THIS EXPOSURE WAS CAUSED BY THE AC POWER CORD BEING EXPOSED TO FORCES BEYOND THE INTENDED DESIGN. THE INNOSPIRE ELEGANCE DEVICE IS INTENDED TO PROVIDE A SOURCE OF COMPRESSED AIR FOR MEDICAL PURPOSES. IT IS TO BE USED WITH A PNEUMATIC (JET) NEBULIZER TO PRODUCE AEROSOL PARTICLES OF MEDICATION FOR RESPIRATORY THERAPY FOR CHILDREN AND ADULTS. LABELING WARNS THE "IT IS RECOMMENDED TO HAVE A BACKUP DEVICE FOR RESPIRATORY DELIVERY IN CASE A SITUATION ARISES WHEN YOUR NEBULIZER CAN NOT BE USED". THIS DEVICE MEETS ALL RELEVANT UL AND IEC STANDARDS. THIS DEVICE IS NOT LABELED FOR LIFE SUPPORT. BASED ON THE INFORMATION AVAILABLE, THE MANUFACTURER CONCLUDES NO FURTHER ACTION IS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11554633 ↗MalfunctionMar 23, 2021Material Split, Cut or Torn; Material Split, Cut or Torn
THE MANUFACTURER RECEIVED AN INNOSPIRE ELEGANCE DEVICE FOR EVALUATION WITH THE COMPLAINT THAT THE DEVICE WOULD NOT SWITCH ON. THERE WAS NO PATIENT HARM OR INJURY REPORTED. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, AN ISSUE RELATED TO THE DEVICE'S AC POWER CORD WAS OBSERVED. THERE WAS EVIDENCE OF EXPOSURE TO THE CONDUCTING WIRES. THE MANUFACTURER CONCLUDES THAT THIS EXPOSURE WAS CAUSED BY THE AC POWER CORD BEING EXPOSED TO FORCES BEYOND THE INTENDED DESIGN. THE INNOSPIRE ELEGANCE DEVICE IS INTENDED TO PROVIDE A SOURCE OF COMPRESSED AIR FOR MEDICAL PURPOSES. IT IS TO BE USED WITH A PNEUMATIC (JET) NEBULIZER TO PRODUCE AEROSOL PARTICLES OF MEDICATION FOR RESPIRATORY THERAPY FOR CHILDREN AND ADULTS. LABELING WARNS THE "IT IS RECOMMENDED TO HAVE A BACKUP DEVICE FOR RESPIRATORY DELIVERY IN CASE A SITUATION ARISES WHEN YOUR NEBULIZER CAN NOT BE USED". THIS DEVICE MEETS ALL RELEVANT UL AND IEC STANDARDS. THIS DEVICE IS NOT LABELED FOR LIFE SUPPORT. BASED ON THE INFORMATION AVAILABLE, THE MANUFACTURER CONCLUDES NO FURTHER ACTION IS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11554603 ↗MalfunctionMar 23, 2021Material Split, Cut or Torn; Material Split, Cut or Torn
THE MANUFACTURER RECEIVED AN INNOSPIRE ELEGANCE DEVICE FOR EVALUATION WITH THE COMPLAINT THAT THE DEVICE WOULD NOT SWITCH ON. THERE WAS NO PATIENT HARM OR INJURY REPORTED. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, AN ISSUE RELATED TO THE DEVICE'S AC POWER CORD WAS OBSERVED. THERE WAS EVIDENCE OF EXPOSURE TO THE CONDUCTING WIRES. THE MANUFACTURER CONCLUDES THAT THIS EXPOSURE WAS CAUSED BY THE AC POWER CORD BEING EXPOSED TO FORCES BEYOND THE INTENDED DESIGN. THE INNOSPIRE ELEGANCE DEVICE IS INTENDED TO PROVIDE A SOURCE OF COMPRESSED AIR FOR MEDICAL PURPOSES. IT IS TO BE USED WITH A PNEUMATIC (JET) NEBULIZER TO PRODUCE AEROSOL PARTICLES OF MEDICATION FOR RESPIRATORY THERAPY FOR CHILDREN AND ADULTS. LABELING WARNS THE "IT IS RECOMMENDED TO HAVE A BACKUP DEVICE FOR RESPIRATORY DELIVERY IN CASE A SITUATION ARISES WHEN YOUR NEBULIZER CAN NOT BE USED". THIS DEVICE MEETS ALL RELEVANT UL AND IEC STANDARDS. THIS DEVICE IS NOT LABELED FOR LIFE SUPPORT. BASED ON THE INFORMATION AVAILABLE, THE MANUFACTURER CONCLUDES NO FURTHER ACTION IS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11554595 ↗
Other Compressor, air, portable devices
- Medline HCS70004Medline Industries, Inc. · Compressor, air, portable386 reports
- Medline HCS70004RDMedline Industries, Inc. · Compressor, air, portable108 reports
- Compressor Mini 115V 60Hz 6481779Maquet Critical Care AB · Compressor, Air, Portable108 reports
- Life2000 MS-01-0093Welch Allyn, Inc. · Compressor, air, portable1 recall1 death5 reports
- Life2000 MS-01-0121Welch Allyn, Inc. · Compressor, air, portable1 recall0 reports
- Life2000 MS-01-0125Welch Allyn, Inc. · Compressor, air, portable1 recall0 reports
- InnoSpire Elegance 00Respironics Respiratory Drug Delivery (Uk) Ltd · Compressor, air, portable19 reports
- InnoSpire Essence 00Respironics Respiratory Drug Delivery (Uk) Ltd · Compressor, air, portable18 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.