Skip to content

InnoSpire Essence 00 (REF 1095060)Recalls & reported problems

Respironics Respiratory Drug Delivery (Uk) Ltd · Innospire Essence Assembly, Plug A with120V 60HZ MOTOR · Class 2

All models of InnoSpire Essence (4)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00383730001807), received between Jul 19, 2018 and Mar 11, 2026 — classified as 1 death, 7 malfunctions.

The most frequently reported specific device problem is “Unexpected Shutdown” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unexpected Shutdown2
  • Adverse Event Without Identified Device or Use Problem1
  • Contamination /Decontamination Problem1
  • Degraded1
  • Electrical /Electronic Property Problem1
  • Failure to Deliver1
  • Failure to Power Up1
  • Fire1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient2
  • Respiratory Failure1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • DeathMar 11, 2026Adverse Event Without Identified Device or Use Problem

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING A PATIENT WITH AN INNOSPIRE ESSENCE NEBULIZER HAS PASSED AWAY. IT WAS ALLEGED THE NEBULIZER WAS DEFECTIVE AND STATED THAT THE DEVICE WAS SUBSEQUENTLY RECALLED. IT IS ALLEGED THAT USE OF THE DEVICE CAUSED OR CONTRIBUTED TO WORSENING OF THE PATIENT¿S RESPIRATORY CONDITION. THE PATIENT SUBSEQUENTLY PASSED ON (B)(6) 2025. THE DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP REPORT WILL BE FILED

    Patient problems: Respiratory Failure

    MAUDE report 24571611 ↗
  • MalfunctionJul 23, 2024Failure to Power Up; Smoking; Failure to Power Up; Smoking

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO AN ASSAY FOR ESSENCE PLUG DEVICE WITH ALLEGATION OF DEVICE IS SMOKING AND NOT TURNING ON. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19804822 ↗
  • MalfunctionMay 2, 2024Failure to Deliver; Failure to Deliver

    THE MANUFACTURER BECAME AWARE OF A USER OF AN ESSENCE NEBULIZER ALLEGING THAT MEDICATION WAS NOT COMING OUT OF THE DEVICE. THE PATIENT STATES THAT THEY RECENTLY REPLACED THE FILTER AND CLEANED THE PARTS FOLLOWING THE INSTRUCTIONS. THERE WAS NO REPORT OF SERIOUS HARM OR INJURY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19230935 ↗
  • MalfunctionNov 16, 2023Unexpected Shutdown; Unexpected Shutdown

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING THAT THE NEBULIZER STOPPED RUNNING DURING THERAPY. IT WAS ALSO MENTIONED THAT THE DEVICE WAS ABOUT 3 YEARS OLD. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. AT THIS TIME, NO FURTHER INVESTIGATION CAN BE PERFORMED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP REPORT WILL BE FILED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18146186 ↗
  • MalfunctionOct 28, 2022Electrical /Electronic Property Problem; Contamination /Decontamination Problem; Electrical /Electronic Property Problem; Contamination /Decontamination Problem

    THE MANUFACTURER BECAME AWARE THAT AN ESSENCE NEBULIZER WAS NOT WORKING. THERE WAS NO REPORT OF SMOKING, FLAMES OR EXPOSED WIRES. THE NEBULIZER WAS RETURNED FOR EVALUATION AND THE TECHNICIAN FOUND EVIDENCE OF A MILKY CONTAMINATE THROUGHOUT THE INSIDE AND OUTSIDE OF THE DEVICE. IT APPEARS THE LIQUID CONTAMINATION MADE ITS WAY INSIDE THE SWITCH FROM THE TOP OF THE ENCLOSURE CAUSING VERDIGRIS AND A SHORT. THIS SHORT MELTED THE INTERNAL POWER CABLES AT THE ON/OFF SWITCH, AND MELTED INTO THE NPC BUFFER TUBE CAUSING A HOLE IN THE BUFFER CYLINDER AND DAMAGED THE SURROUNDING AREA. THERE WAS NO ALLEGATIONS OF HARM OR INJURY. THE MANUFACTURER CONFIRMED THE COMPLAINT OF DEVICE NOT WORKING DUE TO THE THERMAL DAMAGE INSIDE THE MACHINE AT THE ON/OFF SWITCH. THE ROOT CAUSE COULD BE THE SWITCH ITSELF HAD AN INTERNAL SHORT, THE INSULATED CONNECTION TABS AS THE END OF THE WHITE AND BLACK INTERNAL POWER WIRES MAY HAVE HAD A FUILURE OR THE LIQUID CONTAMINATION HAD CAUSED THE THERMAL EVENT. THE NEBULIZER HAD EVIDENCE OF GRIME, DIRT, DEBRIS AND A BLACK POWDERY SUBSTANCE FALLING FROM THE VENTS. THERE WAS NO AIR FILTER IN PLACE. LABELING WARNS: ANY LIQUID SPILLED ON OR INSIDE THE UNIT SHOULD BE ALLOWED TO

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15692566 ↗
  • MalfunctionFeb 9, 2021Material Split, Cut or Torn; Unexpected Shutdown; Material Split, Cut or Torn; Unexpected Shutdown

    THE MANUFACTURER RECEIVED A REPORT OF A POSSIBLE THERMAL EVENT TO AN INNOSPIRE ESSENCE NEBULIZER. THE USER REPORTED SHE HIT THE POWER BUTTON, THE DEVICE STOPPED WORKING AND A SECTION OF THE POWER CORD STARTED TO SMOKE. THERE WAS EVIDENCE OF EXPOSED WIRES ON THE POWER CORD. THERE WAS NO REPORT OF FLAMES, OR DAMAGE TO THE NEBULIZER. THERE WAS NO PATIENT HARM OR INJURY ASSOCIATED WITH THIS REPORTED EVENT. UPON RECEIPT OF THE DEVICE, THE MANUFACTURER FOUND THE DEVICE WORKING AS DESIGNED. THE POWER CORD REVEALED A CUT OR PINCHED MIDDLE SECTION WITH THE EXPOSED BARE INNER WIRES WITH EVIDENCE OF MELTING. THE ROOT CAUSE OF THE EVENT WAS UNABLE TO BE DETERMINED, HOWEVER, IT IS LIKELY THE AC POWER CORD HAD BEEN EXPOSED TO FORCES BEYOND INTENDED DESIGN. THERE WERE NO VOIDS TO THE ENCLOSURE, NO THERMAL DAMAGE TO THE ACTUAL NEBULIZER. THE INNOSPIRE ESSENCE IS INTENDED TO PROVIDE A SOURCE OF COMPRESSED AIR FOR MEDICAL PURPOSES. IT IS TO BE USED WITH A PNEUMATIC (JET) NEBULIZER TO PRODUCE AEROSOL PARTICLES OF MEDICATION FOR RESPIRATORY THERAPY FOR BOTH CHILDREN AND ADULTS. THIS DEVICE IS NOT LABELED FOR LIFE SUPPORT. LABELING INCLUDED WITH THE DEVICE WARNS THE USER, "IT IS RECOMMENDED TO HAVE A B

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11304859 ↗
  • MalfunctionApr 14, 2020Fire; Fire

    THE MANUFACTURER WAS MADE AWARE BY AN END USER THAT A THERMAL EVENT OCCURRED WITH AN INNOSPIRE ESSENCE DEVICE. THERE WAS NO PATIENT HARM OR INJURY ASSOCIATED WITH THE REPORTED EVENT. THE USER REPORTED THAT THE MACHINE WAS TURNED ON AND IN USE WHEN THEY NOTICED THE DEVICE BEGAN TO SMOKE AND FLAMES WERE COMING FROM THE DEVICE. THE USER UNPLUGGED THE DEVICE AND TOOK IT OUTSIDE WHERE THE FLAMES THEN SELF-EXTINGUISHED. THE MANUFACTURER REQUESTED THE RETURN OF THE DEVICE FOR INVESTIGATION; HOWEVER, THE USER DISPOSED OF THE DEVICE AND WILL NOT BE RETURNING IT TO THE MANUFACTURER. THE ROOT CAUSE OF THE REPORTED EVENT IS UNABLE TO BE CONFIRMED. THIS PRODUCT IS INTENDED TO PROVIDE A SOURCE OF COMPRESSED AIR FOR MEDICAL PURPOSES. IT IS TO BE USED WITH A PNEUMATIC (JET) NEBULIZER TO PRODUCE AEROSOL PARTICLES OF MEDICATION FOR RESPIRATORY THERAPY FOR BOTH CHILDREN AND ADULTS. THIS IS NOT A LIFE SUPPORT DEVICE. THE USER IS WARNED IN THE LABELING TO HAVE A BACKUP DEVICE FOR RESPIRATORY DELIVERY IF THIS DEVICE IS NOT OPERABLE. BASED ON THE INFORMATION AVAILABLE, THE MANUFACTURER CONCLUDES NO FURTHER ACTION IS NECESSARY, AND WILL CONTINUE TO MONITOR COMPLAINTS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9957768 ↗
  • MalfunctionJul 19, 2018Degraded; Degraded

    IT WAS REPORTED THAT THIS DEVICE HAD EXPOSED WIRE. ON RECEIPT OF THE DEVICE FOR EVALUATION IT WAS OBSERVED THAT THE MAINS CORD WAS DAMAGED LEADING TO EXPOSED WIRES. THE IFU FOR THIS DEVICE CONTAINS THE WARNING 'THE POWER SUPPLY CORD CANNOT BE REPLACED BY THE USER. IN CASE THE POWER SUPPLY CORD BECOMES DAMAGED, CONTACT PHILIPS RESPIRONICS CUSTOMER SERVICE FOR REPLACEMENT (THERE IS NO SERVICE OPTION)'. NO PATIENT HARM OR DEATH HAS BEEN REPORTED AND THE DEVICE IS NOT LIFE SUSTAINING/SUPPORTING. (B)(4)

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7702222 ↗

FDA clearance and classification

Cleared through 510(k) K042655 on Oct 13, 2004 (decision: Substantially Equivalent). Submitted by Respironics New Jersey, Inc. as “Inspiration Elite Nebulizer Compressor, With Nebulizer, Model Hs456”. The FDA review took 15 days. FDA record ↗

The FDA regulates product code BTI (compressor, air, portable) under 21 CFR 868.6250 as a Class II device. 21 CFR 868.6250 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the InnoSpire Essence 00

Has the InnoSpire Essence 00 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00383730001807), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the InnoSpire Essence 00?

The FDA's MAUDE database lists 8 adverse event reports for this device, received between Jul 19, 2018 and Mar 11, 2026. In those reports, the most common specific device problems are “Unexpected Shutdown” (2), “Contamination /Decontamination Problem” (1) and “Degraded” (1); the most common patient problem is “Respiratory Failure” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the InnoSpire Essence 00?

1 report is classified as a death, 0 as injuries and 7 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the InnoSpire Essence 00 FDA approved?

It was cleared through 510(k) K042655 on Oct 13, 2004. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the InnoSpire Essence 00, and is it still sold?

The FDA's device database lists Respironics Respiratory Drug Delivery (Uk) Ltd as the company responsible for it (brand InnoSpire Essence, model 00). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Compressor, air, portable devices

All Compressor, air, portable devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page