Skip to content

InnoSpire Elegance 00 (REF 1104624)Recalls & reported problems

Respironics Respiratory Drug Delivery (Uk) Ltd · Innospire Elegance, Custom Majors Medical, 6 pack. · Class 2

All models of InnoSpire Elegance (10)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00383730003160), received between Mar 9, 2017 and Mar 1, 2019 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Degraded” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Degraded1
  • Electrical /Electronic Property Problem1

Patient problems

  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 1, 2019Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE MANUFACTURER RECEIVED AN INNOSPIRE ELEGANCE DEVICE FOR EVALUATION OF "LOW AIR FLOW". THERE WAS NO PATIENT HARM OR INJURY REPORTED. UPON RECEIPT OF THE DEVICE, THE MANUFACTURER NOTED PHYSICAL DAMAGE TO THE AC POWER SOURCE, WHICH RESULTED IN EXPOSED CONDUCTING WIRES, AND A RISK OF SHOCK TO THE USER. THERE WAS NO EVIDENCE OF THERMAL DAMAGE. ALTHOUGH THE ROOT CAUSE OF THE OBSERVED DAMAGE WAS NOT CONFIRMED, THIS TYPE OF DAMAGE IS TYPICALLY A RESULT OF THE AC POWER CORD BEING SUBJECTED TO FORCES BEYOND INTENDED DESIGN. THE INNOSPIRE ELEGANCE DEVICE IS INTENDED TO PROVIDE A SOURCE OF COMPRESSED AIR FOR MEDICAL PURPOSES. IT IS TO BE USED WITH A PNEUMATIC (JET) NEBULIZER TO PRODUCE AEROSOL PARTICLES OF MEDICATION FOR RESPIRATORY THERAPY FOR CHILDREN AND ADULTS. LABELING WARNS THE USER "IT IS RECOMMENDED TO HAVE A BACKUP DEVICE FOR RESPIRATORY DELIVERY IN CASE A SITUATION ARISES WHEN YOUR NEBULIZER CAN NOT BE USED". THE USER IS ALSO CAUTIONED THAT THE POWER SUPPLY CORD CANNOT BE REPLACED BY THE USER. IN CASE THE POWER SUPPLY CORD BECOMES DAMAGED, CONTACT PHILIPS RESPIRONICS CUSTOMER SERVICE FOR REPLACEMENT (THERE IS NO SERVICE OPTION). THIS DEVICE MEETS ALL RELEVANT UL AND IEC STANDARDS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8383088 ↗
  • MalfunctionMar 9, 2017Degraded; Degraded

    IT WAS REPORTED THAT THIS DEVICE HAD LOW AIR FLOW. ON RETURN OF THE DEVICE FOR EVALUATION, IT WAS NOTED THAT THE MAINS CORD HAD BEEN SEVERED AND THE PLUG WAS MISSING. THIS HAS LED TO EXPOSED MAINS WIRES, WHICH POSES NO HAZARD ITSELF AS THE DEVICE CANNOT NOW BE CONNECTED TO A MAINS SUPPLY. HOWEVER, IT HAS NOT BEEN ESTABLISHED HOW THE PLUG CAME TO BE MISSING, AND THERE IS A POSSIBILITY THAT IT BROKE OFF IN THE MAINS SOCKET - IF THIS WAS THE CASE THEN A RISK OF ELECTROCUTION WOULD BE CREATED. OUR EVALUATION CONCLUDED THAT IN ALL LIKELIHOOD THE CORD WAS SEVERED BY THE END USER APPLYING EXCESSIVE FORCE. INTERNAL REFERENCE (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6390703 ↗

FDA clearance and classification

Cleared through 510(k) K042655 on Oct 13, 2004 (decision: Substantially Equivalent). Submitted by Respironics New Jersey, Inc. as “Inspiration Elite Nebulizer Compressor, With Nebulizer, Model Hs456”. The FDA review took 15 days. FDA record ↗

The FDA regulates product code BTI (compressor, air, portable) under 21 CFR 868.6250 as a Class II device. 21 CFR 868.6250 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the InnoSpire Elegance 00

Has the InnoSpire Elegance 00 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00383730003160), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the InnoSpire Elegance 00?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Mar 9, 2017 and Mar 1, 2019. In those reports, the most common specific device problems are “Degraded” (1) and “Electrical /Electronic Property Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the InnoSpire Elegance 00?

None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the InnoSpire Elegance 00 FDA approved?

It was cleared through 510(k) K042655 on Oct 13, 2004. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the InnoSpire Elegance 00, and is it still sold?

The FDA's device database lists Respironics Respiratory Drug Delivery (Uk) Ltd as the company responsible for it (brand InnoSpire Elegance, model 00). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Compressor, air, portable devices

All Compressor, air, portable devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page