NIPRO Cellentia™-H Hemodialyzer Cellentia™-17HRecalls & reported problems
Nipro Medical Corporation · CELLULOSE TRIACETATE HEMODIALYZER 1.7m² / Max.TMP 66kPa (500mmHg) · Class 2
All models of NIPRO Cellentia-H Hemodialyzer (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00383790004398), received between Dec 28, 2018 and Jun 3, 2026 — classified as 7 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak5
- Patient-Device Incompatibility5
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Itching Sensation4
- No Clinical Signs, Symptoms or Conditions4
- Dyspnea2
- Hemorrhage/Blood Loss/Bleeding2
- Syncope/Fainting2
- Hypersensitivity/Allergic reaction1
- Insufficient Information1
- Rash1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
InjuryJun 3, 2026Patient-Device Incompatibility
PATIENT HAS BEEN DIALYZING WITH THE USE OF NIPRO CELLENTIA 17H DIALYZER AS PRESCRIBED SINCE APPROXIMATELY (B)(6) 2025. PATIENT COMPLAINED OF SEVERE ITCHING ON (B)(6) 2025, PATIENT WAS ADMINISTERED BENADRYL 50 MG IVP. NIPRO CELLENTIA 17H DISCONTINUED AND ADDED AS AN ALLERGY. ADDITIONAL INFORMATION RECEIVED 02/25/2026: THE PATIENT WAS USING AN 180NRE OPTIFLUX DIALYZER ON (B)(6) 2025. THE DIALYZER WAS CHANGED TO OPTIFLUX 180NR ON (B)(6) 2026. THEN THE PATIENT WAS CHANGED AGAIN, TO THE CELLENTIA 17H ON (B)(6) 2025 (NO RATIONALE PROVIDED). THE AFFECTED LOT NUMBER IS 25D21E. BENADRYL, BOTH ORAL AND IV, WAS PRESCRIBED AND ADMINISTERED DURING THE TREATMENT, BUT IT DID NOT RESOLVE THE ITCHING. THE PATIENT REPORTED SEVERE ITCHING APPROXIMATELY 40 TO 60 MINUTES INTO TREATMENT, WHICH CONTINUED EVEN AFTER THE PATIENT RETURNED HOME. THE ITCHING SUBSIDED THE FOLLOWING DAY WHEN THE PATIENT DID NOT RECEIVE TREATMENT. THE PATIENT ARRIVED AT THE CLINIC WITHOUT ANY VISIBLE SIGNS OR SYMPTOMS OF ITCHING.
Patient problems: Itching Sensation
MAUDE report 25383716 ↗InjuryJun 3, 2026Patient-Device Incompatibility
PATIENT HAS BEEN DIALYZING WITH THE USE OF NIPRO CELLENTIA 17H DIALYZER AS PRESCRIBED SINCE APPROXIMATELY (B)(6) 2025. PATIENT COMPLAINED OF SEVERE ITCHING ON (B)(6) 2025, PATIENT WAS ADMINISTERED BENADRYL 50 MG IVP. NIPRO CELLENTIA 17H DISCONTINUED AND ADDED AS AN ALLERGY.
Patient problems: Itching Sensation
MAUDE report 25382972 ↗MalfunctionJan 22, 2026Fluid/Blood Leak
MW5168729 RECEIVED VIA EMAIL WITH THE FOLLOWING: IT WAS REPORTED TO (B)(6) VIA EMAIL THAT CELLENTIA HEMODIALYZER - SINGLE USE 17H HAD A BLOOD LEAK OCCURRENCE AT START OF TREATMENT. THIS REPORT REFLECTS INFORMATION RECEIVED BY FDA IN THE FORM OF A NOTIFICATION PER 803.22 (B)(2). NO CUSTOMER INFORMATION IS AVAILABLE TO REACH OUT TO THE CUSTOMER TO REQUEST ADDITIONAL INFORMATION RELATE TO THE EVENT.
Patient problems: Insufficient Information
MAUDE report 24158064 ↗InjuryMay 8, 2025Fluid/Blood Leak; Fluid/Blood Leak
CUSTOMER REPORTED BLOOD LEAK OCCURRENCE AS SOON AS THE TREATMENT STARTED, MACHINE ALARMED FOR BLOOD LEAK RIGHT AWAY. NO MEDICATIONS WERE GIVEN, NO PATIENT SYMPTOMS PRESENT, NO MEDICAL INTERVENTION REQUIRED. BLOOD WAS NOT VISIBLE, NO DEFECT OR DAMAGE DETECTED ON THE DIALYZER. BLOOD LEAK TEST STRIP WAS NEGATIVE. TREATMENT INTERRUPTED, APPROXIMATELY 200ML OF BLOOD LOSS. NEW MACHINE WAS SET UP WITH NEW SUPPLIES. PATIENT COMPLETED TREATMENT. DIALYSIS MACHINE: 2008T, FRESENIUS BLOODLINE USED, DIALYSATE FLOW RATE (DFR): AUTOFLOW 1.5, BLOOD FLOW RATE (BFR): 300. GENDER: MALE, WEIGHT: 93KG, DIAGNOSIS: AKI (ACUTE KIDNEY INJURY), PATIENT HAS BEEN ON DIALYSIS 3 TIMES A WEEK SINCE (B)(6) 2025.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 21985373 ↗MalfunctionMay 8, 2025Fluid/Blood Leak; Fluid/Blood Leak
CUSTOMER REPORTED BLOOD LEAK OCCURRENCE AS SOON AS THE TREATMENT STARTED, MACHINE ALARMED FOR BLOOD LEAK RIGHT AWAY. NO MEDICATIONS WERE GIVEN, NO PATIENT SYMPTOMS PRESENT, NO MEDICAL INTERVENTION REQUIRED. BLOOD WAS NOT VISIBLE, NO DEFECT OR DAMAGE DETECTED ON THE DIALYZER. BLOOD LEAK TEST STRIP WAS NEGATIVE. TREATMENT INTERRUPTED, APPROXIMATELY 200ML OF BLOOD LOSS. NEW MACHINE WAS SET UP WITH NEW SUPPLIES. PATIENT COMPLETED TREATMENT. DIALYSIS MACHINE: 2008T, FRESENIUS BLOODLINE USED, DIALYSATE FLOW RATE (DFR): AUTOFLOW 1.5, BLOOD FLOW RATE (BFR): 300 GENDER: MALE, WEIGHT: 93KG, DIAGNOSIS: AKI (ACUTE KIDNEY INJURY), PATIENT HAS BEEN ON DIALYSIS 3 TIMES A WEEK SINCE (B)(6) 2025.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 21984879 ↗MalfunctionOct 10, 2022Fluid/Blood Leak; Fluid/Blood Leak
2HOURS AND 30 MINUTES INTO HEMODIALYSIS TREATMENT, MACHINE ALARMED. TREATMENT STOPPED, BLOOD LEAK CONFIRMED WITH BLOOD LEAK TEST STRIP. NO VISIBLE BLOOD ON DIALYZER. BLOOD WAS NOT RETURNED, UNDETERMINED AMOUNT OF BLOOD LOSS, ESTIMATED MINIMAL. NO MEDICAL INTERVENTIONS OR HOSPITALIZATION. PRIMING CONDITIONS: BFR=150ML/MIN AND PRIMING VOLUME=1000ML. DIALYSIS CONDITIONS: TX TIME=3HRS, BFR=350ML/MIN, DFR=SEQUENTIAL (UF ONLY), ACCESS= AVF WITH 16G NEEDLE 1.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15573241 ↗MalfunctionOct 10, 2022Fluid/Blood Leak; Fluid/Blood Leak
2HOURS AND 30 MINUTES INTO HEMODIALYSIS TREATMENT, MACHINE ALARMED. TREATMENT STOPPED, BLOOD LEAK CONFIRMED WITH BLOOD LEAK TEST STRIP. NO VISIBLE BLOOD ON DIALYZER. BLOOD WAS NOT RETURNED, UNDETERMINED AMOUNT OF BLOOD LOSS, ESTIMATED MINIMAL. NO MEDICAL INTERVENTIONS OR HOSPITALIZATION. PRIMING CONDITIONS: BFR=150ML/MIN AND PRIMING VOLUME=1000ML. DIALYSIS CONDITIONS: TX TIME=3HRS, BFR=350ML/MIN, DFR=SEQUENTIAL (UF ONLY), ACCESS= AVF WITH 16G NEEDLE 1"
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15569515 ↗InjurySep 15, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
CHARGE RN STATED THAT CELLENTIA WAS PRIMED WITH 250 NS AND PATIENT TREATMENT STARTED. WITHIN 5 MINUTES OF TREATMENT START, PATIENT COMPLAINED OF FEELING FAINT AND SHORT OF BREATH. CHARGE NURSE STATED THAT THEY STOPPED THE TREATMENT AND DID NOT RETURN BLOOD TO PATIENT. STAFF REPORTED THAT NO MEDICATIONS WERE GIVEN TO THE PATIENT AND THE PATIENT VITAL SIGNS WERE STABLE. PATIENT REQUESTED TO GO TO THE EMERGENCY ROOM AND AMBULANCE WAS CALLED. PATIENT WAS DISCHARGED FROM HOSPITAL AND RETURNED TO THE OUTPATIENT CLINIC ON WEDNESDAY (B)(6) 2022 AT WHICH TIME SHE HAD A DIALYSIS TREATMENT USING CELLENTIA DD AGAIN AND COMPLETED THE TREATMENT WITHOUT ANY COMPLAINTS. PATIENT HAS BEEN USING CELLENTIA 17 DIALYZERS FOR A YEAR NOW AND THIS IS THE FIRST TIME SHE COMPLAINED OF PROBLEMS.
Patient problems: Dyspnea; Syncope/Fainting
MAUDE report 15431374 ↗InjurySep 15, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
CHARGE RN STATED THAT CELLENTIA WAS PRIMED WITH 250 NS AND PATIENT TREATMENT STARTED. WITHIN 5 MINUTES OF TREATMENT START, PATIENT COMPLAINED OF FEELING FAINT AND SHORT OF BREATH. CHARGE NURSE STATED THAT THEY STOPPED THE TREATMENT AND DID NOT RETURN BLOOD TO PATIENT. STAFF REPORTED THAT NO MEDICATIONS WERE GIVEN TO THE PATIENT AND THE PATIENT VITAL SIGNS WERE STABLE. PATIENT REQUESTED TO GO TO THE EMERGENCY ROOM AND AMBULANCE WAS CALLED. PATIENT WAS DISCHARGED FROM HOSPITAL AND RETURNED TO THE OUTPATIENT CLINIC ON WEDNESDAY (B)(6) 2022 AT WHICH TIME SHE HAD A DIALYSIS TREATMENT USING CELLENTIA DD AGAIN AND COMPLETED THE TREATMENT WITHOUT ANY COMPLAINTS. PATIENT HAS BEEN USING CELLENTIA 17 DIALYZERS FOR A YEAR NOW AND THIS IS THE FIRST TIME SHE COMPLAINED OF PROBLEMS.
Patient problems: Dyspnea; Syncope/Fainting
MAUDE report 15430976 ↗InjuryDec 28, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
CLINIC MANAGER REPORTED A PATIENT WHO HAD DIALYZER REACTIONS ON UNSPECIFIED DATES SINCE (B)(6) 2018. PATIENT'S FIRST DIALYSIS WAS (B)(6) 2018. NO REACTIONS WERE ENCOUNTERED BETWEEN (B)(6) 2018. PATIENT WAS USING THE OPTIFLUX 180NRE. THE CLINIC MANAGER FIRST THOUGHT THE REACTION WAS DUE TO THE GLOVES, GLOVES WERE CHANGED AND THE PATIENT CONTINUED TO HAVE ALLERGIC REACTION. DIALYZER WAS THEN CHANGED TO THE OPTIFLUX 180NR, BUT THE PATIENT CONTINUED TO HAVE ALLERGIC REACTION. THE CLINIC MANAGER THEN CHANGED TO THE NIPRO CELLENTIA 17H ON (B)(6) 2018, THE PATIENT REPORTEDLY HAD A FEW SMALL HIVES AND SLIGHT ITCHING DURING MID-TREATMENT TO LAST HOUR OF TREATMENT. THE CLINIC MANAGER CONFIRMED THAT THE DIALYZER WAS FLUSHED WITH NORMAL SALINE SOLUTION PRIOR TO TREATMENT, PRIMING FLUSHING DIALYZER WITH 2 LITER NSS EVERY TREATMENT PER PHYSICIAN ORDERS AND CONFIRMED THAT BENADRYL WAS ADMINISTERED TO THE PATIENT INTRAVENOUSLY BOTH PRE-TREATMENT AND MID-TREATMENT. THE CLINIC MANAGER CONFIRMED THE PATIENT COMPLETED THEIR TREATMENTS, AND CONFIRMED THE PATIENT'S SYMPTOMS RESOLVE AFTER TREATMENT IS COMPLETED. LOT NUMBERS WERE NOT RECORDED. PATIENT CONTINUES TO HAVE REACTION BUT IS IMPROVED IN SEVERITY
Patient problems: Itching Sensation; Rash; Reaction
MAUDE report 8203581 ↗InjuryDec 28, 2018Patient-Device Incompatibility
CLINIC MANAGER REPORTED A PATIENT WHO HAD DIALYZER REACTIONS ON UNSPECIFIED DATES SINCE (B)(6) 2018. PATIENT'S FIRST DIALYSIS WAS (B)(6) 2018. NO REACTIONS WERE ENCOUNTERED BETWEEN (B)(6) 2018 AND (B)(6) 2018. PATIENT WAS USING THE OPTIFLUX 180NRE. THE CLINIC MANAGER FIRST THOUGHT THE REACTION WAS DUE TO THE GLOVES, GLOVES WERE CHANGED AND THE PATIENT CONTINUED TO HAVE ALLERGIC REACTION. DIALYZER WAS THEN CHANGED TO THE OPTIFLUX 180NR, BUT THE PATIENT CONTINUED TO HAVE ALLERGIC REACTION. THE CLINIC MANAGER THEN CHANGED TO THE NIPRO CELLENTIA 17H ON (B)(6) 2018, THE PATIENT REPORTEDLY HAD A FEW SMALL HIVES AND SLIGHT ITCHING DURING MID-TREATMENT TO LAST HOUR OF TREATMENT. THE CLINIC MANAGER CONFIRMED THAT THE DIALYZER WAS FLUSHED WITH NORMAL SALINE SOLUTION PRIOR TO TREATMENT, PRIMING FLUSHING DIALYZER WITH 2 LITER NSS EVERY TREATMENT PER PHYSICIAN ORDERS AND CONFIRMED THAT BENADRYL WAS ADMINISTERED TO THE PATIENT INTRAVENOUSLY BOTH PRE-TREATMENT AND MID-TREATMENT. THE CLINIC MANAGER CONFIRMED THE PATIENT COMPLETED THEIR TREATMENTS, AND CONFIRMED THE PATIENT'S SYMPTOMS RESOLVE AFTER TREATMENT IS COMPLETED. LOT NUMBERS WERE NOT RECORDED. PATIENT CONTINUES TO HAVE REACTION BUT IS IMPR
Patient problems: Hypersensitivity/Allergic reaction; Itching Sensation; Urticaria
MAUDE report 8203353 ↗
Other Dialyzer, high permeability with or without sealed dialysate system devices
- Optiflux® 0500318EFresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system1 recall6 deaths2,849 reports
- 2008® Series Hemodialysis Machines 190713Fresenius Usa, Inc. · Dialyzer, high permeability with or without sealed dialysate system3 recalls37 deaths893 reports
- 2008T BlueStar Hemodialysis Machine 191126Fresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system5 recalls10 deaths441 reports
- 2008® Series Hemodialysis Machines 190766Fresenius Usa, Inc. · Dialyzer, high permeability with or without sealed dialysate system4 recalls15 deaths370 reports
- 2008T BlueStar Hemodialysis Machine 191130Fresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system4 recalls7 deaths272 reports
- 2008® Series Hemodialysis Machines 190610Fresenius Usa, Inc. · Dialyzer, high permeability with or without sealed dialysate system1 recall15 deaths335 reports
- Optiflux® 0500320EFresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system2 recalls242 reports
- REVACLEAR 300Gambro Renal Products, Inc. · Dialyzer, high permeability with or without sealed dialysate system3 recalls125 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page