NIPRO Cellentia™-H Hemodialyzer Cellentia™-17HRecalls & reported problems

Nipro Medical Corporation · CELLULOSE TRIACETATE HEMODIALYZER 1.7m² / Max.TMP 66kPa (500mmHg) · Class 2

All models of NIPRO Cellentia-H Hemodialyzer (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00383790004398), received between Dec 28, 2018 and Jun 3, 2026 — classified as 7 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak5
  • Patient-Device Incompatibility5
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Itching Sensation4
  • No Clinical Signs, Symptoms or Conditions4
  • Dyspnea2
  • Hemorrhage/Blood Loss/Bleeding2
  • Syncope/Fainting2
  • Hypersensitivity/Allergic reaction1
  • Insufficient Information1
  • Rash1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 3, 2026Patient-Device Incompatibility

    PATIENT HAS BEEN DIALYZING WITH THE USE OF NIPRO CELLENTIA 17H DIALYZER AS PRESCRIBED SINCE APPROXIMATELY (B)(6) 2025. PATIENT COMPLAINED OF SEVERE ITCHING ON (B)(6) 2025, PATIENT WAS ADMINISTERED BENADRYL 50 MG IVP. NIPRO CELLENTIA 17H DISCONTINUED AND ADDED AS AN ALLERGY. ADDITIONAL INFORMATION RECEIVED 02/25/2026: THE PATIENT WAS USING AN 180NRE OPTIFLUX DIALYZER ON (B)(6) 2025. THE DIALYZER WAS CHANGED TO OPTIFLUX 180NR ON (B)(6) 2026. THEN THE PATIENT WAS CHANGED AGAIN, TO THE CELLENTIA 17H ON (B)(6) 2025 (NO RATIONALE PROVIDED). THE AFFECTED LOT NUMBER IS 25D21E. BENADRYL, BOTH ORAL AND IV, WAS PRESCRIBED AND ADMINISTERED DURING THE TREATMENT, BUT IT DID NOT RESOLVE THE ITCHING. THE PATIENT REPORTED SEVERE ITCHING APPROXIMATELY 40 TO 60 MINUTES INTO TREATMENT, WHICH CONTINUED EVEN AFTER THE PATIENT RETURNED HOME. THE ITCHING SUBSIDED THE FOLLOWING DAY WHEN THE PATIENT DID NOT RECEIVE TREATMENT. THE PATIENT ARRIVED AT THE CLINIC WITHOUT ANY VISIBLE SIGNS OR SYMPTOMS OF ITCHING.

    Patient problems: Itching Sensation

    MAUDE report 25383716 ↗
  • InjuryJun 3, 2026Patient-Device Incompatibility

    PATIENT HAS BEEN DIALYZING WITH THE USE OF NIPRO CELLENTIA 17H DIALYZER AS PRESCRIBED SINCE APPROXIMATELY (B)(6) 2025. PATIENT COMPLAINED OF SEVERE ITCHING ON (B)(6) 2025, PATIENT WAS ADMINISTERED BENADRYL 50 MG IVP. NIPRO CELLENTIA 17H DISCONTINUED AND ADDED AS AN ALLERGY.

    Patient problems: Itching Sensation

    MAUDE report 25382972 ↗
  • MalfunctionJan 22, 2026Fluid/Blood Leak

    MW5168729 RECEIVED VIA EMAIL WITH THE FOLLOWING: IT WAS REPORTED TO (B)(6) VIA EMAIL THAT CELLENTIA HEMODIALYZER - SINGLE USE 17H HAD A BLOOD LEAK OCCURRENCE AT START OF TREATMENT. THIS REPORT REFLECTS INFORMATION RECEIVED BY FDA IN THE FORM OF A NOTIFICATION PER 803.22 (B)(2). NO CUSTOMER INFORMATION IS AVAILABLE TO REACH OUT TO THE CUSTOMER TO REQUEST ADDITIONAL INFORMATION RELATE TO THE EVENT.

    Patient problems: Insufficient Information

    MAUDE report 24158064 ↗
  • InjuryMay 8, 2025Fluid/Blood Leak; Fluid/Blood Leak

    CUSTOMER REPORTED BLOOD LEAK OCCURRENCE AS SOON AS THE TREATMENT STARTED, MACHINE ALARMED FOR BLOOD LEAK RIGHT AWAY. NO MEDICATIONS WERE GIVEN, NO PATIENT SYMPTOMS PRESENT, NO MEDICAL INTERVENTION REQUIRED. BLOOD WAS NOT VISIBLE, NO DEFECT OR DAMAGE DETECTED ON THE DIALYZER. BLOOD LEAK TEST STRIP WAS NEGATIVE. TREATMENT INTERRUPTED, APPROXIMATELY 200ML OF BLOOD LOSS. NEW MACHINE WAS SET UP WITH NEW SUPPLIES. PATIENT COMPLETED TREATMENT. DIALYSIS MACHINE: 2008T, FRESENIUS BLOODLINE USED, DIALYSATE FLOW RATE (DFR): AUTOFLOW 1.5, BLOOD FLOW RATE (BFR): 300. GENDER: MALE, WEIGHT: 93KG, DIAGNOSIS: AKI (ACUTE KIDNEY INJURY), PATIENT HAS BEEN ON DIALYSIS 3 TIMES A WEEK SINCE (B)(6) 2025.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21985373 ↗
  • MalfunctionMay 8, 2025Fluid/Blood Leak; Fluid/Blood Leak

    CUSTOMER REPORTED BLOOD LEAK OCCURRENCE AS SOON AS THE TREATMENT STARTED, MACHINE ALARMED FOR BLOOD LEAK RIGHT AWAY. NO MEDICATIONS WERE GIVEN, NO PATIENT SYMPTOMS PRESENT, NO MEDICAL INTERVENTION REQUIRED. BLOOD WAS NOT VISIBLE, NO DEFECT OR DAMAGE DETECTED ON THE DIALYZER. BLOOD LEAK TEST STRIP WAS NEGATIVE. TREATMENT INTERRUPTED, APPROXIMATELY 200ML OF BLOOD LOSS. NEW MACHINE WAS SET UP WITH NEW SUPPLIES. PATIENT COMPLETED TREATMENT. DIALYSIS MACHINE: 2008T, FRESENIUS BLOODLINE USED, DIALYSATE FLOW RATE (DFR): AUTOFLOW 1.5, BLOOD FLOW RATE (BFR): 300 GENDER: MALE, WEIGHT: 93KG, DIAGNOSIS: AKI (ACUTE KIDNEY INJURY), PATIENT HAS BEEN ON DIALYSIS 3 TIMES A WEEK SINCE (B)(6) 2025.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21984879 ↗
  • MalfunctionOct 10, 2022Fluid/Blood Leak; Fluid/Blood Leak

    2HOURS AND 30 MINUTES INTO HEMODIALYSIS TREATMENT, MACHINE ALARMED. TREATMENT STOPPED, BLOOD LEAK CONFIRMED WITH BLOOD LEAK TEST STRIP. NO VISIBLE BLOOD ON DIALYZER. BLOOD WAS NOT RETURNED, UNDETERMINED AMOUNT OF BLOOD LOSS, ESTIMATED MINIMAL. NO MEDICAL INTERVENTIONS OR HOSPITALIZATION. PRIMING CONDITIONS: BFR=150ML/MIN AND PRIMING VOLUME=1000ML. DIALYSIS CONDITIONS: TX TIME=3HRS, BFR=350ML/MIN, DFR=SEQUENTIAL (UF ONLY), ACCESS= AVF WITH 16G NEEDLE 1.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15573241 ↗
  • MalfunctionOct 10, 2022Fluid/Blood Leak; Fluid/Blood Leak

    2HOURS AND 30 MINUTES INTO HEMODIALYSIS TREATMENT, MACHINE ALARMED. TREATMENT STOPPED, BLOOD LEAK CONFIRMED WITH BLOOD LEAK TEST STRIP. NO VISIBLE BLOOD ON DIALYZER. BLOOD WAS NOT RETURNED, UNDETERMINED AMOUNT OF BLOOD LOSS, ESTIMATED MINIMAL. NO MEDICAL INTERVENTIONS OR HOSPITALIZATION. PRIMING CONDITIONS: BFR=150ML/MIN AND PRIMING VOLUME=1000ML. DIALYSIS CONDITIONS: TX TIME=3HRS, BFR=350ML/MIN, DFR=SEQUENTIAL (UF ONLY), ACCESS= AVF WITH 16G NEEDLE 1"

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15569515 ↗
  • InjurySep 15, 2022Patient-Device Incompatibility; Patient-Device Incompatibility

    CHARGE RN STATED THAT CELLENTIA WAS PRIMED WITH 250 NS AND PATIENT TREATMENT STARTED. WITHIN 5 MINUTES OF TREATMENT START, PATIENT COMPLAINED OF FEELING FAINT AND SHORT OF BREATH. CHARGE NURSE STATED THAT THEY STOPPED THE TREATMENT AND DID NOT RETURN BLOOD TO PATIENT. STAFF REPORTED THAT NO MEDICATIONS WERE GIVEN TO THE PATIENT AND THE PATIENT VITAL SIGNS WERE STABLE. PATIENT REQUESTED TO GO TO THE EMERGENCY ROOM AND AMBULANCE WAS CALLED. PATIENT WAS DISCHARGED FROM HOSPITAL AND RETURNED TO THE OUTPATIENT CLINIC ON WEDNESDAY (B)(6) 2022 AT WHICH TIME SHE HAD A DIALYSIS TREATMENT USING CELLENTIA DD AGAIN AND COMPLETED THE TREATMENT WITHOUT ANY COMPLAINTS. PATIENT HAS BEEN USING CELLENTIA 17 DIALYZERS FOR A YEAR NOW AND THIS IS THE FIRST TIME SHE COMPLAINED OF PROBLEMS.

    Patient problems: Dyspnea; Syncope/Fainting

    MAUDE report 15431374 ↗
  • InjurySep 15, 2022Patient-Device Incompatibility; Patient-Device Incompatibility

    CHARGE RN STATED THAT CELLENTIA WAS PRIMED WITH 250 NS AND PATIENT TREATMENT STARTED. WITHIN 5 MINUTES OF TREATMENT START, PATIENT COMPLAINED OF FEELING FAINT AND SHORT OF BREATH. CHARGE NURSE STATED THAT THEY STOPPED THE TREATMENT AND DID NOT RETURN BLOOD TO PATIENT. STAFF REPORTED THAT NO MEDICATIONS WERE GIVEN TO THE PATIENT AND THE PATIENT VITAL SIGNS WERE STABLE. PATIENT REQUESTED TO GO TO THE EMERGENCY ROOM AND AMBULANCE WAS CALLED. PATIENT WAS DISCHARGED FROM HOSPITAL AND RETURNED TO THE OUTPATIENT CLINIC ON WEDNESDAY (B)(6) 2022 AT WHICH TIME SHE HAD A DIALYSIS TREATMENT USING CELLENTIA DD AGAIN AND COMPLETED THE TREATMENT WITHOUT ANY COMPLAINTS. PATIENT HAS BEEN USING CELLENTIA 17 DIALYZERS FOR A YEAR NOW AND THIS IS THE FIRST TIME SHE COMPLAINED OF PROBLEMS.

    Patient problems: Dyspnea; Syncope/Fainting

    MAUDE report 15430976 ↗
  • InjuryDec 28, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CLINIC MANAGER REPORTED A PATIENT WHO HAD DIALYZER REACTIONS ON UNSPECIFIED DATES SINCE (B)(6) 2018. PATIENT'S FIRST DIALYSIS WAS (B)(6) 2018. NO REACTIONS WERE ENCOUNTERED BETWEEN (B)(6) 2018. PATIENT WAS USING THE OPTIFLUX 180NRE. THE CLINIC MANAGER FIRST THOUGHT THE REACTION WAS DUE TO THE GLOVES, GLOVES WERE CHANGED AND THE PATIENT CONTINUED TO HAVE ALLERGIC REACTION. DIALYZER WAS THEN CHANGED TO THE OPTIFLUX 180NR, BUT THE PATIENT CONTINUED TO HAVE ALLERGIC REACTION. THE CLINIC MANAGER THEN CHANGED TO THE NIPRO CELLENTIA 17H ON (B)(6) 2018, THE PATIENT REPORTEDLY HAD A FEW SMALL HIVES AND SLIGHT ITCHING DURING MID-TREATMENT TO LAST HOUR OF TREATMENT. THE CLINIC MANAGER CONFIRMED THAT THE DIALYZER WAS FLUSHED WITH NORMAL SALINE SOLUTION PRIOR TO TREATMENT, PRIMING FLUSHING DIALYZER WITH 2 LITER NSS EVERY TREATMENT PER PHYSICIAN ORDERS AND CONFIRMED THAT BENADRYL WAS ADMINISTERED TO THE PATIENT INTRAVENOUSLY BOTH PRE-TREATMENT AND MID-TREATMENT. THE CLINIC MANAGER CONFIRMED THE PATIENT COMPLETED THEIR TREATMENTS, AND CONFIRMED THE PATIENT'S SYMPTOMS RESOLVE AFTER TREATMENT IS COMPLETED. LOT NUMBERS WERE NOT RECORDED. PATIENT CONTINUES TO HAVE REACTION BUT IS IMPROVED IN SEVERITY

    Patient problems: Itching Sensation; Rash; Reaction

    MAUDE report 8203581 ↗
  • InjuryDec 28, 2018Patient-Device Incompatibility

    CLINIC MANAGER REPORTED A PATIENT WHO HAD DIALYZER REACTIONS ON UNSPECIFIED DATES SINCE (B)(6) 2018. PATIENT'S FIRST DIALYSIS WAS (B)(6) 2018. NO REACTIONS WERE ENCOUNTERED BETWEEN (B)(6) 2018 AND (B)(6) 2018. PATIENT WAS USING THE OPTIFLUX 180NRE. THE CLINIC MANAGER FIRST THOUGHT THE REACTION WAS DUE TO THE GLOVES, GLOVES WERE CHANGED AND THE PATIENT CONTINUED TO HAVE ALLERGIC REACTION. DIALYZER WAS THEN CHANGED TO THE OPTIFLUX 180NR, BUT THE PATIENT CONTINUED TO HAVE ALLERGIC REACTION. THE CLINIC MANAGER THEN CHANGED TO THE NIPRO CELLENTIA 17H ON (B)(6) 2018, THE PATIENT REPORTEDLY HAD A FEW SMALL HIVES AND SLIGHT ITCHING DURING MID-TREATMENT TO LAST HOUR OF TREATMENT. THE CLINIC MANAGER CONFIRMED THAT THE DIALYZER WAS FLUSHED WITH NORMAL SALINE SOLUTION PRIOR TO TREATMENT, PRIMING FLUSHING DIALYZER WITH 2 LITER NSS EVERY TREATMENT PER PHYSICIAN ORDERS AND CONFIRMED THAT BENADRYL WAS ADMINISTERED TO THE PATIENT INTRAVENOUSLY BOTH PRE-TREATMENT AND MID-TREATMENT. THE CLINIC MANAGER CONFIRMED THE PATIENT COMPLETED THEIR TREATMENTS, AND CONFIRMED THE PATIENT'S SYMPTOMS RESOLVE AFTER TREATMENT IS COMPLETED. LOT NUMBERS WERE NOT RECORDED. PATIENT CONTINUES TO HAVE REACTION BUT IS IMPR

    Patient problems: Hypersensitivity/Allergic reaction; Itching Sensation; Urticaria

    MAUDE report 8203353 ↗

Other Dialyzer, high permeability with or without sealed dialysate system devices

All Dialyzer, high permeability with or without sealed dialysate system devices →