NIPRO Cellentia™-H Hemodialyzer Cellentia™-21HRecalls & reported problems
Nipro Medical Corporation · CELLULOSE TRIACETATE HEMODIALYZER 2.1m² / Max.TMP 66kPa (500mmHg) · Class 2
All models of NIPRO Cellentia-H Hemodialyzer (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00383790004411), received between Aug 14, 2019 and Dec 22, 2023 — classified as 4 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak2
- Obstruction of Flow2
- Patient-Device Incompatibility2
- Device Alarm System1
- Insufficient Flow or Under Infusion1
Patient problems
- Dysphasia2
- Dyspnea2
- No Clinical Signs, Symptoms or Conditions2
- Blood Loss1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryDec 22, 2023Patient-Device Incompatibility; Patient-Device Incompatibility
PATIENT ATTENDED REGULARLY SCHEDULED IN-CENTER DIALYSIS TREATMENT. PRE VITAL SIGNS BP-149/82, PULSE-101, RESPIRATION-21, TEMPERATURE - 97.8. AT APPROX 0954 DIALYSIS WAS INITIATED WITH A USE OF TREATMENT ENDED, PATIENT GIVEN DIPHENHYDRAMINE 50 MG IVP(INTRAVENOUS PUSH). PATIENT DEVELOPED SHORTNESS OF BREATH, SLURRED SPEECH AND RIGHT FACIAL DROOP. 911 CALLED, PATIENT TRANSPORTED TO HOSPITAL VIA EMS(EMERGENCY MEDICAL SERVICES) AND LATER RELEASED. PATIENT RETURNED TO FACILITY (B)(6) 2023, TREATED ON OPTIFLUX 180NR DIALYZER WITHOUT INCIDENT. NIPRO CELLENTIA 21 ADDED AS ALLERGY.
Patient problems: Dysphasia; Dyspnea
MAUDE report 18389304 ↗InjuryDec 22, 2023Patient-Device Incompatibility; Patient-Device Incompatibility
PATIENT ATTENDED REGULARLY SCHEDULED IN-CENTER DIALYSIS TREATMENT. PRE VITAL SIGNS BP-149/82, PULSE-101, RESPIRATION-21, TEMPERATURE - 97.8. AT APPROX 0954 DIALYSIS WAS INITIATED WITH A USE OF TREATMENT ENDED, PATIENT GIVEN DIPHENHYDRAMINE 50 MG IVP(INTRAVENOUS PUSH). PATIENT DEVELOPED SHORTNESS OF BREATH, SLURRED SPEECH AND RIGHT FACIAL DROOP. 911 CALLED, PATIENT TRANSPORTED TO HOSPITAL VIA EMS(EMERGENCY MEDICAL SERVICES) AND LATER RELEASED. PATIENT RETURNED TO FACILITY (B)(6) 2023, TREATED ON OPTIFLUX 180NR DIALYZER WITHOUT INCIDENT. NIPRO CELLENTIA 21 ADDED AS ALLERGY.
Patient problems: Dysphasia; Dyspnea
MAUDE report 18388822 ↗InjuryMar 23, 2021Fluid/Blood Leak; Fluid/Blood Leak
1 PATIENT / 2 OCCURRENCES: WHERE THE STAFF HAVE NOTICED MASSIVE AIR BUBBLES IN THE DIALYZER, STAFF DID WHATEVER THEY COULD BUT WERE NOT ABLE TO CLEAR THE BUBBLES AND HAD TO WASTE THE PATIENT'S BLOOD. ADDITIONAL INFORMATION RECEIVED 03/02/2021: PRIMING CONDITIONS - THEY PRIME WITH A BFR OF 100 AND FLUSHED THE DIALYZER WITH AT LEAST A L OF SALINE. DIALYSIS CONDITIONS: BFR 400, DFR: 800, TREATMENT TIME 4 HRS AND NO HEPARIN BRAND/MANUFACTURER NAME & MODELS OF DEVICES USED TOGETHER WITH DIALYZER: GAMBRO MACHINE AND LINES. HOW MUCH BLOOD DID THE PATIENT LOST DUE TO CHANGING THE DIALYZER - 300 ML WAS ANY MEDICAL ACTIONS TAKEN- NONE AS THEIR HGB (AS CLAIMED WHEN THEY RECHECKED) WAS STABLE. WHERE DID THE BUBBLES GENERATE: MIDDLE TO TOP PORTION OF THE DIALYZER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11558902 ↗MalfunctionMar 23, 2021Fluid/Blood Leak; Fluid/Blood Leak
1 PATIENT / 2 OCCURRENCES - WHERE THE STAFF HAVE NOTICED MASSIVE AIR BUBBLES IN THE DIALYZER, STAFF DID WHATEVER THEY COULD BUT WERE NOT ABLE TO CLEAR THE BUBBLES AND HAD TO WASTE THE PATIENT'S BLOOD. ADDITIONAL INFORMATION RECEIVED 03/02/2021: 1. PRIMING CONDITIONS - THEY PRIME WITH A BFR OF 100 AND FLUSHED THE DIALYZER WITH AT LEAST A L OF SALINE. 2. DIALYSIS CONDITIONS - BFR 400, DFR - 800, TREATMENT TIME 4 HRS AND NO HEPARIN. 3. BRAND/MANUFACTURER NAME & MODELS OF DEVICES USED TOGETHER WITH DIALYZER - GAMBRO MACHINE AND LINES. 4. HOW MUCH BLOOD DID THE PATIENT LOST DUE TO CHANGING THE DIALYZER - 300 ML. 5. WAS ANY MEDICAL ACTIONS TAKEN- NONE AS THEIR HGB (AS CLAIMED WHEN THEY RECHECKED) WAS STABLE. 6. WHERE DID THE BUBBLES GENERATE - MIDDLE TO TOP PORTION OF THE DIALYZER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11558860 ↗MalfunctionAug 14, 2019Obstruction of Flow; Obstruction of Flow
ON (B)(6) 2019, WITH 1HR AND 20 MINUTES LEFT INTO A 4.25HR TREATMENT, THE SYSTEM CLOTTED AND THE STAFF WAS UNABLE TO RETURN THE PATIENT'S BLOOD. THE CLINIC SET UP A NEW SYSTEM AND PATIENT WAS ABLE TO COMPLETE THE REMAINING 1HR AND 20 MINUTES. PATIENT USED TO RECEIVE HEPARIN BUT WAS STOPPED DUE TO VAGINAL BLEEDING. STAFF DID NOT RECORD THE LOT NUMBER FOR THE DIALYZER USED. CLINIC CHECKED THE PATIENT'S HEMOGLOBIN LEVELS AND THEY WERE 9.2G/DL. (DIALYZER WAS DISPOSED) PATIENT HAS A CVC(DIALYSIS CATHETER), BFR IS 400, DFR IS 800. ON (B)(6): 2ND INCIDENT, SAME PATIENT, WITH 1HR AND 30 MINUTES LEFT INTO A 4.25HR TREATMENT, MACHINE ALARMED FOR HIGH VENOUS PRESSURE, CLOTS WERE NOTED ON THE BLOODLINES AND THE DIALYZER, BUT THE STAFF WAS ABLE TO RETURN THE PATIENT'S BLOOD. PATIENT RECEIVED 2000 UNITS OF BOLUS HEPARIN, CLINIC IS GIVING THE PATIENT 100ML OF HEPARIN FLUSHES EVERY HOUR, AND WILL DO CONTINUOUS FLUSHES OF 300ML/HR WITH HER NEXT SCHEDULED TREATMENT. HEMOGLOBIN LEVELS WERE CHECKED AND THE RESULTS WERE 9 G/DL. PATIENT IS NOW ON MAX DOES OF ARANESP.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8895285 ↗InjuryAug 14, 2019Device Alarm System; Insufficient Flow or Under Infusion; Obstruction of Flow
ON (B)(6) 2019, WITH 1 HR AND 20 MINUTES LEFT INTO A 4.25 HR TREATMENT, THE SYSTEM CLOTTED AND THE STAFF WAS UNABLE TO RETURN THE PATIENT'S BLOOD. THE CLINIC SET UP A NEW SYSTEM AND PATIENT WAS ABLE TO COMPLETE THE REMAINING 1 HR AND 20 MINUTES. PATIENT USED TO RECEIVE HEPARIN BUT WAS STOPPED DUE TO VAGINAL BLEEDING. STAFF DID NOT RECORD THE LOT NUMBER FOR THE DIALYZER USED. CLINIC CHECKED THE PATIENT'S HEMOGLOBIN LEVELS AND THEY WERE 9.2 G/DL. (DIALYZER WAS DISPOSED). PATIENT HAS A CVC(DIALYSIS CATHETER), BFR IS 400, DFR IS 800. ON (B)(6): 2ND INCIDENT, SAME PATIENT, WITH 1 HR AND 30 MINUTES LEFT INTO A 4.25 HR TREATMENT, MACHINE ALARMED FOR HIGH VENOUS PRESSURE, CLOTS WERE NOTED ON THE BLOODLINES AND THE DIALYZER, BUT THE STAFF WAS ABLE TO RETURN THE PATIENT'S BLOOD. PATIENT RECEIVED 2000 UNITS OF BOLUS HEPARIN, CLINIC IS GIVING THE PATIENT 100 ML OF HEPARIN FLUSHES EVERY HOUR, AND WILL DO CONTINUOUS FLUSHES OF 300 ML/HR WITH HER NEXT SCHEDULED TREATMENT. HEMOGLOBIN LEVELS WERE CHECKED AND THE RESULTS WERE 9 G/DL. PATIENT IS NOW ON MAX DOES OF ARANESP.
Patient problems: Blood Loss
MAUDE report 8895172 ↗
Other Dialyzer, high permeability with or without sealed dialysate system devices
- Optiflux® 0500318EFresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system1 recall6 deaths2,849 reports
- 2008® Series Hemodialysis Machines 190713Fresenius Usa, Inc. · Dialyzer, high permeability with or without sealed dialysate system3 recalls37 deaths893 reports
- 2008T BlueStar Hemodialysis Machine 191126Fresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system5 recalls10 deaths441 reports
- 2008® Series Hemodialysis Machines 190766Fresenius Usa, Inc. · Dialyzer, high permeability with or without sealed dialysate system4 recalls15 deaths370 reports
- 2008T BlueStar Hemodialysis Machine 191130Fresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system4 recalls7 deaths272 reports
- 2008® Series Hemodialysis Machines 190610Fresenius Usa, Inc. · Dialyzer, high permeability with or without sealed dialysate system1 recall15 deaths335 reports
- Optiflux® 0500320EFresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system2 recalls242 reports
- REVACLEAR 300Gambro Renal Products, Inc. · Dialyzer, high permeability with or without sealed dialysate system3 recalls125 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.