NIPRO ELISIO™ -H HEMODIALYZER ELISIO™-15HRecalls & reported problems
Nipro Medical Corporation · Surface area 1.5m² Max.TMP:66kPa(500mmHg) · Class 2
All models of NIPRO ELISIO -H HEMODIALYZER (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00383790007092), received between Oct 2, 2018 and Dec 28, 2018 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Leak/Splash” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Leak/Splash2
- Device Displays Incorrect Message1
- Patient-Device Incompatibility1
Patient problems
- Blood Loss2
- Fever1
- Infiltration into Tissue1
- Loss of consciousness1
- Low Blood Pressure/ Hypotension1
- Low Oxygen Saturation1
- Rash1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 28, 2018Leak/Splash; Leak/Splash
27 MIN INTO THE DIALYSIS TREATMENT, THE MACHINE ALARMED FOR MINOR BLOOD LEAK, THE STAFF TESTED IT WITH A BLOOD LEAK TEST STRIP AND RESULTS WERE POSITIVE. THEY WERE UNABLE TO RETURN THE PATIENT'S BLOOD, THE MACHINE WAS SWITCHED, AND WHILE SETTING UP NEW DIALYZER AND BLOODLINES, THE PATIENT INFILTRATED AND COULD NOT RESTART TREATMENT. PATIENT LEFT THE DIALYSIS UNIT STABLE, TEMPERATURE WAS 96.5 RIGHT AFTER MACHINE ALARMED FOR BLOOD LEAK, TEMPERATURE WAS 96.4 WHEN DISCHARGED FROM UNIT. SINCE THE STAFF WAS UNABLE TO RETURN THE PATIENT'S BLOOD, 300ML OF BLOOD WAS LOST. PATIENT'S PHYSICIAN AND CLINICAL COORDINATOR WERE AWARE OF THE ISSUE. PHYSICIAN INCREASED DOSE OF VENOFER AND ORDER TO RECHECK HEMOGLOBIN LEVEL ON NEXT SCHEDULED TREATMENT. PATIENT'S REGULAR DIALYSIS TREATMENT WAS RESCHEDULED FOR (B)(6) 2018 DUE TO INFILTRATION, STAFF WAS UNABLE TO DRAW LABS. PATIENT IS BEING SCHEDULED FOR VASCULAR ACCESS INTERVENTION (ANGIOPLASTY) ON (B)(6) 2018. TREATMENT WAS STARTED AT 6:07 AM AND MEDICATION WAS GIVEN AT 6:28 AM. PATIENT'S ACCESS IS A LEFT UPPER AV FISTULA. TREATMENT ORDERS: 210 MINS TREATMENT, 3K, 2.5CA, BICRAB 32, SODIUM 138, BFR 400, DFR 600, HEPARIN BOLUS 2000 UNITS AND 600 UNITS/HO
Patient problems: Blood Loss
MAUDE report 8202384 ↗InjuryDec 27, 2018Leak/Splash; Device Displays Incorrect Message
27 MIN INTO THE DIALYSIS TREATMENT, THE MACHINE ALARMED FOR MINOR BLOOD LEAK, THE STAFF TESTED IT WITH A BLOOD LEAK TEST STRIP AND RESULTS WERE POSITIVE. THEY WERE UNABLE TO RETURN THE PATIENT'S BLOOD, THE MACHINE WAS SWITCHED, AND WHILE SETTING UP NEW DIALYZER AND BLOODLINES, THE PATIENT INFILTRATED AND COULD NOT RESTART TREATMENT. PATIENT LEFT THE DIALYSIS UNIT STABLE, TEMPERATURE WAS 96.5 RIGHT AFTER MACHINE ALARMED FOR BLOOD LEAK, TEMPERATURE WAS 96.4 WHEN DISCHARGED FROM UNIT. SINCE THE STAFF WAS UNABLE TO RETURN THE PATIENT'S BLOOD, 300 ML OF BLOOD WAS LOST. PATIENT'S PHYSICIAN AND CLINICAL COORDINATOR WERE AWARE OF THE ISSUE. PHYSICIAN INCREASED DOSE OF VENOFER AND ORDER TO RECHECK HEMOGLOBIN LEVEL ON NEXT SCHEDULED TREATMENT. PATIENT'S REGULAR DIALYSIS TREATMENT WAS RESCHEDULED FOR (B)(6) 2018 DUE TO INFILTRATION, STAFF WAS UNABLE TO DRAW LABS. PATIENT IS BEING SCHEDULED FOR VASCULAR ACCESS INTERVENTION (ANGIOPLASTY) ON (B)(6) 2018. TREATMENT WAS STARTED AT 6:07 AM AND MEDICATION WAS GIVEN AT 6:28 AM. PATIENT'S ACCESS IS A LEFT UPPER AV FISTULA. TREATMENT ORDERS: 210 MINS TREATMENT, 3K, 2.5 CA, BICARB 32, SODIUM 138, BFR 400, DFR 600, HEPARIN BOLUS 2000 UNITS AND 600 UNITS/
Patient problems: Blood Loss; Infiltration into Tissue
MAUDE report 8201183 ↗InjuryOct 2, 2018Patient-Device Incompatibility; Patient-Device Incompatibility
ON JUNE 29, 2018 THE MANUFACTURER WAS ADVISED OF AN ADVERSE EVENT THAT OCCURRED IN (B)(6) 2016 (EXACT DATE UNKNOWN), WHILE USING A DIALYZER MODEL THAT WAS NOT SIMILAR TO ANY MODEL SOLD IN THE UNITED STATES. COMPLAINT DETAILS: THE PATIENT WAS USING A CTA DIALYZER AND SHE WAS CHANGED TO NIPRO PES DIALYZER. AFTER THE CHANGE, THE PATIENT HAD DETERIORATION OF THE SKIN AND ITCHING. ON (B)(6) OF 2017, AFTER INITIATION OF TREATMENT, THE PATIENT'S BLOOD PRESSURE AND OXYGEN LEVEL DECREASED. THE PATIENT WAS HOSPITALIZED, HAD FEVER AND SKIN RASH WITH INCREASING EOSINOPHILS AND HIGH LEVEL OF IGE. ALLERGY TO PES DIALYZER WAS SUSPECTED. THE DIALYZER WAS CHANGED TO THE CTA DIALYZER. DIALYSIS WAS THEN COMPLETED WITHOUT SYMPTOMS. EOSINOPHILS AND IGE LEVELS DECREASED, SKIN ITCHING WAS IMPROVED. THE PATIENT WAS THEN DISCHARGED FROM THE HOSPITAL, DIALYSIS HAS CONTINUED TO BE STABLE USING THE CTA DIALYZER. ON SEPTEMBER 28, 2018, THE MANUFACTURER WAS ADVISED THAT THE PRODUCT CODE REPORTED WAS INCORRECT. ONCE CORRECTED, IT WAS DETERMINED THAT THE CORRECT CODE HAS A SIMILAR MODEL SOLD IN THE UNITED STATES. MDR REPORTABILITY WAS THEN INITIATED.
Patient problems: Fever; Loss of consciousness; Low Blood Pressure/ Hypotension; Low Oxygen Saturation; Rash
MAUDE report 7927938 ↗
Other Dialyzer, high permeability with or without sealed dialysate system devices
- Optiflux® 0500318EFresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system1 recall6 deaths2,849 reports
- 2008® Series Hemodialysis Machines 190713Fresenius Usa, Inc. · Dialyzer, high permeability with or without sealed dialysate system3 recalls37 deaths893 reports
- 2008T BlueStar Hemodialysis Machine 191126Fresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system5 recalls10 deaths441 reports
- 2008® Series Hemodialysis Machines 190766Fresenius Usa, Inc. · Dialyzer, high permeability with or without sealed dialysate system4 recalls15 deaths370 reports
- 2008T BlueStar Hemodialysis Machine 191130Fresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system4 recalls7 deaths272 reports
- 2008® Series Hemodialysis Machines 190610Fresenius Usa, Inc. · Dialyzer, high permeability with or without sealed dialysate system1 recall15 deaths335 reports
- Optiflux® 0500320EFresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system2 recalls242 reports
- REVACLEAR 300Gambro Renal Products, Inc. · Dialyzer, high permeability with or without sealed dialysate system3 recalls125 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page