Skip to content

Unistik 3 AT 1012Recalls & reported problems

Owen Mumford Limited · Unistik 3 Extra - 21G x 2.0mm, sterile, single use safety lancet · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00384701012013), received between Apr 26, 2019 and Feb 24, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Retraction Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Retraction Problem2
  • Fail-Safe Did Not Operate1
  • Fail-Safe Problem1
  • Protective Measures Problem1

Patient problems

  • Insufficient Information2
  • Anxiety1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 24, 2025Retraction Problem; Retraction Problem

    HOSPITAL SENT EMAIL OF AN ISSUE THEY HAD EXPERIENCED WITH OUR LANCETS. THE EMAIL STATED SOME OF THE LANCETS, DEVICE DOES NOT RETRACT NEEDLE AFTER USE. THEY REPORTED THE NEEDLE LOOSELY DROPS OUT BUT DOESN'T REMAIN WITH THE ORANGE BASE. NO OTHER DETAILS WERE PROVIDED, NOR A SAMPLE SENT.

    Patient problems: Insufficient Information

    MAUDE report 21453337 ↗
  • MalfunctionFeb 24, 2025Retraction Problem; Retraction Problem

    HOSPITAL SENT EMAIL OF AN ISSUE THEY HAD EXPERIENCED WITH OUR LANCETS. THE EMAIL STATED SOME OF THE LANCETS, DEVICE DOES NOT RETRACT NEEDLE AFTER USE. THEY REPORTED THE NEEDLE LOOSELY DROPS OUT BUT DOESN'T REMAIN WITH THE ORANGE BASE. NO OTHER DETAILS WERE PROVIDED, NOR A SAMPLE SENT.

    Patient problems: Insufficient Information

    MAUDE report 21453284 ↗
  • MalfunctionApr 26, 2019Fail-Safe Problem; Protective Measures Problem; Fail-Safe Did Not Operate; Fail-Safe Problem; Protective Measures Problem; Fail-Safe Did Not Operate

    LABORATORY TECHNICAL SUPERVISOR EMAILED AFTER AN EMPLOYEE REPORTED TO HER THAT SHE NOTICED SOME OF THE LANCETS WERE NOT RETRACTING AFTER USE.

    Patient problems: Anxiety

    MAUDE report 8557828 ↗

Other Lancet, Blood devices

All Lancet, Blood devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page