STAT PROFILE PRIME CCS COMP (REF 52857)Recalls & reported problems
Nova Biomedical Corporation · Prime CCS Comp Analyzer · Class 2
All models of STAT PROFILE (13)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00385480528573), received between Mar 31, 2022 and Apr 5, 2022 — classified as 5 malfunctions.
The most frequently reported specific device problem is “High Test Results” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K141907); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- High Test Results3
- Low Test Results2
- Incorrect Measurement1
- No Apparent Adverse Event1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 5, 2022Incorrect Measurement; High Test Results; No Apparent Adverse Event; Incorrect Measurement; High Test Results; No Apparent Adverse Event
NOVA BIOMEDICAL (NOVA) WAS MADE AWARE OF A POTENTIAL ISSUE REGARDING PCO2 RESULTS DURING THE TREATMENT OF A PATIENT ON A STATPROFILE PRIME CCS COMP ANALYZER WITH SERIAL NUMBER (B)(4). THE USER DID NOT BELIEVE THE RESULTS PRODUCED WERE ACCURATE AND QUESTIONED THE RESULTS BEFORE PROVIDING ANY TREATMENT. THE SAME SAMPLE WAS RUN ACROSS ANOTHER STATPROFILE PRIME PLUS CCS COMP ANALYZER WHICH CONFIRMED THE INITIAL RESULTS WERE INCORRECT. NO PATIENT HARM OR ADVERSE TREATMENT DUE TO THE INACCURATE RESULTS WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14014864 ↗MalfunctionMar 31, 2022Low Test Results; Low Test Results
THE CUSTOMER REPORTED DISCREPANT RESULTS ON PRIME ANALYZER (B)(4) FOR PCO2. THE CLINICIAN QUESTIONED THE RESULTS AND RAN THE SAMPLE AGAIN ON A DIFFERENT PRIME ANALYZER. THIS SECOND TEST WAS CLOSER TO THE PATIENT HISTORY AND CENTRAL LAB DATA. THERE WAS NO PATIENT HARM OR INTERVENTION REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13964673 ↗MalfunctionMar 31, 2022High Test Results; High Test Results
THE CUSTOMER REPORTED DISCREPANT RESULTS ON PRIME ANALYZER (B)(4) FOR PCO2. THE CLINICIAN QUESTIONED THE RESULTS AND RAN THE SAMPLE AGAIN ON A DIFFERENT PRIME ANALYZER. THIS SECOND TEST WAS CLOSER TO THE PATIENT HISTORY AND CENTRAL LAB DATA. THERE WAS NO PATIENT HARM OR INTERVENTION REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13964415 ↗MalfunctionMar 31, 2022High Test Results; High Test Results
THE CUSTOMER REPORTED DISCREPANT RESULTS ON PRIME ANALYZER P07A17210C FOR PCO2. THE CLINICIAN QUESTIONED THE RESULTS AND RAN THE SAMPLE AGAIN ON A DIFFERENT PRIME ANALYZER. THIS SECOND TEST WAS CLOSER TO THE PATIENT HISTORY AND CENTRAL LAB DATA. THERE WAS NO PATIENT HARM OR INTERVENTION REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13964315 ↗MalfunctionMar 31, 2022Low Test Results; Low Test Results
THE CUSTOMER REPORTED DISCREPANT RESULTS ON PRIME ANALYZER P07A17210C FOR PCO2. THE CLINICIAN QUESTIONED THE RESULTS AND RAN THE SAMPLE AGAIN ON A DIFFERENT PRIME ANALYZER. THIS SECOND TEST WAS CLOSER TO THE PATIENT HISTORY AND CENTRAL LAB DATA. THERE WAS NO PATIENT HARM OR INTERVENTION REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13964206 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K141907. One submission often covers many models, so check the product and lot details in each record.
- Z-2732-2017 · Jun 19, 2017 · TerminatedNova Biomedical Stat Profile Prime Analyzer as follows: PN 52855 Stat Profile Prime ABG Analyzer PN 53655 Stat Profile Prime ABG Analyzer with Scanner PN 52856 Stat Profile Prime C
Other ELECTRODE, ION SPECIFIC, CALCIUM devices
- RAPIDPoint® 500e Blood Gas System 11416751Siemens Healthcare Diagnostics Inc. · ELECTRODE, ION SPECIFIC, CALCIUM2 recalls7 reports
- epoc Host2 11413524Epocal Inc · ELECTRODE, ION SPECIFIC, CALCIUM1 recall1 report
- Roche 9180 Electrolyte Analyzer 3157334001Roche Diagnostics GmbH · ELECTRODE, ION SPECIFIC, CALCIUM50 reports
- RAPIDLab® 1200 Series 1265 10321852Siemens Healthcare Diagnostics Inc. · ELECTRODE, ION SPECIFIC, CALCIUM12 reports
- STAT PROFILE PHOX ULTRANova Biomedical Corporation · ELECTRODE, ION SPECIFIC, CALCIUM8 reports
- STAT PROFILE Prime CCS CompNova Biomedical Corporation · ELECTRODE, ION SPECIFIC, CALCIUM3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page