Skip to content

epoc Host2 11413524Recalls & reported problems

Epocal Inc · The epoc Blood Analysis System, is intended for use by trained medical professionals as an in vitro diagnostic device for thequantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory orat the point of care in hospitals, nursing homes or other clinical care institutions. · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00630414606460), received between Nov 4, 2022 and Nov 4, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “High Test Results” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

4 other recalls cover devices cleared under the same FDA submission (K061597, K090109, K092849, K093297, K113726, K171247); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Dec 12, 2022Class IIOpen, Classified

    pO2 High Recovery in Quality Control When Adjusted for Barometric Pressure

    Z-1214-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • High Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 4, 2022High Test Results; High Test Results

    THE CUSTOMER REPORTED THAT THEIR EPOC INSTRUMENT GAVE A DISCREPANT HIGH SODIUM RESULT COMPARED TO RETESTING OF THE SAME SAMPLE ON A NON-SIEMENS INSTRUMENT. THERE IS NO REPORT OF INJURY DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15729710 ↗

FDA clearance and classification

Cleared through 510(k) K171247 on Jan 17, 2018 (decision: Substantially Equivalent). Submitted by Epocal, Inc. as “epoc Blood Urea Nitrogen Test, epoc Total Carbon Dioxide Test”. The FDA review took 264 days. FDA record ↗

Cleared through 510(k) K113726 on Oct 5, 2012 (decision: Substantially Equivalent). Submitted by Epocal, Inc. as “Epoc Chloride Test And Epoc Creatinine Test”. The FDA review took 291 days. FDA record ↗

Cleared through 510(k) K093297 on Jun 9, 2010 (decision: Substantially Equivalent). Submitted by Epocal, Inc. as “Epoc Bgem, Epoc Reader, Epoc Host, Epoc Care-Fill Capillary Tube”. The FDA review took 231 days. FDA record ↗

…and 3 more FDA submissions listed for this device.

The FDA regulates product code JFP (electrode, ion specific, calcium) under 21 CFR 862.1145 as a Class II device. 21 CFR 862.1145 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K061597, K090109, K092849, K093297, K113726, K171247. One submission often covers many models, so check the product and lot details in each record.

  • Z-1971-2013 · Mar 13, 2013 · Terminated
    epoc BGEM Test Card (Catalogue number CT-1004-00-00). quantitative testing of samples for pH, pCO2, pO2, Sodium Potassium , Ionized Calcium, Glucose, Lactate and Hematocrit.
  • Z-2034-2012 · Jun 28, 2011 · Terminated
    epoc Blood Analysis BGEM Test Cards for use with the epoc(R) Blood Analysis System Product Usage: Usage: epoc Blood Analysis BGEM Test Cards are used with the epoc(R) Blood
  • Z-1287-2011 · Oct 27, 2010 · Terminated
    02.122.585, 85mm Spiral Blade, 14mm, Right - Synthes Spiral Blade Component of the Synthes Modular Blade Plate System For fixation of fractures and osteotomies of the proximal and
  • Z-0952-2009 · Oct 16, 2008 · Terminated
    EPOC BGE Test Card The EPOC Blood Analysis System is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of whole blo

Common questions about the epoc Host2 11413524

Has the epoc Host2 11413524 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1214-2023 (Class II), began on Dec 12, 2022: pO2 High Recovery in Quality Control When Adjusted for Barometric Pressure. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the epoc Host2 11413524?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Nov 4, 2022. That report lists device problem “High Test Results” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the epoc Host2 11413524?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the epoc Host2 11413524 FDA approved?

It was cleared through 510(k) K061597 on Sep 27, 2006; its record lists 6 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the epoc Host2 11413524, and is it still sold?

The FDA's device database lists Epocal Inc as the company responsible for it (brand epoc Host2, model 11413524). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other ELECTRODE, ION SPECIFIC, CALCIUM devices

All ELECTRODE, ION SPECIFIC, CALCIUM devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page