Stelo Biosensor 9500-174Recalls & reported problems
Dexcom, Inc. · Stelo Biosensor · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Dexcom Stelo biosensor is an over-the-counter continuous glucose monitor. Most of the 290 reports describe parts of the sensor coming loose or detaching, and some describe a broken needle or cannula, sometimes with parts left in the body. The number of reports fell from 288 in the prior 12 months to 2 in the last 12 months. The reports include no deaths and 11 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 290
- Deaths reported
- 0
- Injuries reported
- 11
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 290 adverse event reports that list this exact device (DI 00386270004338), received between Apr 30, 2025 and Dec 15, 2025 — classified as 11 injuries, 279 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (214 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component214
- Break49
- Appropriate Device Problem Term/Code Not Available18
- Adverse Event Without Identified Device or Use Problem11
- Wireless Communication Problem3
Patient problems
- No Clinical Signs, Symptoms or Conditions214
- Foreign Body In Patient45
- Pain21
- Erythema17
- Bacterial Infection10
- Skin Inflammation/ Irritation10
- Abscess4
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available3
Recent adverse event reports
The 25 most recent of 290 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 15, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE NEEDLE WAS BROKEN. NO PRODUCT OR DATA WAS PROVIDED FOR INVESTIGATION. PROBLEM COULD NOT BE CONFIRMED AND PROBABLE CAUSE CANNOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23809375 ↗MalfunctionDec 15, 2025Break; Break
IT WAS REPORTED THAT A BROKEN NEEDLE OR CANNULA OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2025. NO RELEVANT PRODUCT OR DATA RELATED TO THE ISSUE WAS PROVIDED FOR INVESTIGATION. CONFIRMATION OF THE COMPLAINT IS UNDETERMINED AND PROBABLE CAUSE CANNOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23804108 ↗MalfunctionSep 19, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT AN APPLICATOR DEPLOYMENT OCCURRED. THE SENSOR WAS INSERTED INTO THE BACK OF UPPER ARM. PRODUCT WAS PROVIDED FOR INVESTIGATION. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED DUE TO NO DAMAGE. THE INTERNAL VISUAL INSPECTION WAS PERFORMED AND FAILED DUE DETACHED NEEDLE. AN APPLICATOR DEPLOYMENT WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. HOWEVER, A BROKEN OR DETACHED NEEDLE OR CANNULA WAS FOUND IN CONNECTION TO THE REPORTED EVENT. THE PROBABLE CAUSE OF THE BROKEN OR DETACHED NEEDLE OR CANNULA COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23096324 ↗MalfunctionSep 15, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A DETACHED OR MISSING SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED ON (B)(6) 2025. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 23050625 ↗MalfunctionSep 15, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A DETACHED OR MISSING SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED INTO THE ABDOMEN, WHICH IS OFF LABEL USAGE OF THE DEVICE, ON (B)(6) 2025. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23050184 ↗MalfunctionSep 10, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A DETACHED OR MISSING SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED INTO THE BACK OF UPPER ARM ON (B)(6) 2025. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 23016385 ↗MalfunctionSep 9, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A DETACHED OR MISSING SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED INTO THE BACK OF UPPER ARM ON (B)(6) 2025. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23009300 ↗MalfunctionSep 9, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A DETACHED OR MISSING SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED INTO THE ABDOMEN, WHICH IS OFF LABEL USAGE OF THE DEVICE, ON (B)(6) 2025. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23004168 ↗MalfunctionSep 9, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A DETACHED OR MISSING SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED INTO THE BACK OF UPPER ARM ON (B)(6) 2025. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23001523 ↗MalfunctionSep 8, 2025Break; Break
AFTER SUBMISSION OF THE INITIAL MDR, PRODUCT WAS RECEIVED ON 9/10/2025 AND IT WAS DETERMINED THIS COMPLAINT IS NOT REPORTABLE PER (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22999344 ↗MalfunctionSep 8, 2025Break; Break
IT WAS REPORTED THAT A BROKEN SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED INTO THE BACK OF UPPER ARM ON (B)(6) 2025. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22999337 ↗MalfunctionSep 5, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT A BROKEN OR DETACHED WIRE OCCURRED. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22986637 ↗MalfunctionSep 5, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A DETACHED OR MISSING SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED INTO THE BACK OF UPPER ARM ON (B)(6) 2025. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: Discomfort; Foreign Body In Patient
MAUDE report 22984102 ↗MalfunctionSep 5, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT A DETACHED OR MISSING SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED INTO THE BACK OF UPPER ARM ON (B)(6) 2025. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: Erythema
MAUDE report 22976399 ↗MalfunctionSep 5, 2025Break; Break
IT WAS REPORTED THAT A BROKEN SENSOR WIRE OCCURRED. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22976049 ↗MalfunctionSep 4, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT A BROKEN OR DETACHED WIRE OCCURRED. THE SENSOR WAS INSERTED ON (B)(6) 2025. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22968614 ↗MalfunctionSep 4, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A DETACHED OR MISSING SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED INTO THE BACK OF UPPER ARM ON (B)(6) 2025. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: Pain
MAUDE report 22964172 ↗MalfunctionSep 4, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A DETACHED OR MISSING SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED INTO THE BACK OF UPPER ARM ON (B)(6) 2025. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22963652 ↗MalfunctionSep 4, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A DETACHED OR MISSING SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED INTO THE BACK OF UPPER ARM ON (B)(6) 2025. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22962714 ↗MalfunctionSep 2, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A DETACHED OR MISSING SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED INTO THE BACK OF UPPER ARM ON (B)(6)2025. PRODUCT WAS PROVIDED FOR INVESTIGATION. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. AN INTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. THE ALLEGATION WAS UNDETERMINED VIA PRODUCT. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22941449 ↗MalfunctionSep 2, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A DETACHED OR MISSING SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED INTO THE BACK OF UPPER ARM ON (B)(6) 2025. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22936714 ↗MalfunctionSep 2, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT A BROKEN OR DETACHED WIRE OCCURRED. THE SENSOR WAS INSERTED INTO THE BACK OF UPPER ARM ON (B)(6) 2025. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF A BROKEN SENSOR WIRE. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22936486 ↗MalfunctionSep 2, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A DETACHED OR MISSING SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED INTO THE BACK OF UPPER ARM ON (B)(6) 2025. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. PRODUCT HAS BEEN RECEIVED BUT IS PENDING EVALUATION. A FOLLOW UP REPORT WILL BE SUBMITTED UPON COMPLETION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: Erythema; Foreign Body In Patient
MAUDE report 22936391 ↗InjuryAug 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
SUBSEQUENT TO THE INITIAL MDR, A CORRECTION IS REQUIRED. IT WAS REPORTED THAT A SKIN REACTION OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2025. DATE OF EVENT IS AN APPROXIMATION. ON 07/28/2025, IT WAS REPORTED THAT A SKIN REACTION AT THE PUNCTURE SITE OCCURRED. AFTER SENSOR INSERTION THE CUSTOMER DEVELOPED AN INFECTION AT THE INSERTION SITE. THEY STATED, ¿GOT AN INFECTION AT THE SITE OF INSERTION. HAD TO GO TO URGENT CARE AND TAKE ANTIBIOTICS.¿ THE NAME OF THE ANTIBIOTIC, DATE AND SITE OF INSERTION WAS NOT SPECIFIED. NO OTHER CUSTOMER DETAILS WERE AVAILABLE. NO DATA OR PRODUCT WAS PROVIDED FOR INVESTIGATION. THE ALLEGATION WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE.
Patient problems: Bacterial Infection
MAUDE report 22926993 ↗MalfunctionAug 29, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A DETACHED OR MISSING SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED INTO THE BACK OF UPPER ARM ON (B)(6) 2025. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 22925372 ↗
FDA clearance and classification
The FDA regulates product code SAF (integrated continuous glucose monitor for non-intensive glucose monitoring, over-the-counter) under 21 CFR 862.1355 as a Class II device. 21 CFR 862.1355 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the Stelo Biosensor 9500-174
Has the Stelo Biosensor 9500-174 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00386270004338), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Stelo Biosensor 9500-174?
The FDA's MAUDE database lists 290 adverse event reports for this device, received between Apr 30, 2025 and Dec 15, 2025. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (214), “Break” (49) and “Wireless Communication Problem” (3); the most common patient problems are “Foreign Body In Patient” (45), “Pain” (21) and “Erythema” (17). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Stelo Biosensor 9500-174?
None of the reports are classified as deaths, 11 as injuries and 279 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Stelo Biosensor 9500-174 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the Stelo Biosensor 9500-174, and is it still sold?
The FDA's device database lists Dexcom, Inc. as the company responsible for it (brand Stelo Biosensor, model 9500-174). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Integrated Continuous Glucose Monitor For Non-Intensive Glucose Monitoring, Over-The-Counter devices
- Stelo Glucose Biosensor (2pk) STP-XN-002Dexcom, Inc. · Integrated Continuous Glucose Monitor For Non-Intensive Glucose Monitoring, Over-The-Counter610 reports
- Stelo Glucose Biosensor (1pk) STP-XN-001Dexcom, Inc. · Integrated Continuous Glucose Monitor For Non-Intensive Glucose Monitoring, Over-The-Counter537 reports
- Lingo 77263Abbott Diabetes Care Inc · Integrated continuous glucose monitor for non-intensive glucose monitoring, Over-The-Counter35 reports
- Stelo Glucose Biosensor (2pk) STP-XN-003Dexcom, Inc. · Integrated Continuous Glucose Monitor For Non-Intensive Glucose Monitoring, Over-The-Counter4 reports
- Dexcom Stelo iOS Mobile App SW14116Dexcom, Inc. · Integrated Continuous Glucose Monitor For Non-Intensive Glucose Monitoring, Over-The-Counter3 reports
- Stelo Glucose Biosensor (2pk) UPC STP-XN-002Dexcom, Inc. · Integrated Continuous Glucose Monitor For Non-Intensive Glucose Monitoring, Over-The-Counter3 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page