Stelo Glucose Biosensor (1pk) STP-XN-001Recalls & reported problems

Dexcom, Inc. · Stelo Glucose Biosensor US (1pk) · Class 2

Quick summary

This Stelo glucose biosensor from Dexcom is an over-the-counter continuous glucose sensor worn on the back of the upper arm. The FDA database holds 540 reports for it, many describing sensor parts detaching or breaking, and some mentioning pieces left in the body. There were 38 injuries and no deaths. No recalls name this device. Reports rose from 96 in the previous 12 months to 444 in the last 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
551
Deaths reported
0
Injuries reported
39
Recalls naming it
0

Summary

The FDA's MAUDE database contains 551 adverse event reports that list this exact device (DI 00386270004345), received between May 1, 2025 and Sep 30, 2026 — classified as 39 injuries, 512 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (295 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 8 reportsJun 2025: 4 reportsJul 2025: 3 reportsAug 2025: 2 reportsSep 2025: 29 reportsOct 2025: 50 reportsNov 2025: 39 reportsDec 2025: 57 reportsJan 2026: 44 reportsFeb 2026: 50 reportsMar 2026: 51 reportsApr 2026: 69 reportsMay 2026: 52 reportsJun 2026: 39 reportsJul 2026: 10 reportsAug 2026: 29 reportsSep 2026: 15 reportsOct 2026: 0 reports20242025202669
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component295
  • Appropriate Device Problem Term/Code Not Available125
  • Break93
  • Adverse Event Without Identified Device or Use Problem24
  • Wireless Communication Problem12
  • Unintended Electrical Shock4
  • Incorrect, Inadequate or Imprecise Result or Readings3
  • Application Program Freezes or Fails to Launch1

Patient problems

  • No Clinical Signs, Symptoms or Conditions390
  • Foreign Body In Patient103
  • Pain38
  • Erythema37
  • Bacterial Infection18
  • Skin Inflammation/ Irritation16
  • Hemorrhage/Blood Loss/Bleeding15
  • Insufficient Information10

Recent adverse event reports

The 25 most recent of 551 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026Appropriate Device Problem Term/Code Not Available

    It was reported that a wearable failure occurred. The biosensor was inserted into the back of upper arm on (b)(6) 2026. On (b)(6) 2026 the wearable failed. The customer replied ¿yes¿ to the medical intervention question, and indicated they had unspecified hypoglycemia symptoms. The consumer described the medical intervention as ¿hospital.¿ At the time of the report on (b)(6) 2026 they were recovering. Although requested, no other details were provided, and they did not respond to additional requests for information. Data has been requested but is pending evaluation. A follow up report will be submitted upon completion. No additional event or patient information is available.

    Patient problems: Hypoglycemia; Insufficient Information

    MAUDE report 50947727 ↗
  • MalfunctionSep 30, 2026Detachment of Device or Device Component

    It was reported that a detached or missing sensor wire occurred. The sensor was inserted on (b)(6) 2026. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947348 ↗
  • MalfunctionSep 29, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50939726 ↗
  • MalfunctionSep 29, 2026Detachment of Device or Device Component

    It was reported that a detached or missing sensor wire occurred. The sensor was inserted on (b)(6)2026. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50936222 ↗
  • MalfunctionSep 28, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923347 ↗
  • MalfunctionSep 28, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923094 ↗
  • MalfunctionSep 28, 2026Detachment of Device or Device Component

    It was reported that a detached or missing sensor wire occurred. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50916822 ↗
  • MalfunctionSep 28, 2026Detachment of Device or Device Component

    It was reported that a detached or missing sensor wire occurred. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50916812 ↗
  • MalfunctionSep 28, 2026Detachment of Device or Device Component

    It was reported that a detached or missing sensor wire occurred. The sensor was inserted into the back of the upper arm on (b)(6)-2026. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50915800 ↗
  • MalfunctionSep 11, 2026Wireless Communication Problem

    It was reported that "Signal Loss" occurred. No product or data was provided for investigation. The allegation and the probable cause cannot be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50735344 ↗
  • MalfunctionSep 1, 2026Detachment of Device or Device Component

    It was reported that a detached or missing sensor wire occurred. The sensor was inserted on (b)(6)2026. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50554613 ↗
  • MalfunctionSep 1, 2026Appropriate Device Problem Term/Code Not Available

    It was reported that a broken or detached wire occurred. The sensor was inserted on (b)(6) 2026. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50554436 ↗
  • MalfunctionSep 1, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50552529 ↗
  • MalfunctionSep 1, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50549337 ↗
  • MalfunctionSep 1, 2026Detachment of Device or Device Component

    It was reported that a detached or missing sensor wire occurred. The sensor was inserted into the back of the upper arm on (b)(6) 2026. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: Erythema

    MAUDE report 50548662 ↗
  • MalfunctionAug 31, 2026Detachment of Device or Device Component

    It was reported that a detached or missing sensor wire occurred. The sensor was inserted into the back of the upper arm on (b)(6) 2026. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50524561 ↗
  • MalfunctionAug 31, 2026Appropriate Device Problem Term/Code Not Available

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50523756 ↗
  • MalfunctionAug 29, 2026Unintended Electrical Shock

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50509984 ↗
  • MalfunctionAug 28, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50506637 ↗
  • MalfunctionAug 28, 2026Appropriate Device Problem Term/Code Not Available

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50503518 ↗
  • MalfunctionAug 27, 2026Appropriate Device Problem Term/Code Not Available

    It was reported that a broken or detached wire occurred. The sensor was inserted on (b)(6) 2026. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50487971 ↗
  • MalfunctionAug 27, 2026Appropriate Device Problem Term/Code Not Available

    It was reported that a broken or detached wire occurred. The sensor was inserted on (b)(6) 2026. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50487344 ↗
  • MalfunctionAug 27, 2026Break

    It was reported that a broken sensor wire occurred. The sensor was inserted into the back of the upper arm on (b)(6) 2026. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50483051 ↗
  • MalfunctionAug 27, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50465122 ↗
  • MalfunctionAug 26, 2026Appropriate Device Problem Term/Code Not Available

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50463592 ↗

Other Integrated Continuous Glucose Monitor For Non-Intensive Glucose Monitoring, Over-The-Counter devices

All Integrated Continuous Glucose Monitor For Non-Intensive Glucose Monitoring, Over-The-Counter devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.