TR BAND TRB24-REG (UDI 00389701011349)Recalls & reported problems
Terumo Medical Corporation · TR BAND Radial Artery Compression Device · Class 2
All models of TR BAND (4)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Terumo TR Band is a device used to compress the radial artery at the wrist after a procedure. The 284 reports often describe the band losing air and deflating, and some describe hematomas and blood loss at the access site. The reports include no deaths and 86 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 284
- Deaths reported
- 0
- Injuries reported
- 86
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 284 adverse event reports that list this exact device (DI 00389701011349), received between Apr 25, 2017 and May 11, 2026 — classified as 86 injuries, 198 malfunctions.
The most frequently reported specific device problem is “Gas/Air Leak” (134 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Gas/Air Leak134
- Deflation Problem86
- Unintended Deflation34
- Leak/Splash21
- Inflation Problem16
- Adverse Event Without Identified Device or Use Problem8
- Air Leak7
- Break5
Patient problems
- No Clinical Signs, Symptoms or Conditions145
- Hematoma67
- Blood Loss27
- Hemorrhage/Blood Loss/Bleeding22
- No Consequences Or Impact To Patient19
- Bruise/Contusion8
- No Information5
- No Patient Involvement5
Recent adverse event reports
The 25 most recent of 284 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 11, 2026Gas/Air Leak; Unintended Deflation
TERUMO RECEIVED THE FOLLOWING REPORTED INFORMATION: THE TR BAND WAS APPLIED PER PROTOCOL FOLLOWING A CORONARY ANGIOGRAM; HOWEVER, LESS THAN ONE MINUTE LATER, THE BAND DID NOT HOLD AIR/DEFLATED. THE BAND WAS REMOVED AND A NEW BAND WAS PLACED WITHOUT INCIDENT. THE PROCEDURE WAS INVASIVE, AND THERE WAS UNEXPECTED OR PROLONGED CARE. A 6 FRENCH RADIAL SHEATH WAS USED AT THE ACCESS SITE DURING THE PROCEDURE. THERE WAS NO HARM TO THE PATIENT. THERE WAS NO NOTICEABLE DAMAGE TO THE DEVICE. NO BLOOD WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25137491 ↗MalfunctionMay 5, 2026Gas/Air Leak; Unintended Deflation
TERUMO RECEIVED THE FOLLOWING REPORTED INFORMATION: FOLLOWING A LEFT HEART CATHETERIZATION (LHC), THE BAND WAS APPLIED AND 12ML AIR WAS INJECTED INTO THE TR BAND. IMMEDIATELY AFTER THE PROCEDURE, THE LARGE AND SMALL BALLOON ASSEMBLY BEGAN TO LOSE AIR AND DEFLATED, RESULTING IN A NEW BAND BEING PLACED. HEMOSTASIS WAS ACHIEVED, THE PATIENT WAS STABLE, AND NO BLOOD LOSS WAS REPORTED. THERE WAS NO NOTICEABLE DAMAGE TO THE DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25083104 ↗MalfunctionApr 23, 2026Deflation Problem; Gas/Air Leak
TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING INFORMATION: IT WAS REPORTED THAT AFTER RECEIVING A RADIAL ACCESS HEART CATHETERIZATION THE FEMALE PATIENT HAD A TR BAND PLACED ON THE WRIST. AS THE PATIENT WAS BEING TRANSPORTED FROM THE ROOM STAFF NOTICED THAT THERE WAS BLOOD AROUND THE ACCESS SITE, THE BALLOON DEFLATED BEFORE PATIENT LEFT THE LAB. THE TR BAND HAD LOST AIR. THE DOCTOR REMOVED THE TR BAND AND HELD PRESSURE UNTIL A NEW BAND WAS PLACED ON THE WRIST. THERE WERE NO ISSUES WITH THE NEW BAND. THERE WAS NO RECORDED BLOOD LOSS NOR REPORTED FURTHER PATIENT INJURY. THE NURSE STATED THAT THEY TESTED THE TR BAND AFTER REMOVAL AND FOUND THAT IT WAS LEAKING AIR. THE LEAK WAS COMING FROM THE BALLOON. THE PATIENT WAS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24975304 ↗MalfunctionApr 17, 2026Deflation Problem; Gas/Air Leak
TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING INFORMATION: IT WAS REPORTED THAT ORIGINAL A TR BAND WAS APPLIED POST RADIAL PROCEDURE. A HEMATOMA DEVELOPED PROXIMAL TO THE BAND AFTER A COUPLE OF MINUTES. MANUAL PRESSURE WAS APPLIED AND THEN THE TR BAND WAS REPOSITIONED. IT WAS NOTED AT THAT POINT THAT THE TR BAND APPEARED TO BE LOSING AIR. MORE AIR WAS ADDED, HOWEVER, AGAIN THERE WAS A SLOW LEAK. A NEW TR BAND WAS PLACED ON THE PATIENT AND THERE WERE NO FURTHER COMPLICATIONS. HEMOSTASIS WAS ACHIEVED USING THE SECOND TR BAND THAT WAS APPLIED. THE FIRST TR BAND INITIALLY HAD 15ML OF AIR, HOWEVER, DURING REPOSITIONING, IT WAS NOTICED THAT ONLY 12ML WAS ABLE TO BE REMOVED. 15ML WAS ADDED FOR THE REPOSITIONING AND THEN WHEN AIR WAS LEAKING AGAIN, A NEW TR BAND WAS APPLIED WITH 11ML OF FINAL AIR VOLUME (WHICH HELD JUST FINE). THE TR BAND STARTED TO DEFLATE WITHIN A FEW MINUTES, ONCE IT WAS DETERMINED THAT THEY WANTED TO REPOSITION THE BAND. THEY SUSPECTED THE LEAK WAS COMING FROM THE "PILLOW" HOWEVER, THEY WERE UNABLE TO CONFIRM EXACTLY WHERE. THE SIZE OF THE HEMATOMA WAS ABOUT THE SIZE OF A HALF DOLLAR COIN. THE PATIENT REMAINED IN STABLE CONDITION. THE SECOND TR BAND THAT WAS APPLIE
Patient problems: Hematoma
MAUDE report 24917532 ↗MalfunctionApr 13, 2026Gas/Air Leak; Unintended Deflation
TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING INFORMATION: IT WAS REPORTED THAT THE TR BAND DID NOT HOLD AIR, THEY FINISHED WITH A NEW TR BAND. APPROXIMATELY 15ML OF AIR WAS INJECTED WHEN THE TR BAND WAS APPLIED, HOWEVER IT WOULD NOT MAINTAIN AIR AT THE BEGINNING OF THE PROCEDURE. A GLIDESLENDER SHEATH WAS USED AT THE ACCESS SITE. THE ESTIMATED BLOOD LOSS WAS LESS THAN 250CC. THE PATIENT REMAINED STABLE AND THERE WAS NO PATIENT HARM. THE PROCEDURE PERFORMED PRIOR TO THE USE OF THE TR BAND WAS A PERCUTANEOUS CORONARY INTERVENTION (PCI). THERE WAS NOT ANY NOTICEABLE DAMAGE TO THE DEVICE. IT IS UNKNOWN HOW HEMOSTASIS WAS ACHIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24865738 ↗MalfunctionApr 11, 2026Gas/Air Leak; Unintended Deflation
TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING INFORMATION: IT WAS REPORTED THAT THE TR BAND WOULD NOT HOLD AIR PRESSURE. THERE WAS 18ML OF AIR INJECTED INTO THE TR BAND WHEN IT WAS APPLIED. THE TR BAND STARTED TO DEFLATE AFTER INITIAL INFLATION. THERE WAS NO NOTICEABLE DAMAGE TO THE DEVICE. THE LOCATION OF THE LEAK IS UNKNOWN. THE PATIENT REMAINED IN STABLE CONDITION AND THERE WAS NO PATIENT HARM. HEMOSTASIS WAS ACHIEVED BY A SECOND TR BAND. THE PROCEDURE PERFORMED PRIOR TO THE USE OF THE TR BAND WAS LEFT HEART CATHETERIZATION. THE ESTIMATED BLOOD LOSS WAS LESS THAN 250CC.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24850671 ↗MalfunctionApr 8, 2026Deflation Problem; Gas/Air Leak
TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING REPORTED INFORMATION: FOLLOWING A CARDIAC CATH, THE TR BAND WAS PLACED AND 15ML'S OF AIR WAS INJECTED INTO THE BAND. THEN 5 WAS THEN REPLACED. TWO MINUTES, THE TR BAND STARTED TO DEFLATE/LOST PRESSURE IN THE BALLOON AREA. THERE WAS NO NOTICEABLE DAMAGE TO THE DEVICE. THE BLOOD LOSS WAS LESS THAN 250CC, AND THE PATIENT WAS STABLE. THE PATIENT DID BLEED AND DEVELOPED A HEMATOMA; HOWEVER, WAS STABLE AND OKAY AFTERWARDS. THE SIZE OF THE HEMATOMA WAS UNKNOWN. A SECOND TR BAND WAS PLACED ON TO PROVIDE PRESSURE TO THE HEMATOMA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24816277 ↗MalfunctionApr 2, 2026Gas/Air Leak; Unintended Deflation
TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING INFORMATION: IT WAS REPORTED THAT UPON PLACING THE TR BAND AND FILLING WITH AIR, IT IMMEDIATELY BEGAN TO SELF-DEFLATE. A NEW TR BAND WAS PLACED. THERE WAS NO PATIENT INJURY REPORTED AND NO BLOOD LOSS. THE PROCEDURE PERFORMED PRIOR TO THE USE OF THE TR BAND WAS LEFT HEART CATHETERIZATION (LHC).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24772670 ↗MalfunctionApr 2, 2026Gas/Air Leak; Unintended Deflation
TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING INFORMATION: IT WAS REPORTED THAT WHEN PUTTING AIR INTO THE TR BAND, IT WAS NOT HOLDING AIR AND SLOWLY DEFLATED. A NEW TR BAND WAS PLACED AND HEMOSTASIS WAS ACHIEVED. THE PATIENT REMAINED IN STABLE CONDITION. THERE WAS THE NORMAL AMOUNT OF AIR INJECTED UNTIL BLEEDING STOPPED. THERE WAS NO VISIBLE DAMAGE NOTED ON THE TR BAND. THE PROCEDURE PERFORMED FOR THE PATIENT WAS A PERCUTANEOUS CORONARY INTERVENTION (PCI). THE ESTIMATED BLOOD LOSS WAS LESS THAN 250CC.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24772667 ↗MalfunctionApr 2, 2026Gas/Air Leak; Unintended Deflation
TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING REPORTED INFORMATION: FOLLOWING A RIGHT HEART CATH, THE TR BAND WAS APPLIED AND 13ML'S (11CC) OF AIR WAS INJECTED INTO THE BAND. FORTY-FIVE (45) MINUTES LATER, THE PATIENT WAS MOVED FROM THE OPERATING ROOM (OR) TO THE HOLDING AREA. WHEN THE NURSE CHECKED THE TR BAND, THEY NOTICED THE AIR WAS DISSOLVED. A SLOW LEAK OVER TIME OCCURRED. THE BALLOON WITH 5CC OF AIR LEAKED OVER A TWO (2) HOUR PERIOD. THERE WAS NO NOTICEABLE DAMAGE TO THE DEVICE. THE DEVICE WAS REMOVED AND REPLACED WITH NEW TR BAND AND HEMOSTASIS WAS ACHIEVED. THERE WAS MINIMAL BLOOD LOSS, LESS THAN 5CC, AND THE PATIENT WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24766348 ↗MalfunctionMar 31, 2026Gas/Air Leak
TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING REPORTED INFORMATION: FOLLOWING A LEFT HEART CATH, THE BALLOON OF THE TR BAND LOST AIR IMMEDIATELY AFTER PLACEMENT, SLOW LEAK AT 5CC/HR. THERE WAS NO NOTICEABLE DAMAGE TO THE DEVICE. HEMOSTASIS WAS ACHIEVED BY USING A SECOND TR BAND, AND THE PATIENT WAS IN STABLE CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24743505 ↗MalfunctionMar 31, 2026Gas/Air Leak; Unintended Deflation
TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING INFORMATION: IT WAS REPORTED THAT THE TR BAND WAS APPLIED. IT WAS NOTICED AFTER BALLOON INFLATION THAT THE BALLOONS DEFLATED IMMEDIATELY AFTER REMOVING THE SYRINGE AND BLEEDING WAS NOTICED. MULTIPLE ATTEMPTS TO INFLATE THE TR BAND DID NOT WORK. THE TR BAND WAS REPLACED WITH NEW TR BAND AND RECOVERY PROCEEDED WITHOUT ISSUE. THERE WAS NO NOTICEABLE DAMAGE. THE PATIENT REMAINED IN STABLE CONDITION, WITH MINIMAL BLEEDING AND NO HEMATOMA. THE LOCATION OF THE LEAK AND AMOUNT OF AIR THAT WAS INITIALLY APPLIED ARE BOTH UNKNOWN. THE PROCEDURE PERFORMED PRIOR TO THE USE OF THE TR BAND WAS CARDIAC CATHETERIZATION AND PERCUTANEOUS CORONARY INTERVENTION (PCI).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24741703 ↗MalfunctionMar 30, 2026Unintended Deflation
TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING REPORTED INFORMATION: FOLLOWING A LEFT HEART CATH, THE TR BAND WAS APPLIED. HOWEVER, A DEFLATION ISSUE WITH THE BALLOON OCCURRED AND LED TO A HEMATOMA. THE ESTIMATED BLOOD LOSS WAS LESS THAN 250CC.
Patient problems: Hematoma
MAUDE report 24731576 ↗MalfunctionMar 30, 2026Unintended Deflation
TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING REPORTED INFORMATION: FOLLOWING A LEFT HEART CATH, THE TR BAND WAS APPLIED. HOWEVER, A DEFLATION ISSUE WITH THE BALLOON OCCURRED AND LED TO A HEMATOMA. THE ESTIMATED BLOOD LOSS WAS LESS THAN 250CC.
Patient problems: Hematoma
MAUDE report 24730491 ↗MalfunctionMar 30, 2026Unintended Deflation
TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING REPORTED INFORMATION: FOLLOWING A LEFT HEART CATH, THE TR BAND WAS APPLIED. HOWEVER, A DEFLATION ISSUE WITH THE BALLOON OCCURRED AND LED TO A HEMATOMA. THE ESTIMATED BLOOD LOSS WAS LESS THAN 250CC.
Patient problems: Hematoma
MAUDE report 24730408 ↗MalfunctionMar 30, 2026Unintended Deflation
TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING REPORTED INFORMATION: FOLLOWING A LEFT HEART CATH, THE TR BAND WAS APPLIED. HOWEVER, A DEFLATION ISSUE WITH THE BALLOON OCCURRED AND LED TO A HEMATOMA. THE ESTIMATED BLOOD LOSS WAS LESS THAN 250CC.
Patient problems: Hematoma
MAUDE report 24730357 ↗MalfunctionMar 30, 2026Gas/Air Leak
TERUMO MEDICAL RECEIVED AN FDA MEDWATCH REPORT # (B)(4). THE EVENT DESCRIPTION STATES: "APPLYING TR BAND ON A PATIENT POST CARDIAC CATH, THE PHYSICIAN ATTEMPTED TO INFLATE THE DEVICE AND IT FAILED TO HOLD THE AIR. DEVICE REMOVED FROM PATIENT AND ANOTHER ONE WAS APPLIED." ADDITIONAL INFORMATION WAS RECEIVED ON 11MAR2026: THE TR BAND WAS REMOVED FROM THE PATIENT AND ANOTHER ONE WAS APPLIED. THE SHEATH REMAINED IN PLACE UNTIL ANOTHER TR BAND WAS APPLIED SUCCESSFULLY. THE FAILURE WAS SEEN IN THE CATH LAB AND WAS NOTED DURING INFLATION. THE DEVICE WAS NOT MANIPULATED BEFORE USE IN ANY WAY, PRE-USE OR DURING STORAGE. THE PRODUCT WAS STORED PRIOR TO USE IN THE CABINET. THE PATIENT REMAINED IN STABLE CONDITION AND THERE WAS NO PATIENT HARM. THERE WAS NO BLOOD LOSS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24725531 ↗MalfunctionMar 24, 2026Gas/Air Leak; Unintended Deflation
TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING REPORTED INFORMATION: THE NURSE TESTED THE BAND BY PUTTING 15ML OF AIR ON IT AND PRESSING ON THE BALLOON, DEFLATED THE BAND, AND THEN APPLIED TO THE PATIENT. UPON INFLATION AND APPLICATION OF BAND, THE BAND DID NOT MAINTAIN AIR AND THE PATIENT BEGAN TO BLEED. THE BAND WAS TAKEN OFF THE PATIENT AND OPENED A NEW TR BAND WAS USED WITHOUT ISSUE. THE BLOOD LOSS WAS LESS THAN 250CC.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24675350 ↗MalfunctionMar 19, 2026Gas/Air Leak; Unintended Deflation
TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING REPORTED INFORMATION: FOLLOWING A LEFT HEART CATHETERIZATION WHILE THE PATIENT WAS IN THE ROOM, THE BLADDER OF THE TR BAND DID NOT HOLD/LOST AIR. THE BLADDER WAS NOTED TO BE FLATTER UPON REMOVAL. A TERUMO 5/6 FRENCH SLENDER SHEATH WAS USED IN THE ACCESS SITE DURING THE PROCEDURE. THE TR BAND STARTED TO LOSE AIR APPROXIMATELY 5 MINUTES AFTER INITIAL INFLATION. 15 MLS OF AIR WAS INITIALLY INJECTED INTO THE TR BAND. MANUAL PRESSURE WAS HELD UNTIL A SECOND BAND WAS APPLIED; HOWEVER, THERE WAS ACTIVE BLEEDING WHEN APPLYING THE NEW BAND. HEMOSTASIS WAS ACHIEVED. THE ESTIMATED BLOOD LOSS WAS ABOUT 5CC. THE PATIENT WAS IN STABLE CONDITION AND RELEASED AS PLANNED. NO SWELLING, PAIN, NUMBNESS, FAINT PULSE, OR WEAKNESS WAS NOTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24642867 ↗MalfunctionMar 19, 2026Gas/Air Leak; Unintended Deflation
TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING REPORTED INFORMATION: FOLLOWING A LEFT HEART CATHETERIZATION WHILE THE PATIENT WAS IN THE ROOM, THE BLADDER OF THE TR BAND DID NOT HOLD/LOST AIR. THE BLADDER WAS NOTED TO BE FLATTER UPON REMOVAL. A TERUMO 5/6 FRENCH SLENDER SHEATH WAS USED IN THE ACCESS SITE DURING THE PROCEDURE. THE TR BAND STARTED TO LOSE AIR APPROXIMATELY 5-10 MINUTES AFTER INITIAL INFLATION. IT WAS UNKNOWN HOW MANY ML'S WERE INITIALLY INJECTED. MANUAL PRESSURE WAS HELD UNTIL A SECOND BAND WAS APPLIED; HOWEVER, THERE WAS ACTIVE BLEEDING WHEN APPLYING THE NEW BAND. HEMOSTASIS WAS ACHIEVED. THE ESTIMATED BLOOD LOSS WAS ABOUT 5-10CC. THE PATIENT WAS IN STABLE CONDITION AND RELEASED AS PLANNED. NO SWELLING, PAIN, NUMBNESS, FAINT PULSE, OR WEAKNESS WAS NOTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24642866 ↗MalfunctionMar 19, 2026Gas/Air Leak; Unintended Deflation
TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING REPORTED INFORMATION: FOLLOWING A LEFT HEART CATHETERIZATION WHILE THE PATIENT WAS IN THE ROOM, THE BLADDER OF THE TR BAND DID NOT HOLD/LOST AIR. THE BLADDER WAS NOTED TO BE FLATTER UPON REMOVAL. A TERUMO 5/6 FRENCH SLENDER SHEATH WAS USED IN THE ACCESS SITE DURING THE PROCEDURE. THE TR BAND STARTED TO LOSE AIR APPROXIMATELY 5 MINUTES AFTER INITIAL INFLATION. 15 MLS OF AIR WAS INITIALLY INJECTED INTO THE TR BAND. MANUAL PRESSURE WAS HELD UNTIL A SECOND BAND WAS APPLIED; HOWEVER, THERE WAS ACTIVE BLEEDING WHEN APPLYING THE NEW BAND. HEMOSTASIS WAS ACHIEVED. THE ESTIMATED BLOOD LOSS WAS ABOUT 5CC. THE PATIENT WAS IN STABLE CONDITION AND RELEASED AS PLANNED. NO SWELLING, PAIN, NUMBNESS, FAINT PULSE, OR WEAKNESS WAS NOTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24642865 ↗MalfunctionMar 19, 2026Gas/Air Leak; Unintended Deflation
TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING REPORTED INFORMATION: FOLLOWING A LEFT HEART CATHETERIZATION WHILE THE PATIENT WAS IN THE ROOM, THE BLADDER OF THE TR BAND DID NOT HOLD/LOST AIR. THE BLADDER WAS NOTED TO BE FLATTER UPON REMOVAL. A TERUMO 5/6 FRENCH SLENDER SHEATH WAS USED IN THE ACCESS SITE DURING THE PROCEDURE. THE TR BAND STARTED TO LOSE AIR APPROXIMATELY 5 MINUTES AFTER INITIAL INFLATION. 15 MLS OF AIR WAS INITIALLY INJECTED INTO THE TR BAND. MANUAL PRESSURE WAS HELD UNTIL A SECOND BAND WAS APPLIED; HOWEVER, THERE WAS ACTIVE BLEEDING WHEN APPLYING THE NEW BAND. HEMOSTASIS WAS ACHIEVED. THE ESTIMATED BLOOD LOSS WAS ABOUT 5CC. THE PATIENT WAS IN STABLE CONDITION AND RELEASED AS PLANNED. NO SWELLING, PAIN, NUMBNESS, FAINT PULSE, OR WEAKNESS WAS NOTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24642864 ↗MalfunctionMar 2, 2026
APPLYING TR BAND ON A PATIENT POST CARDIAC CATH, THE PHYSICIAN ATTEMPTED TO INFLATE THE DEVICE AND IT FAILED TO HOLD THE AIR. DEVICE REMOVED FROM PATIENT AND ANOTHER ONE WAS APPLIED. MANUFACTURER RESPONSE FOR CLAMP, VASCULAR, TR BAND (PER SITE REPORTER). CLINICAL REPRESENTATIVE FROM TERUMO WILL BE INTO EVALUATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24477533 ↗MalfunctionJan 29, 2026Deflation Problem
TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING REPORTED INFORMATION: PERCUTANEOUS CORONARY INTERVENTION VIA RADIAL ACCESS USING A 6 FRENCH GLIDESHEATH SLENDER SHEATH. THE PATIENT BEGAN TO DEVELOP HEMATOMA FOLLOWING CONCLUSION OF PROCEDURE AND PLACEMENT OF THE TR BAND. THE TR BAND WAS PLACED PER PROTOCOL AND INSTRUCTION FOR USE (IFU). THE BALLOON DEFLATED PREMATURELY AND BLOOD APPEARED. THE TR BAND WAS RE-INFLATED FOR HEMOSTASIS. THE PATIENT SHOWED HEMATOMA SYMPTOMS UPON REACHING POST-PROCEDURAL ROOM & CARE. PROCEDURE WAS DEEMED SUCCESSFUL. THERE WAS NO BLOOD LOSS.
Patient problems: Hematoma
MAUDE report 24217879 ↗MalfunctionJan 15, 2026Gas/Air Leak
TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING REPORTED INFORMATION: TR BAND WAS APPLIED BY THE DOCTOR AND AN EXPERIENCED TECH. IT SEEMED OK AT FIRST, HOWEVER, THEN THEY NOTICED A HEMATOMA AND THAT THE TR BAND HAD DEFLATED. MANUAL PRESSURE AND NEW TR BAND WAS APPLIED. ADDITIONAL INFORMATION WAS RECEIVED ON 23DEC2025: THERE WAS 15ML OF AIR INJECTED INTO THE TR BAND WHEN IT WAS APPLIED. A 5FR SLENDER WAS USED IN THE ACCESS SITE DURING THE PATIENT PROCEDURE. THE TR BAND DEFLATED INSTANTLY, IT HAD A HOLE IN IT. THE HEMATOMA WAS TREATED BY MANUAL PRESSURE. THE SIZE OF THE HEMATOMA WAS TWO INCHES. THERE WAS SWELLING INITIALLY. THERE WAS NO PAIN OR NUMBNESS. THE PULSE WAS NOT FAINT AND NO WEAKNESS IN MOVEMENT. THERE WAS NO NOTICEABLE DAMAGE TO THE DEVICE. THE PATIENT WAS STABLE.
Patient problems: Hematoma; Swelling/ Edema
MAUDE report 24090315 ↗
FDA clearance and classification
Cleared through 510(k) K152525 on Nov 10, 2015 (decision: Substantially Equivalent). Submitted by Terumo Medical Corporation as “TR BAND Radial Compression Device”. The FDA review took 68 days. FDA record ↗
The FDA regulates product code DXC (clamp, vascular) under 21 CFR 870.4450 as a Class II device. 21 CFR 870.4450 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the TR BAND TRB24-REG
Has the TR BAND TRB24-REG been recalled?
No FDA recall record in our data names this exact device identifier (DI 00389701011349), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the TR BAND TRB24-REG?
The FDA's MAUDE database lists 284 adverse event reports for this device, received between Apr 25, 2017 and May 11, 2026. In those reports, the most common specific device problems are “Gas/Air Leak” (134), “Deflation Problem” (86) and “Unintended Deflation” (34); the most common patient problems are “Hematoma” (67), “Blood Loss” (27) and “Hemorrhage/Blood Loss/Bleeding” (22). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the TR BAND TRB24-REG?
None of the reports are classified as deaths, 86 as injuries and 198 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the TR BAND TRB24-REG FDA approved?
It was cleared through 510(k) K152525 on Nov 10, 2015. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the TR BAND TRB24-REG, and is it still sold?
The FDA's device database lists Terumo Medical Corporation as the company responsible for it (brand TR BAND, model TRB24-REG). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CLAMP, VASCULAR devices
- Heartstring III System 3.8MM HSK-3038Maquet Cardiovascular LLC · Clamp, Vascular1 recall4 deaths814 reports
- Heartstring III System 4.3MM HSK-3043Maquet Cardiovascular LLC · Clamp, Vascular1 recall265 reports
- TR BAND TRB24-REGTerumo Medical Corporation · CLAMP, VASCULAR255 reports
- HeartString III Proximal Seal HS-3045Maquet Cardiovascular LLC · Clamp, Vascular1 recall187 reports
- TR BAND TRB29-LRGTerumo Medical Corporation · CLAMP, VASCULAR67 reports
- TR BAND TRB29-LRGTerumo Medical Corporation · CLAMP, VASCULAR63 reports
- Flo-Thru Intraluminal Shunt® FT-12125Synovis Life Technologies, Inc. · CLAMP, VASCULAR1 recall1 report
- Flo-Thru Intraluminal Shunt® FT-12225Synovis Life Technologies, Inc. · CLAMP, VASCULAR1 recall1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page