HeartString III Proximal Seal HS-3045Recalls & reported problems

Maquet Cardiovascular LLC · HeartString III Proximal Seal · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Maquet HeartString III seal is a clamp used in heart bypass surgery to close an opening in the aorta. Many of its 187 reports describe seals that would not load, fit, or unfold properly, and some describe bleeding that continued. There are 29 injury reports and no deaths. There is 1 open recall, from August 2025, over seals that may not load, deploy, or stop bleeding.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
187
Deaths reported
0
Injuries reported
29
Recalls naming it
1

Summary

The FDA's MAUDE database contains 187 adverse event reports that list this exact device (DI 00607567700307), received between Jul 14, 2020 and May 7, 2025 — classified as 29 injuries, 158 malfunctions.

The most frequently reported specific device problem is “Fitting Problem” (120 reports).

1 FDA recall record name this device's identifier, 1 still open.

2 other recalls cover devices cleared under the same FDA submission (K130382); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 3 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 15, 2025Class IIOpen, Classified

    Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.

    Z-2586-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fitting Problem120
  • Activation Problem38
  • Failure to Unfold or Unwrap24
  • Insufficient Device Problem Information6
  • Break2
  • Peeled/Delaminated2
  • Activation Failure1
  • Activation, Positioning or Separation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions122
  • No Consequences Or Impact To Patient59
  • Hemorrhage/Blood Loss/Bleeding13
  • Blood Loss1
  • Foreign Body In Patient1

Recent adverse event reports

The 25 most recent of 187 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 7, 2025Peeled/Delaminated; Peeled/Delaminated

    THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, VASOVIEW HEMOPRO 2 JAW PEELED OFF DESPITE NOT BEING USED FOR A LONG TIME. IT DID NOT ENTER THE PATIENT'S BODY. THE PROCEDURE WAS COMPLETED USING A NEW DEVICE. THERE WAS NO SIGNIFICANT DELAY. NO HEALTH HAZARDS TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21981649 ↗
  • InjuryMay 2, 2025Peeled/Delaminated; Failure to Cut; Peeled/Delaminated; Failure to Cut

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21950114 ↗
  • MalfunctionMay 1, 2025Melted; Melted

    N/A

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21943191 ↗
  • InjuryApr 4, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    RELATED TO (B)(4). THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, HST III SEAL (4.5MM) WERE LOADED APPROPRIATELY BUT WERE NOT HEMOSTATIC AT ALL AND BLED MUCH MORE THAN USUAL. THE SEAL WAS REMOVED AND ANOTHER LOADED HS WAS INSERTED. THE SAME SCENARIO OCCURRED AND THE SECOND HS WAS REMOVED. A THIRD WAS INSERTED, AGAIN WITH THE SAME OUTCOME AND A FOURTH WAS INSERTED. THE SURGEON REMOVED THE FOURTH SINCE THERE WAS STILL THE EXCESSIVE BLEEDING AND PLACED A SIDE-BITING CLAMP ON THE AORTA TO COMPLETE THE SINGLE ANASTOMOSIS. ANASTOMOSIS WAS COMPLETED WITHOUT FURTHER INCIDENT. NO PATIENT INJURY.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21778525 ↗
  • MalfunctionAug 1, 2023Fitting Problem; Fitting Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17432495 ↗
  • MalfunctionJul 31, 2023Fitting Problem; Fitting Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17431148 ↗
  • InjuryJul 10, 2023Activation Problem; Activation Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17292930 ↗
  • MalfunctionJul 10, 2023Fitting Problem; Fitting Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17290282 ↗
  • MalfunctionJun 23, 2023Activation Problem; Activation Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17192322 ↗
  • MalfunctionApr 10, 2023Activation Problem; Activation Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16709764 ↗
  • MalfunctionFeb 9, 2023Activation Problem; Activation Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16336726 ↗
  • MalfunctionOct 12, 2022Fitting Problem; Fitting Problem

    THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE THE HST III SEAL (4.5MM), 5-PACK SEAL DEVELOPMENT DID NOT GO WELL. THE SEAL COULD NOT BE LOADED INTO THE DELIVERY TUBE AND REMAINED IN THE LOADER AND COULD NOT BE DEPLOYED. THE WHITE PLUNGER IS NOT PRESSED BECAUSE THE SEAL COULD NOT BE DEPLOYED WHEN LOADING IT INTO THE DELIVERY DEVICE. THE BLUE SLIDE IS STILL LOCKED AND NOT UNLOCKED. THE NEW PRODUCT WAS OPENED AND THE OPE WAS SUCCESSFULLY COMPLETED. THE PATIENT HAS NO HEALTH HAZARDS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15592070 ↗
  • MalfunctionSep 13, 2022Fitting Problem; Fitting Problem

    THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE THE HST III SEAL (4.5MM) FELL OFF WHEN THE DELIVERY SYSTEM WAS WITHDRAWN FROM THE LOADING DEVICE. THE BLUE SLIDE WAS LOCKED. SINCE THE SEAL WAS COMPLETELY OUT OF THE DELIVERY TUBE, IT WAS NOT IN CONTACT WITH THE PATIENT'S AORTA. A NEW HSIII WAS ISSUED AND THE OPERATION WAS COMPLETED. THE PATIENT HAD NO HEALTH HAZARDS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15414323 ↗
  • MalfunctionSep 1, 2022Fitting Problem; Fitting Problem

    THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE THE HST III SEAL (4.5MM), WAS NOT PROPERLY LOADED INTO THE DEVICE, PREVENTING IT FROM DEPLOYING. A NEW HEART STRING WAS OPENED AND THE SURGERY WAS COMPLETED WITHOUT ANY PROBLEMS. THE PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15342168 ↗
  • MalfunctionAug 30, 2022Activation Problem; Activation Problem

    THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE THE HST III SEAL (4.5MM), WAS USED AND THE HEMOSTATIC SEAL WAS NOT PUSHED OUT OF THE ROTARY DEVICE WHEN THE WHITE PLUNGER BUTTON WAS PRESSED AND PULLED OUT. A NEW DEVICE WAS USED AND THE OPERATION WAS COMPLETED WITHOUT ANY PROBLEMS AND THERE WAS NO IMPACT ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15325273 ↗
  • MalfunctionAug 23, 2022Fitting Problem; Fitting Problem

    THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE HST III SEAL (4.5MM), 5-PACK THE SEAL COULD NOT BE LOADED, AND THE PROCEDURE WAS COMPLETED WITHOUT ANY PROBLEMS USING A SPARE HEART STRING. NO HEALTH HAZARD TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15277971 ↗
  • InjuryJul 20, 2022Fitting Problem; Activation Problem; Fitting Problem; Activation Problem

    THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, SEAL DID NOT FILL INTO THE TUBE OF THE DELIVERY DEVICE. THE SEAL DID NOT LOAD PROPERLY. AT THE SETTING STAGE, THERE WAS NO SEAL INSIDE THE TUBE. THEY SET IT ACCORDING TO THE PROCEDURE, BUT COULD NOT DO IT MANY TIMES, SO THEY ROLLED THE SEAL BY HAND AND FILLED IT IN THE TUBE AND TRIED TO DEPLOY IT IN THE AORTA, BUT THE SEAL WAS CAUGHT IN THE TUBE AND COULD NOT BE DEPLOYED. THEREFORE, PREPARE AND USE AN ALTERNATIVE HEARTSTRING. THERE WAS A DELAY OF A FEW MINUTES. THERE WAS SOME INCREASE IN BLEEDING, BUT NO ADDITIONAL BLOOD TRANSFUSIONS WERE NEEDED. NO EFFECT ON THE PATIENT.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions

    MAUDE report 15071025 ↗
  • MalfunctionJul 15, 2022Fitting Problem; Fitting Problem

    THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE THE HST III SEAL DID NOT LOAD CORRECTLY. THEY USED A SECOND DEVICE AND IT WORKED AS INTENDED. THERE WAS NO HARM OR ADVERSE EVENT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15027089 ↗
  • MalfunctionJul 12, 2022Fitting Problem; Fitting Problem

    RELATED TO (B)(4). THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, HST III SEAL (4.5MM), 5-PACK THE SEAL IN THE DELIVERY SHEATH IS TWISTED WHEN PREPARED FOR HS USE AND CANNOT BE LOADED. THE OPERATION WAS COMPLETED WITH NO PROBLEMS USING THE NEW DEVICE AND THE PATIENT IS FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14987717 ↗
  • MalfunctionJul 12, 2022Fitting Problem; Fitting Problem

    RELATED TO (B)(4). SUMMARY: THE HOSPITAL REPORTED THAT DURING PREPARATION FOR A CORONARY ARTERY BYPASS PROCEDURE AFTER AN HS FAILURE, HST III SEAL (4.5MM), 5-PACK THE SEAL IN THE DELIVERY SHEATH IS KINKED AND CANNOT BE LOADED. THE OPERATION WAS COMPLETED WITH NO PROBLEMS USING THE NEW DEVICE AND THE PATIENT IS FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14987643 ↗
  • MalfunctionJun 29, 2022Activation Problem; Activation Problem

    THE HOSPITAL REPORTED THAT DURING PREPARATION FOR A CORONARY ARTERY BYPASS PROCEDURE USING HST III SEAL (4.5MM), 5-PACK. THEY FOLLOWED THE PROCEDURE, BUT A STICKER (SEAL) REMAINED IN THE DELIVERY DEVICE. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14853043 ↗
  • MalfunctionJun 28, 2022Fitting Problem; Fitting Problem

    THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, HST III SEAL (4.5MM), 5-PACK SEAL HAD FALLEN OFF WHEN THE DELIVERY SYSTEM WAS PULLED OUT OF THE LOADING DEVICE. THE OPERATION WAS COMPLETED WITH THE NEW HS#. THE PATIENT HAD NO HEALTH HAZARD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14844733 ↗
  • MalfunctionJun 17, 2022Fitting Problem; Audible Prompt/Feedback Problem; Fitting Problem; Audible Prompt/Feedback Problem

    THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE HST III SEAL (4.5MM), 5-PACK, THE SEAL WAS NOT LOADED AND REMAINED IN THE LOADER, MAKING IT UNUSABLE, SO THE NEW ITEM WAS OPENED AND THE SURGERY WAS COMPLETED SUCCESSFULLY. THERE WAS NO PROBLEM FOR THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14734394 ↗
  • InjuryMay 16, 2022Activation Problem; Activation Problem

    THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE USING HST III SEAL (4.5MM). THERE WAS NO CALCIFICATION AND LOADING WAS PERFORMED AS USUAL, BUT BLEEDING WAS ABNORMALLY LARGE AND IT BECAME A PROCEDURE TO SWITCH TO THE SIDE CLAMP. THE AMOUNT OF BLEEDING WAS NOT THE AMOUNT THAT REQUIRED A BLOOD TRANSFUSION. THE SEAL WAS DEPLOYED IN THE AORTA, BUT THE BLEEDING WAS SEVERE, WHETHER THE SEAL WAS IN A BAD SHAPE OR IN A BAD POSITION. IT IS CLEARLY DIFFERENT FROM WHEN THEY USE IT ALL THE TIME, AND IT SEEMS THAT THE PRODUCT IS DEFECTIVE. THE NEW HEART STRING IS NOT USED AND IT IS SUPPORTED BY THE SIDE CLAMP. NO HARM TO THE PATIENT.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 14409048 ↗
  • MalfunctionMay 10, 2022Fitting Problem; Fitting Problem

    THE HOSPITAL REPORTED THAT AT THE TIME OF OPENING HST III SEAL (4.5MM), 5-PACK, THE SEAL WAS NOT IN THE NORMAL POSITION AND IT WAS NOT READY FOR LOADING, SO THEY DID NOT USE IT FOR THE PATIENT AND FINISHED THE SURGERY WITH A NEW PRODUCT. THE SEAL DID NOT FIT PROPERLY AFTER OPENING. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14347549 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K130382. One submission often covers many models, so check the product and lot details in each record.

  • Z-2587-2025 · Aug 15, 2025 · Open, Classified
    Heartstring III Proximal Seal System, 3.8 MM. Intravascular anastomosis occluder.
  • Z-2588-2025 · Aug 15, 2025 · Open, Classified
    Heartstring III Proximal Seal System, 4.3 MM. Intravascular anastomosis occluder.

Other Clamp, Vascular devices