Skip to content

REMstar System One 60 Series REMstar Pro C-Flex+, 60 Series, AURecalls & reported problems

Respironics, Inc. · REMstar Pro C-Flex+, System One 60 Series, Australia · Class 2

All models of REMstar System One 60 Series (92)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Respironics REMstar Pro C-Flex+ System One 60 Series is a non-continuous ventilator (CPAP/BiPAP type machine) sold in Australia. Its 43 reports mostly describe the sound abatement foam breaking down, leaving black particles and bits of foam in the humidifier, mask, or air path, along with coughs, sore throat, headaches and nose bleeds. No deaths and 5 injuries were reported, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
43
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 43 adverse event reports that list this exact device (DI 00606959005440), received between Jul 2, 2021 and May 14, 2025 — classified as 5 injuries, 38 malfunctions.

The most frequently reported specific device problem is “Degraded” (42 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Degraded42
  • Adverse Event Without Identified Device or Use Problem1
  • Unraveled Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions22
  • Cough7
  • Headache5
  • Cancer3
  • Asthma2
  • Sore Throat2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Bone Fracture(s)1

Recent adverse event reports

The 25 most recent of 43 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 14, 2025Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE PATIENT ALLEGED NOSE BLEEDS, HEADACHES, PAIN AROUND HEART. THERE WAS NO REPORT OF SERIOUS OR PERMANENT PATIENT HARM OR INJURY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: Chest Pain; Epistaxis; Headache

    MAUDE report 22018594 ↗
  • MalfunctionNov 19, 2024Degraded; Adverse Event Without Identified Device or Use Problem; Degraded; Adverse Event Without Identified Device or Use Problem

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING THAT THE REMSTAR PRO C-FLEX+ DEVICE HAD CAUSED THE DEVELOPMENT OF LUNG CANCER. LYMPH NODES WERE BIOPSIED AND AWAITING RESULTS. THE MANUFACTURER'S INVESTIGATION IS ONGOING. THE DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: Cancer; Fatigue

    MAUDE report 20727183 ↗
  • MalfunctionJul 22, 2024Unraveled Material; Unraveled Material

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT ALLEGED DRY HACKING COUGH. THERE WAS NO REPORT OF SERIOUS PATIENT HARM OR INJURY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: Cough

    MAUDE report 19793182 ↗
  • MalfunctionJun 3, 2024Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT HAS ALLEGED VISUALIZATION OF PARTICLES. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19445564 ↗
  • MalfunctionMay 14, 2024Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING VISUALIZATION OF PARTICLES IN THE AIR PATH. THE PATIENT ALLEGES FEELING "QUITE UNWELL" AND ALSO ALLEGES TO HAVE NOTICED BLACK PARTICLES AND WAS CONCERNED. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE PATIENT ADDITIONALLY MENTIONED SEEKING REPLACEMENT FOR A RESMED F20 MEDIUM - FULL FACE MASK. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: Malaise

    MAUDE report 19310811 ↗
  • MalfunctionOct 4, 2023Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING VISUALIZATION OF BLACK PARTICLES ON INLINE FILTER. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17871062 ↗
  • MalfunctionAug 10, 2023Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING VISUALIZATION OF PARTICLES IN THE AIR PATH. THE PATIENT ALLEGES COUGHING, GAGGING, DRY THROAT. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Cough

    MAUDE report 17502051 ↗
  • MalfunctionJul 13, 2023Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT HAS ALLEGED TO SEEING SMALL PARTICLES IN THE MASK, BITS OF BLACK FOAM IN THE HUMIDIFIER AND THE FOAM/FILTER IS DISINTEGRATING. THE PATIENT ALLEGES HARM FROM THE DEVICE. THERE WAS NO REPORT OF SERIOUS PATIENT HARM OR INJURY. THERE WAS NO MEDICAL INTERVENTION REQUIRED BY THE PATIENT. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: Insufficient Information

    MAUDE report 17317780 ↗
  • MalfunctionJan 25, 2023Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING VISUALIZATION OF BLACK PARTICLES IN THE TUBING. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16239947 ↗
  • MalfunctionNov 4, 2022Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING VISUALIZATION OF BLACK PARTICLES IN THE MASK. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15736414 ↗
  • MalfunctionSep 30, 2022Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING VISUALIZATION OF BLACK PARTICLES IN THE TUBING AND MASK. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15519042 ↗
  • MalfunctionSep 13, 2022Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING THE PATIENT IS EXPERIENCING A BURNING IN THE NOSE WHILE USING THE DEVICE. PATIENT ALSO ALLEGED VISUALIZATION OF PARTICLES IN THE HUMIDIFIER. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: Burning Sensation

    MAUDE report 15411048 ↗
  • MalfunctionAug 22, 2022Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING VISUALIZATION OF BLACK PARTICLES/DOTS IN THE DEVICE. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15266889 ↗
  • MalfunctionJul 26, 2022Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT ALLEGED VISUALIZATION OF BLACK PARTICLES IN THE HUMIDIFIER AND WATER CHAMBER. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15106176 ↗
  • MalfunctionJun 29, 2022Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT ALLEGED HEADACHES WHILE USING THE DEVICE. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: Headache

    MAUDE report 14852935 ↗
  • MalfunctionMay 24, 2022Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT ALLEGED LUNG ISSUES WHILE USING THE DEVICE. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: Unspecified Respiratory Problem

    MAUDE report 14488069 ↗
  • MalfunctionMay 19, 2022Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT ALLEGED VISUALIZATION OF BLACK PARTICLES ON MASK. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14436647 ↗
  • MalfunctionMay 11, 2022Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT ALLEGED LOW OXYGEN LEVELS, DIZZINESS, AND SERIOUS SKIN ERUPTIONS THAT ARE BREAKING OUT CONSTANTLY WHILE USING THE DEVICE. MEDICAL INTERVENTION WAS NOT SPECIFIED. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: Dizziness; Low Oxygen Saturation; Skin Inflammation/ Irritation

    MAUDE report 14360630 ↗
  • MalfunctionMay 4, 2022Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT ALLEGED COUGH. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: Cough

    MAUDE report 14282359 ↗
  • MalfunctionMay 4, 2022Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT ALLEGED VISUALISATION OF BLACK PARTICLES IN THE WATER TANK . THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14278180 ↗
  • MalfunctionApr 29, 2022Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT ALLEGED VISUALIZATION BLACK DEPOSITS FORMING IN MASK. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14254496 ↗
  • MalfunctionApr 25, 2022Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT ALLEGED HUMIDIFIER WAS NOT WORKING AND ALSO ALLEGED TIREDNESS AND HEADACHES . THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: Headache

    MAUDE report 14193846 ↗
  • MalfunctionApr 25, 2022Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT ALLEGED SORE THROAT AND CONSISTENT COUGH . THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: Cough; Sore Throat

    MAUDE report 14193843 ↗
  • MalfunctionApr 20, 2022Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO THE CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT ALLEGED RED BLOTCHES ALL OVER HIS FACE. THERE WAS NO REPORT OF SERIOUS OR PERMANENT PATIENT HARM OR INJURY. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.

    Patient problems: Rash

    MAUDE report 14161051 ↗
  • MalfunctionApr 12, 2022Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT ALLEGED VISUALIZATION OF BLACK PARTICLES IN THE HUMIDIFIER AND MASK. PATIENT ALSO ALLEGED AN ODD FEELING IN HIS THROAT. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.

    Patient problems: Sore Throat

    MAUDE report 14084496 ↗

Other VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices

All VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page