BiPAP autoSV Advanced System One 60 Series BiPAP autoSV Advanced with Humidifier, 60 Series, USRecalls & reported problems

Respironics, Inc. · BiPAP autoSV Advanced System One 60 Series with Humidifier, US · Class 2

All models of BiPAP autoSV Advanced System One 60 Series (19)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Respironics BiPAP autoSV Advanced is a breathing machine, and its 497 reports mostly describe sound-reducing foam inside the device breaking down into particles. Some reports also describe contamination and overheating, and patient problems listed include headaches and dizziness. There are 9 deaths and 91 injuries reported, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
494
Deaths reported
9
Injuries reported
91
Recalls naming it
0

Summary

The FDA's MAUDE database contains 494 adverse event reports that list this exact device (DI 00606959012523), received between Sep 1, 2021 and Jul 3, 2026 — classified as 9 deaths, 91 injuries, 394 malfunctions.

The most frequently reported specific device problem is “Degraded” (491 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 5 reportsDec 2023: 2 reportsJan 2024: 3 reportsFeb 2024: 4 reportsMar 2024: 5 reportsApr 2024: 10 reportsMay 2024: 12 reportsJun 2024: 15 reportsJul 2024: 10 reportsAug 2024: 5 reportsSep 2024: 2 reportsOct 2024: 33 reportsNov 2024: 7 reportsDec 2024: 1 reportsJan 2025: 9 reportsFeb 2025: 30 reportsMar 2025: 20 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 7 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 46 reportsDec 2025: 0 reportsJan 2026: 9 reportsFeb 2026: 8 reportsMar 2026: 14 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 5 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202646
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Degraded491
  • Contamination53
  • Overheating of Device13
  • Adverse Event Without Identified Device or Use Problem6
  • Ambient Noise Problem1
  • Device Emits Odor1
  • Leak/Splash1
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions344
  • Unspecified Respiratory Problem54
  • Cancer46
  • Headache35
  • Dizziness30
  • Unspecified Kidney or Urinary Problem23
  • Dyspnea18
  • Asthma17

Recent adverse event reports

The 25 most recent of 494 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 3, 2026Degraded

    A BIPAP AUTOSV ADV DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION, THE DEVICE WAS VISUALLY INSPECTED AND FOAM WAS OBSERVED IN TECHNICAL FINDINGS. ADDITIONALLY, THE PATIENT'S COMPLAINT WAS CONFIRMED AND THE DEVICE WAS SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25703089 ↗
  • MalfunctionJun 29, 2026Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A BIPAP AUTOSV ADV DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER THE DEVICE WAS VISUALLY INSPECTED AND EVIDENCE OF FOAM DEGRADATION WAS FOUND. MOREOVER, THE DEVICE HAD NO CONTAMINATION AND UNRELATED ERROR CODES. IN ADDITION, THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25659210 ↗
  • MalfunctionJun 12, 2026Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A BIPAP AUTOSV ADV DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, THE THIRD-PARTY SERVICE CENTER VISUALLY INSPECTED THE DEVICE AND OBSERVED VISIBLE FOAM PARTICLES. ADDITIONALLY, THE SERVICE TECHNICIAN FOUND MOISTURE CONTAMINATION AND THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25493647 ↗
  • MalfunctionJun 12, 2026Degraded; Overheating of Device

    A BIPAP AUTOSV DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR SERVICE WITH AN INITIAL ALLEGATION OF A BURNT CONNECTOR. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. THERE WAS NO REPORT OF MEDICAL INTERVENTION. DURING THE EVALUATION OF THE DEVICE, THE COMPLAINT WAS CONFIRMED. THE SERVICE TECHNICIAN ALSO OBSERVED VISIBLE FOAM PARTICLES. DUE TO THE ALLEGED MALFUNCTION, THE DEVICE WILL BE RETURNED TO THE MANUFACTURER'S PRODUCT INVESTIGATION LAB FOR ADDITIONAL TESTING. THE ROOT CAUSE IS NOT CONFIRMED AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25492555 ↗
  • MalfunctionJun 11, 2026Degraded

    A BIPAP AUTOSV ADV WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR SERVICE AS PART OF THE SOUND ABATEMENT FOAM RECALL PROCESS. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM. ADDITIONALLY, UNRELATED TO THE REPORTABLE MALFUNCTION, THE TECHNICIAN OBSERVED AN ERROR CODE. THE DEVICE WAS SCRAPPED AFTER THE EVALUATION WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25474483 ↗
  • MalfunctionJun 5, 2026Degraded

    A BIPAP AUTOSV ADV DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR SERVICE AS PART OF THE SOUND ABATEMENT FOAM RECALL PROCESS. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. THERE WAS NO REPORT OF MEDICAL INTERVENTION. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF VISIBLE FOAM PARTICLES WAS FOUND. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED A CUT/BROKEN POWER CONNECTOR, AND THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25417420 ↗
  • MalfunctionApr 6, 2026Degraded

    A BIPAP AUTOSV ADVANCED DEVICE WAS RETURNED TO A SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND VISIBLE FOAM PARTICLES WERE OBSERVED. ADDITIONALLY, THE DEVICE WAS DISPOSITIONED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24795859 ↗
  • MalfunctionMar 12, 2026Degraded

    A BIPAP AUTOSV DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR SERVICE AS PART OF THE SOUND ABATEMENT FOAM RECALL PROCESS. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES INSIDE THE BLOWER KIT AND BLOWER FOAM DEGRADATION. ADDITIONALLY, UNRELATED TO THE REPORTABLE MALFUNCTION, THE TECHNICIAN OBSERVED DUST CONTAMINATION AND THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24586848 ↗
  • MalfunctionMar 9, 2026Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A BIPAP AUTOSV ADV DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION/PARTICLES WAS FOUND INSIDE THE BLOWER KIT. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST AND FLUIDS FAIL CONTAMINATION, AND THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24552384 ↗
  • MalfunctionMar 9, 2026Degraded

    A BIPAP AUTOSV ADV DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR SERVICE AS PART OF THE SOUND ABATEMENT FOAM RECALL PROCESS. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM PARTICLES INSIDE THE BLOWER KIT AND A FOAM DEGRADATION. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST CONTAMINATION. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24543356 ↗
  • MalfunctionMar 5, 2026Degraded

    A BIPAP AUTOSV ADV DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER AS PART OF THE SOUND ABATEMENT FOAM RECALL PROCESS WITH NO INITIAL ALLEGED COMPLAINT. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. THERE WAS NO REPORT OF MEDICAL INTERVENTION. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION/PARTICLES WAS FOUND INSIDE THE BLOWER KIT. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST CONTAMINATION WAS PRESENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24524163 ↗
  • MalfunctionMar 5, 2026Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A BIPAP AUTOSV ADV DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF VISIBLE FOAM PARTICLES WAS FOUND INSIDE THE BLOWER KIT. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST FAIL CONTAMINATION, AND THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24521717 ↗
  • MalfunctionMar 5, 2026Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A BIPAP AUTOSV ADVANCED DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION/PARTICLES WAS FOUND INSIDE THE BLOWER KIT. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST CONTAMINATION, AND THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24518629 ↗
  • MalfunctionMar 5, 2026Degraded

    A BIPAP AUTOSV DEVICE WAS RETURNED TO A THIRD PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. THERE WAS NO REPORT OF MEDICAL INTERVENTION. DURING THE EVALUATION OF THE DEVICE, THE THIRD PARTY SERVICE CENTER VISUALLY INSPECTED THE DEVICE AND OBSERVED BISIBLE FOAM PARTICLES INSIDE THE BLOWER KIT AND NOTED A BLOWER FOAM DEGRADATION. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST CONTAMINATION, AND THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITION THAT ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BUY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED. THIS EVENT IS REPORTABLE UNDER FDA 21 CFR PART 803 AS IT MEETS THE CRITERIA FOR A SERIOUS INJURY AND OR PRODUCT MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24518004 ↗
  • MalfunctionMar 3, 2026Degraded

    A BIPAP AUTOSV ADV DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. THERE WAS NO REPORT OF MEDICAL INTERVENTION. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION/PARTICLES WAS FOUND INSIDE THE BLOWER KIT. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST AND FLUID CONTAMINATION, AND THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24500751 ↗
  • MalfunctionMar 2, 2026Contamination; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A BIPAP AUTOSV ADV DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND NO FOAM PARTICLES WERE NOTED IN THE AIRPATH, YET FOAM DEGRADATION WAS NOTED. THE DEVICE'S SOUND ABATEMENT FOAM WAS REPLACED TO ADDRESS THE ISSUE. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST CONTAMINATION, AND THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS REPAIRED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24489756 ↗
  • MalfunctionMar 2, 2026Degraded; Adverse Event Without Identified Device or Use Problem

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A BIPAP AUTOSV ADV DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. THERE WAS NO REPORT OF MEDICAL INTERVENTION. DURING EVALUATION NO FOAM PARTICLE WERE NOTED IN THE AIR PATH, YET FOAM DEGRADATION WAS NOTED. THE DEVICE SOUND ABATEMENT FOAM NEEDS REPLACED TO ADDRESS THE ISSUE. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST CONTAMINATION, AND THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24489520 ↗
  • MalfunctionMar 2, 2026Contamination; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A BIPAP AUTOSV DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND NO FOAM PARTICLES WERE NOTED IN THE AIRPATH, YET FOAM DEGRADATION WAS NOTED. THE DEVICES SOUND ABATEMENT FOAM WAS REPLACED TO ADDRESS THE ISSUE. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST CONTAMINATION, AND THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24485831 ↗
  • MalfunctionMar 2, 2026Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A BIPAP AUTOSV ADV DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND NO FOAM PARTICLES WERE NOTED IN THE AIRPATH, YET FOAM DEGRADATION WAS NOTED. THE DEVICES SOUND ABATEMENT FOAM WAS REPLACED TO ADDRESS THE ISSUE. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24484784 ↗
  • MalfunctionMar 2, 2026Degraded

    A BIPAP AUTOSV DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR SERVICE AS PART OF THE SOUND ABATEMENT FOAM RECALL PROCESS. DURING THE EVALUATION, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES INSIDE THE BLOWER KIT AND OBJECT CODE INDICATES THE BLOWER CONTRIBUTED TO THE FOAM DEGRADATION. ADDITIONALLY, DEVICE HAD DUST FAIL CONTAMINATION. THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED. THIS EVENT IS REPORTABLE UNDER FDA 21 CFR PART 803 AS IT MEETS THE CRITERIA FOR A SERIOUS INJURY AND/ OR PRODUCT MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24483531 ↗
  • MalfunctionMar 2, 2026Contamination; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A BIPAP AUTOSV ADV DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND NO FOAM PARTICLES WERE NOTED IN THE AIRPATH, YET FOAM DEGRADATION WAS NOTED. THE DEVICES SOUND ABATEMENT FOAM WAS REPLACED TO ADDRESS THE ISSUE. ADDITIONALLY, THE SERVICE TECHNICIAN FOUND DUST CONTAMINATION AND THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS REPAIRED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24481719 ↗
  • MalfunctionFeb 27, 2026Degraded

    A BIPAP AUTOSV DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER AS PARTI OF THE SOUND ABATEMENT FOAM RECALL PROCESS WITH NO INITIAL ALLEGED COMPLAINT. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED OBJECT CODE INDICATING THE CONTRIBUTED TO BLOWER FOAM DEGRADATION. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST CONTAMINATION, AND THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24475573 ↗
  • MalfunctionFeb 26, 2026Degraded

    A BIPAP AUTOSV ADV DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR SERVICE AS PART OF THE SOUND ABATEMENT FOAM RECALL PROCESS WITH NO INITIAL ALLEGED COMPLAINT. DURING THE DEVICE EVALUATION, THE SERVICE TECHNICIAN NOTED VISIBLE FOAM PARTICLES INSIDE THE BLOWER KIT. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED FLUIDS FAIL CONTAMINATION AND UNRELATED ERROR CODES, WHICH ARE CONSISTENT WITH NORMAL DEVICE OPERATION UNDER EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24457570 ↗
  • MalfunctionFeb 26, 2026Degraded

    A BIPAP AUTOSV DEVICE WAS RETURNED TO A THIRD PARTY SERVIC CENTER AS PART OF THE SOUND ABATEMENT FOAM RECALL PROCESS. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED OBJECT CODE INDICATES THE BLOWER CONTRIBUTED TO THE FOAM DEGRADATION. ADDITIONALLY, DEVICE HAD DUST FAIL CONTAMINATION. THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED. THIS EVENT IS REPORTABLE UNDER FDA 21 CFR PART 803 AS IT MEETS THE CRITERIA FOR A SERIOUS INJURY AND/OR PRODUCT MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24455278 ↗
  • MalfunctionFeb 25, 2026Degraded

    A BIPAP AUTOSV ADV DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. THERE WAS NO REPORT OF MEDICAL INTERVENTION. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION/PARTICLES WAS FOUND INSIDE THE BLOWER KIT. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST CONTAMINATION. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24444559 ↗

Other VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices

All VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices →