Skip to content

System One System One Heated Humidifier, INTLRecalls & reported problems

Respironics, Inc. · System One Heated Humidifier, International · Class 2

All models of System One (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
1
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00606959014244), received between Sep 24, 2024 and Feb 18, 2026 — classified as 1 death, 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Degraded” (6 reports).

2 other recalls cover devices cleared under the same FDA submission (K113068); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 3 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Degraded6
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Cancer1
  • Insufficient Information1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 18, 2026Degraded

    A DEVICE WAS RETURNED TO THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM DEGRADATION. THE DEVICE POWERED ON AND AIRFLOW WAS CONFIRMED. THE DEVICE'S DOWNLOADED LOGS WERE REVIEWED BY THE THIRD-PARTY SERVICE CENTER. THERE WERE THIRTY-TWO ERROR CODE FOUND. THE THIRD-PARTY SERVICE CENTER CONCLUDED THERE WAS A VISIBLE FOAM DEGRADATION AND UNIT GOT SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24374894 ↗
  • MalfunctionFeb 16, 2026Degraded

    A DEVICE WAS RETURNED TO THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM DEGRADATION. THE DEVICE POWERED ON AND AIRFLOW WAS CONFIRMED. THE DEVICE'S DOWNLOADED LOGS WERE REVIEWED BY THE THIRD-PARTY SERVICE CENTER. THERE WERE 12 ERROR CODES FOUND. THE THIRD-PARTY SERVICE CENTER CONCLUDED THERE WAS A VISIBLE FOAM DEGRADATION AND UNIT GOT SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24342474 ↗
  • MalfunctionFeb 13, 2026Degraded

    A DEVICE WAS RETURNED TO THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM DEGRADATION. THE DEVICE POWERED ON AND AIRFLOW WAS CONFIRMED. THE DEVICE'S DOWNLOADED LOGS WERE REVIEWED BY THE THIRD-PARTY SERVICE CENTER. THERE WERE NO ERROR CODES FOUND. THE THIRD-PARTY SERVICE CENTER CONCLUDES THAT THERE WAS A VISIBLE FOAM DEGRADATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24329899 ↗
  • MalfunctionNov 28, 2025Degraded; Degraded

    A DEVICE WAS RETURNED TO A MANUFACTURER SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM DEGRADATION. THE DEVICE POWERED ON AND AIRFLOW WAS CONFIRMED. THE DEVICE'S DOWNLOADED LOGS WERE REVIEWED BY THE MANUFACTURER SERVICE CENTER. THERE WAS THIRTEEN ERROR CODES FOUND. THE MANUFACTURER SERVICE CENTER CONCLUDES THAT THERE WAS VISIBLE FOAM DEGRADATION AND UNIT GOT SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23670467 ↗
  • DeathNov 1, 2024Degraded; Adverse Event Without Identified Device or Use Problem; Degraded; Adverse Event Without Identified Device or Use Problem

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP/ BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING PT HAS PASSED AWAY. NO OTHER CLINICAL INFORMATION OR MEDICAL INTERVENTIONS WERE REPORTED. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20585362 ↗
  • InjurySep 24, 2024Degraded; Degraded

    THE CUSTOMER REPORTED SERIOUS INJURY OF CANCER (LUNG CARCINOMA) AND DAMAGE TO LUNG TISSUE OCCURRED IN RELATION TO THE DEVICE. THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING SERIOUS INJURY OF CANCER AND LUNG DAMAGE. MEDICAL INTERVENTION WAS NOT SPECIFIED. THE MANUFACTURER IS IN THE PROCESS OF OBTAINING ADDITIONAL INFORMATION CONCERNING THIS EVENT AND THE COMPLAINT IS STILL UNDER INVESTIGATION. A FINAL REPORT WILL BE SUBMITTED ONCE THE INVESTIGATION IS COMPLETE.

    Patient problems: Cancer

    MAUDE report 20295096 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K113068. One submission often covers many models, so check the product and lot details in each record.

  • Z-1817-2024 · Apr 1, 2024 · Open, Classified
    OmniLab Advanced + Ventilator Product Numbers 1111122 1111123 1111124 1111125 1111126 1111127 1111138 1111141 1111142 KR1111127 R1111122 R1111123 R1111124 U1111122 U1111124
  • Z-1974-2021 · Jun 14, 2021 · Open, Classified
    SystemOne (Q-Series), DreamStation, DreamStation Go, Dorma 400, Dorma 500, REMstar SE Auto Noncontinuous Ventilators The Philips Respironics E30 ventilator is intended to provi

Other VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices

All VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page