Skip to content

DreamStation 15MM Heated TubeRecalls & reported problems

Respironics, Inc. · DreamStation 15mm Heated Tube · Class 2

All models of Philips DreamStation (190)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00606959022331), received between Apr 24, 2025 and Jun 29, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Excessive Heating” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Excessive Heating3
  • Degraded1
  • Thermal Decomposition of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 29, 2026Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A DREAMSTATION AUTO CPAP DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF VISIBLE FOAM PARTICLES WAS FOUND INSIDE THE BLOWER KIT. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST FAIL CONTAMINATION. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25658816 ↗
  • MalfunctionJun 10, 2026Thermal Decomposition of Device; Excessive Heating

    THE MANUFACTURER RECEIVED AN INFORMATION ALLEGING THAT 15MM HEATED TUBE HAS A BLACK MARK WHICH LOOKS LIKE A BURNED PLASTIC INSIDE THE TUBING MATERIAL. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THE DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25465138 ↗
  • MalfunctionApr 22, 2026Excessive Heating

    THE DEVICE WAS RECEIVED AND EVALUATED BY THE MANUFACTURER¿S PRODUCT INVESTIGATION LAB (PIL). THE 15MM HEATED TUBE RETURNED IN AN OPENED SHIPPING PACKAGING AND COULD OBSERVE A DAMAGED COIL WIRE (PLASTIC ON WIRE LOOKING AS IF IT WAS MELTED). THE PLASTIC TUBING WAS NOT DAMAGED. PIL COULD DETERMINE THE REST OF THE TUBING WAS DAMAGE AND DEFECT FREE AND STILL IN A NORMAL DESIGN ORIENTATION AS INTENDED. PIL CONFIRMED THE COMPLAINT OF ¿BROKEN/PIERCED/EXPOSED HEATING WIRES¿ AND COULD NOT DETERMINE THE CAUSE FOR THE COMPLAINT. PIL COULD CONCLUDE THAT BASED ON THE CAUSE DETERMINATION FOR THE COMPLAINT, NO FURTHER CORRECTIVE ACTION IS REQUIRED FOR THIS COMPLAINT. NO FURTHER INVESTIGATION IS NEEDED. THE 15MM HEATED TUBE HAS BEEN SCRAPPED PER THE REQUIREMENTS SET FORTH IN WORK INSTRUCTION 8.4-1131 REVISION 10 AND DISPOSED OF ACCORDINGLY. NO FURTHER INVESTIGATION WILL BE PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24956441 ↗
  • MalfunctionApr 24, 2025Excessive Heating; Excessive Heating

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING HEATED TUBING. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21901687 ↗

Other VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices

All VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices →