DreamStation 15MM Heated TubeRecalls & reported problems
Respironics, Inc. · DreamStation 15mm Heated Tube · Class 2
All models of Philips DreamStation (190)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00606959022331), received between Apr 24, 2025 and Jun 29, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Excessive Heating” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Excessive Heating3
- Degraded1
- Thermal Decomposition of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 29, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A DREAMSTATION AUTO CPAP DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF VISIBLE FOAM PARTICLES WAS FOUND INSIDE THE BLOWER KIT. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST FAIL CONTAMINATION. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25658816 ↗MalfunctionJun 10, 2026Thermal Decomposition of Device; Excessive Heating
THE MANUFACTURER RECEIVED AN INFORMATION ALLEGING THAT 15MM HEATED TUBE HAS A BLACK MARK WHICH LOOKS LIKE A BURNED PLASTIC INSIDE THE TUBING MATERIAL. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THE DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25465138 ↗MalfunctionApr 22, 2026Excessive Heating
THE DEVICE WAS RECEIVED AND EVALUATED BY THE MANUFACTURER¿S PRODUCT INVESTIGATION LAB (PIL). THE 15MM HEATED TUBE RETURNED IN AN OPENED SHIPPING PACKAGING AND COULD OBSERVE A DAMAGED COIL WIRE (PLASTIC ON WIRE LOOKING AS IF IT WAS MELTED). THE PLASTIC TUBING WAS NOT DAMAGED. PIL COULD DETERMINE THE REST OF THE TUBING WAS DAMAGE AND DEFECT FREE AND STILL IN A NORMAL DESIGN ORIENTATION AS INTENDED. PIL CONFIRMED THE COMPLAINT OF ¿BROKEN/PIERCED/EXPOSED HEATING WIRES¿ AND COULD NOT DETERMINE THE CAUSE FOR THE COMPLAINT. PIL COULD CONCLUDE THAT BASED ON THE CAUSE DETERMINATION FOR THE COMPLAINT, NO FURTHER CORRECTIVE ACTION IS REQUIRED FOR THIS COMPLAINT. NO FURTHER INVESTIGATION IS NEEDED. THE 15MM HEATED TUBE HAS BEEN SCRAPPED PER THE REQUIREMENTS SET FORTH IN WORK INSTRUCTION 8.4-1131 REVISION 10 AND DISPOSED OF ACCORDINGLY. NO FURTHER INVESTIGATION WILL BE PERFORMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24956441 ↗MalfunctionApr 24, 2025Excessive Heating; Excessive Heating
THE MANUFACTURER RECEIVED INFORMATION ALLEGING HEATED TUBING. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21901687 ↗
Other VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices
- REMstar System One 60 Series REMstar Auto A-Flex, w/Humid System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)117 deaths10,075 reports
- DreamStation DreamStation Auto BiPAP w/Humidifier, w/Cell, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)106 deaths5,937 reports
- DreamStation DreamStation Auto CPAP with Humidifier, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)44 deaths5,793 reports
- REMstar System One 60 Series REMstar Auto A-Flex System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)64 deaths5,138 reports
- REMstar System One 60 Series REMstar Pro C-Flex+, w/Humid System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)65 deaths4,520 reports
- REMstar System One 60 Series REMstar Auto A-Flex w/HT Humid System One 60 Series, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)43 deaths4,149 reports
- DreamStation DreamStation Auto CPAP w/Humid w/Heated Tube, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)30 deaths3,981 reports
- DreamStation DreamStation Auto CPAP w/Humidifier, w/Cell, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)86 deaths3,516 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page