Skip to content

DreamStation 00 (REF FRX500S14)Recalls & reported problems

Respironics, Inc. · DreamStation Auto CPAP, France · Class 2

All models of Philips DreamStation (190)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00606959023574), received between Sep 28, 2016 and Mar 13, 2026 — classified as 1 injury, 6 malfunctions.

The most frequently reported specific device problem is “Degraded” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 3 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Degraded6
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Cough3
  • Headache3
  • Fatigue2
  • No Clinical Signs, Symptoms or Conditions2
  • Chest Pain1
  • Dizziness1
  • Eye Irritation1
  • Nausea1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 13, 2026Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP/ BIPAP, AND MECHANICAL VENTILATOR DEVICES. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE DEVICE WAS RETURNED TO THIRD PARTY SERVICE CENTER AND EVALUATED. EVALUATION RESULT STATED THAT FOAM PARTICLES WERE OBSERVED. THE DEVICE HAS BEEN SCRAPPED. ADDITIONALLY, TOBACCO CONTAMINATION WAS PRESENT INSIDE THE DEVICE. THE MANUFACTURER IS SUBMITTING A FINAL REPORT AT THIS TIME. IF PERTINENT INFORMATION BECOMES AVAILABLE TO THE MANUFACTURER AT A LATER DATE, AN ADDENDUM TO THIS FINAL REPORT WILL BE FILED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24594483 ↗
  • MalfunctionDec 31, 2024Degraded; Degraded

    A DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. DURING THE EVALUATION OF THE DEVICE, THE THIRD-PARTY SERVICE CENTER VISUALLY INSPECTED THE DEVICE AND FOUND EVIDENCE OF FOAM DEGRADATION. DURING THE EVALUATION, FOAM PARTICLES WERE OBSERVED. THE DEVICE WAS SCRAPPED. THERE WAS NO INFORMATION ABOUT PRODUCT RETURNED DATE AND TIME. THE MANUFACTURER IS SUBMITTING A FINAL REPORT AT THIS TIME. IF PERTINENT INFORMATION BECOMES AVAILABLE TO THE MANUFACTURER AT A LATER DATE, AN ADDENDUM TO THIS FINAL REPORT WILL BE FILED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21046609 ↗
  • MalfunctionNov 11, 2024Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE/RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A DREAMSTATION AUTO DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT HAS ALLEGED COUGHING, CHEST PAIN, SKIN AND EYE IRRITATION. THERE WAS NO REPORT OF HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER. THE TECHNICIAN COULD NOT CONFIRM FOAM PARTICLES IN THE DEVICE EVALUATION. THE DEVICE WAS SCRAPPED. THE MANUFACTURER IS SUBMITTING A FINAL REPORT AT THIS TIME. IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE TO THE MANUFACTURER AT A LATER DATE, AN ADDENDUM TO THIS FINAL REPORT WILL BE FILED.

    Patient problems: Chest Pain; Cough; Eye Irritation; Skin Inflammation/ Irritation

    MAUDE report 20658241 ↗
  • MalfunctionNov 8, 2024Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE/RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A DREAMSTATION AUTO DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT HAS ALLEGED HEADACHE, DIZZINESS, NAUSEA, SEVERE FATIGUE. THERE WAS NO REPORT OF HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER. THE TECHNICIAN COULD NOT CONFIRM FOAM PARTICLES IN THE DEVICE EVALUATION. THE DEVICE WAS SCRAPPED. THE MANUFACTURER IS SUBMITTING A FINAL REPORT AT THIS TIME. IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE TO THE MANUFACTURER AT A LATER DATE, AN ADDENDUM TO THIS FINAL REPORT WILL BE FILED.

    Patient problems: Dizziness; Fatigue; Headache; Nausea

    MAUDE report 20639025 ↗
  • MalfunctionNov 8, 2024Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE/RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A DREAMSTATION AUTO DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT HAS ALLEGED HEADACHE AND COUGH. THERE WAS NO REPORT OF HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER. THE TECHNICIAN COULD NOT CONFIRM FOAM PARTICLES IN THE DEVICE EVALUATION. THE DEVICE WAS SCRAPPED. THE MANUFACTURER IS SUBMITTING A FINAL REPORT AT THIS TIME. IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE TO THE MANUFACTURER AT A LATER DATE, AN ADDENDUM TO THIS FINAL REPORT WILL BE FILED.

    Patient problems: Cough; Headache

    MAUDE report 20639024 ↗
  • MalfunctionSep 14, 2023Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING "HEADACHES, COUGHING, RESPIRATORY DISCOMFORT, CONGESTED BRONCHI ITCHING, GREAT FATIGUE IMPACT ON BLOOD PRESSURE STOP RESPIRATORY ASSISTANCE." NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: Cough; Fatigue; Headache; Unspecified Respiratory Problem

    MAUDE report 17745486 ↗
  • InjurySep 28, 2016Adverse Event Without Identified Device or Use Problem

    THE MANUFACTURER WAS MADE AWARE THAT AN END USER ALLEGED SHE WAS "POISONED" BY USING A CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE. THE END USER REPORTS BEING HOSPITALIZED. THE DATE OF THE ALLEGED EVENT, DATE OF ADMISSION, ADMITTING DIAGNOSIS, TREATMENT REQUIRED, AND LENGTH OF STAY ARE UNKNOWN AT THIS TIME. A REVIEW OF THE DEVICE HISTORY RECORDS FOR THE CPAP AND ASSOCIATED HUMIDIFIER INDICATE THEY PASSED ALL TESTING AND FUNCTIONED AS DESIGNED PRIOR TO LEAVING THE MANUFACTURING FACILITY. A SEARCH OF THE MANUFACTURER'S COMPLAINT SYSTEM INDICATES NO SIMILAR REPORTS OF "POISONING" OCCURRING WITH THESE DEVICES. THE MANUFACTURER REQUESTED RETURN OF THE DEVICES FOR EVALUATION, AND ADDITIONAL INFORMATION. A FOLLOW UP REPORT WILL BE FILED UPON COMPLETION OF THE MANUFACTURER'S INVESTIGATION.

    Patient problems: No Patient Involvement; Toxicity

    MAUDE report 5983274 ↗

Other VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices

All VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page