DreamStation DreamStation Heated Humidifier, DERecalls & reported problems
Respironics, Inc. · DreamStation Humidifier, Germany · Class 2
All models of Philips DreamStation (190)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00606959024113), received between Oct 21, 2025 and Nov 25, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Degraded” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Degraded1
- Overheating of Device1
Patient problems
- Dehydration1
- Electric Shock1
- No Clinical Signs, Symptoms or Conditions1
- Stroke/CVA1
- Syncope/Fainting1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 25, 2025Degraded; Degraded
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A PATIENT EXPERIENCED ELECTRICAL SHOCK FROM HEATED TUBING WHILE USING A DREAMSTATION AUTO CPAP DEVICE. THE PATIENT ALLEGES HE SWITCHED BACK TO THE NORMAL TUBING DAYS LATER AND PASSED OUT AND WAS IN THE HOSPITAL FOR 8 HOURS. THE DOCTORS COULDN'T TELL HIM WHAT HAD TAKEN PLACE; HOWEVER, THE PATIENT CLAIM HE HAD A STROKE. THE PATIENT ALSO ALLEGES DEHYDRATION DURING THE THERAPY, THE HUMIDIFIER SETTING WAS ON LEVEL 5, THE MASK TEST SHOWED 100% AND AHI VALUE WAS LESS THAN 5. THE DEVICE WAS RETURNED TO THE MANUFACTURER'S PRODUCT INVESTIGATION LABORATORY (PIL) FOR INVESTIGATION. THE DEVICE WAS APPLIED POWER AND VERIFIED AIRFLOW. WHILE VERIFYING AIRFLOW PIL OBSERVED THAT THE PRESSURE READING WAS LOW LIKELY DUE TO THE HUMIDIFIER LID SEAL AND DRY BOX SEAL MISSING FROM THE HUMIDIFIER. USING A KNOWN GOOD, HEATED TUBE PIL VERIFIED THAT THE TUBE FUNCTIONED PROPERLY AND DID NOT OBSERVE ANY TYPE OF SHOCK. DURING THE INVESTIGATION, PIL OBSERVED THE HUMIDIFIER LID SEAL, AND THE DRY BOX SEAL WAS NOT RETURNED WITH THE HUMIDIFIER. A DUST LIKE CONTAMINATION ON THE TOP ENCLOSURE, BOTTOM ENCLOSURE, BOTTOM ENCLOSURE OF THE HUMIDIFIER, BOTTOM PANEL, BACK PANEL, B
Patient problems: Dehydration; Electric Shock; Stroke/CVA; Syncope/Fainting
MAUDE report 23653085 ↗MalfunctionOct 21, 2025Overheating of Device; Overheating of Device
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THIS REPORT WAS RELATED TO A DREAMSTATION HUMIDIFIER. THE MANUFACTURER RECEIVED INFORMATION FROM THE PATIENT ALLEGING "SLIGHT BURN MARKS" ON THE INSIDE OF THE HUMIDIFIER LID, AND A "HOLE MELTED INTO THE LID SEAL." THE PATIENT ALSO ALLEGES AN ODOR DURING USAGE. NO DEATH, SERIOUS HARM, OR INJURY WAS REPORTED. THE DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP REPORT WILL BE FILED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23348998 ↗
Other VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices
- REMstar System One 60 Series REMstar Auto A-Flex, w/Humid System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)117 deaths10,075 reports
- DreamStation DreamStation Auto BiPAP w/Humidifier, w/Cell, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)106 deaths5,937 reports
- DreamStation DreamStation Auto CPAP with Humidifier, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)44 deaths5,793 reports
- REMstar System One 60 Series REMstar Auto A-Flex System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)64 deaths5,138 reports
- REMstar System One 60 Series REMstar Pro C-Flex+, w/Humid System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)65 deaths4,520 reports
- REMstar System One 60 Series REMstar Auto A-Flex w/HT Humid System One 60 Series, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)43 deaths4,149 reports
- DreamStation DreamStation Auto CPAP w/Humid w/Heated Tube, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)30 deaths3,981 reports
- DreamStation DreamStation Auto CPAP w/Humidifier, w/Cell, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)86 deaths3,516 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page