Skip to content

DreamStation DreamStation Heated Humidifier, DERecalls & reported problems

Respironics, Inc. · DreamStation Humidifier, Germany · Class 2

All models of Philips DreamStation (190)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00606959024113), received between Oct 21, 2025 and Nov 25, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Degraded” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Degraded1
  • Overheating of Device1

Patient problems

  • Dehydration1
  • Electric Shock1
  • No Clinical Signs, Symptoms or Conditions1
  • Stroke/CVA1
  • Syncope/Fainting1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 25, 2025Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING A PATIENT EXPERIENCED ELECTRICAL SHOCK FROM HEATED TUBING WHILE USING A DREAMSTATION AUTO CPAP DEVICE. THE PATIENT ALLEGES HE SWITCHED BACK TO THE NORMAL TUBING DAYS LATER AND PASSED OUT AND WAS IN THE HOSPITAL FOR 8 HOURS. THE DOCTORS COULDN'T TELL HIM WHAT HAD TAKEN PLACE; HOWEVER, THE PATIENT CLAIM HE HAD A STROKE. THE PATIENT ALSO ALLEGES DEHYDRATION DURING THE THERAPY, THE HUMIDIFIER SETTING WAS ON LEVEL 5, THE MASK TEST SHOWED 100% AND AHI VALUE WAS LESS THAN 5. THE DEVICE WAS RETURNED TO THE MANUFACTURER'S PRODUCT INVESTIGATION LABORATORY (PIL) FOR INVESTIGATION. THE DEVICE WAS APPLIED POWER AND VERIFIED AIRFLOW. WHILE VERIFYING AIRFLOW PIL OBSERVED THAT THE PRESSURE READING WAS LOW LIKELY DUE TO THE HUMIDIFIER LID SEAL AND DRY BOX SEAL MISSING FROM THE HUMIDIFIER. USING A KNOWN GOOD, HEATED TUBE PIL VERIFIED THAT THE TUBE FUNCTIONED PROPERLY AND DID NOT OBSERVE ANY TYPE OF SHOCK. DURING THE INVESTIGATION, PIL OBSERVED THE HUMIDIFIER LID SEAL, AND THE DRY BOX SEAL WAS NOT RETURNED WITH THE HUMIDIFIER. A DUST LIKE CONTAMINATION ON THE TOP ENCLOSURE, BOTTOM ENCLOSURE, BOTTOM ENCLOSURE OF THE HUMIDIFIER, BOTTOM PANEL, BACK PANEL, B

    Patient problems: Dehydration; Electric Shock; Stroke/CVA; Syncope/Fainting

    MAUDE report 23653085 ↗
  • MalfunctionOct 21, 2025Overheating of Device; Overheating of Device

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THIS REPORT WAS RELATED TO A DREAMSTATION HUMIDIFIER. THE MANUFACTURER RECEIVED INFORMATION FROM THE PATIENT ALLEGING "SLIGHT BURN MARKS" ON THE INSIDE OF THE HUMIDIFIER LID, AND A "HOLE MELTED INTO THE LID SEAL." THE PATIENT ALSO ALLEGES AN ODOR DURING USAGE. NO DEATH, SERIOUS HARM, OR INJURY WAS REPORTED. THE DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP REPORT WILL BE FILED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23348998 ↗

Other VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices