DreamStation DreamStation Auto CPAP w/Humid w/H Tube, AURecalls & reported problems
Respironics, Inc. · DreamStation Auto CPAP with Humidifier, with Heated Tube, Australia · Class 2
All models of Philips DreamStation (190)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00606959026018), received between May 8, 2019 and May 30, 2025 — classified as 4 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Degraded” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Degraded9
- Adverse Event Without Identified Device or Use Problem1
- Contamination1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Abdominal Pain1
- Alteration in Body Temperature1
- Asthma1
- Chest Pain1
- Chronic Obstructive Pulmonary Disease (COPD)1
- Embolism/Embolus1
- Nodule1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 30, 2025Contamination; Degraded; Contamination; Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING VISUALIZATION OF BLACK PARTICLES NEAR THE AIR OUTLET PORT. THIS DEVICE HAS GONE THROUGH THE REMEDIATION PROCESS WHERE BLACK FOAM IS REPAIRED WITH WHITE SILICONE. THIS IS UNLIKELY TO BE FROM FOAM DEGRADATION. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22116653 ↗MalfunctionSep 9, 2021Degraded; Degraded
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE'S SOUND ABATEMENT FOAM BECAME DEGRADED AND CAUSED THE PATIENT TO HAVE SINUSITIS. THE PATIENT DID RECEIVE MEDICAL INTERVENTION IN THE FORM OF A PRESCRIPTION FOR NASAL SPRAY AND SINUS DECONGESTANT. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.
Patient problems: Respiratory Tract Infection
MAUDE report 12449460 ↗MalfunctionSep 8, 2021Degraded; Degraded
THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE'S SOUND ABATEMENT FOAM. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER 88058.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12441068 ↗InjuryAug 24, 2021Degraded; Degraded
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE'S SOUND ABATEMENT FOAM BECAME DEGRADED AND CAUSED A PATIENT TO DEVELOP HEART ATTACK LIKE SYMPTOMS. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.
Patient problems: Asthma; Unspecified Heart Problem
MAUDE report 12359171 ↗MalfunctionAug 16, 2021Degraded; Degraded
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE'S SOUND ABATEMENT FOAM BECAME DEGRADED AND CAUSED THE PATIENT TO HAVE CHEST TIGHTNESS. THE PATIENT DID RECEIVE MEDICAL INTERVENTION IN THE FORM OF A NASAL SPRAY AND AN INHALER. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: Chronic Obstructive Pulmonary Disease (COPD); Sneezing; Sore Throat; Unspecified Respiratory Problem
MAUDE report 12327829 ↗MalfunctionAug 10, 2021Degraded; Degraded
THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE'S SOUND ABATEMENT FOAM. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12301675 ↗MalfunctionJul 7, 2021Degraded; Degraded
THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE'S SOUND ABATEMENT FOAM. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: Abdominal Pain; Alteration in Body Temperature; Chest Pain; No Clinical Signs, Symptoms or Conditions; Shaking/Tremors
MAUDE report 12132687 ↗InjuryJun 30, 2021Degraded; Degraded
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE'S SOUND ABATEMENT FOAM BECAME DEGRADED AND CAUSED A PATIENT TO DEVELOP A BLOOD CLOT IN THEIR LUNG. THE PATIENT DID RECEIVE MEDICAL INTERVENTION IN THE FORM OF A DOCTOR'S ASSESSMENT AND A PRESCRIBED BLOOD THINNER MEDICATION. THE MANUFACTURER IS ATTEMPTING TO HAVE THE DEVICE RETURNED TO ITS PRODUCT INVESTIGATION LAB. THE INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: Embolism/Embolus
MAUDE report 12091806 ↗InjuryJun 25, 2021Degraded; Degraded
THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN END USER IS INVESTIGATING IF A CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE HAS CAUSED A LUMP IN HER THROAT RELATED TO THE SOUND ABATEMENT FOAM. THE DEVICE HAS YET TO BE RETURNED TO THE MANUFACTURER FOR EVALUATION. A FINAL REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: Nodule
MAUDE report 12069909 ↗InjuryMay 8, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED INFORMATION FROM AN END USER ALLEGING WHILE USING A CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE HIS PARTIAL DENTURE PLATE BECAME DISLODGED AND SUBSEQUENTLY REQUIRED SURGERY TO REMOVE THE PARTIAL DENTURE PLATE FROM HIS ESOPHAGUS. THE MANUFACTURER'S INVESTIGATION IS ON-GOING. UPON COMPLETION OF THE INVESTIGATION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: Tissue Damage
MAUDE report 8591988 ↗
Other VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices
- REMstar System One 60 Series REMstar Auto A-Flex, w/Humid System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)117 deaths10,075 reports
- DreamStation DreamStation Auto BiPAP w/Humidifier, w/Cell, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)106 deaths5,937 reports
- DreamStation DreamStation Auto CPAP with Humidifier, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)44 deaths5,793 reports
- REMstar System One 60 Series REMstar Auto A-Flex System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)64 deaths5,138 reports
- REMstar System One 60 Series REMstar Pro C-Flex+, w/Humid System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)65 deaths4,520 reports
- REMstar System One 60 Series REMstar Auto A-Flex w/HT Humid System One 60 Series, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)43 deaths4,149 reports
- DreamStation DreamStation Auto CPAP w/Humid w/Heated Tube, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)30 deaths3,981 reports
- DreamStation DreamStation Auto CPAP w/Humidifier, w/Cell, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)86 deaths3,516 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page