Skip to content

DreamStation DreamStation CPAP w/Humidifier, w/Heated Tube, CARecalls & reported problems

Respironics, Inc. · DreamStation, CPAP with Humidifier, with Heated Tube, Canada · Class 2

All models of Philips DreamStation (190)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00606959026544), received between Oct 12, 2021 and Dec 27, 2024 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Degraded” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Degraded5
  • Temperature Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Cancer1
  • Dry Eye(s)1
  • Epistaxis1
  • Headache1
  • Sore Throat1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 27, 2024Temperature Problem; Temperature Problem

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A DREAMSTATION CPAP DEVICE'S THERMAL EVENT. THE DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR EVALUATION. DURING EVALUATION OF THE DEVICE AT THIRD-PARTY SERVICE CENTER, CONNECTOR WAS BURNT FOUND IN TECHNICAL FINDING. THERE WAS NO SERIOUS PATIENT HARM OR INJURY. THE PATIENT REQUIRED NO MEDICAL INTERVENTION. THE DEVICE WAS SCRAPPED FOLLOWING THE EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21028504 ↗
  • MalfunctionMar 25, 2022Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO THE CPAP DEVICE'S SOUND ABATEMENT FOAM. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13904993 ↗
  • MalfunctionMar 23, 2022Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THERE WAS NO ALLEGATION OF HARM OR INJURY. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13869848 ↗
  • MalfunctionFeb 8, 2022Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT HAS ALLEGED TO PARTICLES IN THE CHAMBER. THE PATIENT HAS ALLEGED TO EXPERIENCING NOSE BLEEDS, HEADACHES, ITCHY EYES AND SORE THROAT. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.

    Patient problems: Dry Eye(s); Epistaxis; Headache; Sore Throat

    MAUDE report 13480250 ↗
  • InjuryDec 22, 2021Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING A CPAP DEVICE'S SOUND ABATEMENT FOAM BECAME DEGRADED AND CAUSED CANCER AND NAUSEA AND VOMITING. THE PATIENT DID REPORT RECEIVING MEDICAL INTERVENTION AND IS RECEIVING CANCER TREATMENT. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.

    Patient problems: Cancer

    MAUDE report 13061026 ↗
  • MalfunctionOct 12, 2021Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12617361 ↗

Other VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices

All VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices →