REMstar System One 60 Series REMstar Plus C-Flex, System One 60 Series, Latin AmericaRecalls & reported problems

Respironics, Inc. · REMstar Plus C-Flex, System One 60 Series, Latin America · Class 2

All models of REMstar System One 60 Series (92)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00606959029910), received between Nov 22, 2024 and Nov 22, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Degraded” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K113068); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Degraded1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 22, 2024Degraded; Degraded

    A REMSTAR PLUS CFLEX WAS RETURNED TO THE MANUFACTURER FOR EVALUATION TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE DEVICE WAS RETURNED TO THE THIRD-PARTY SERVICE CENTER FOR FURTHER EVALUATION. THE DEVICE WAS EVALUATED. THE DEVICE'S DOWNLOADED LOGS WERE REVIEWED BY THE SERVICE CENTER. THERE WAS NO ERROR CODE FOUND. UPON FURTHER EVALUATION, VISIBLE FOAM PARTICLES WERE OBSERVED INSIDE THE BLOWER KIT. IN CONCLUSION, THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE SERVICE CENTER ALSO FOUND EVIDENCE OF DUST PARTICLES, CIGARETTE SMOKE OR INSECT INFESTATIONS AS CONTAMINATION. THE DEVICE WAS SCRAPPED. THERE WAS NO REPORTED PATIENT DEATH OR SERIOUS INJURY, HOWEVER THE REPORTED ISSUE/EVENT IS REPORTABLE DUE TO THE POTENTIAL TO CAUSE OR CONTRIBUTE TO A SERIOUS INJURY AS IDENTIFIED IN CAPA (B)(4), IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE/RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20762232 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K113068. One submission often covers many models, so check the product and lot details in each record.

  • Z-1817-2024 · Apr 1, 2024 · Open, Classified
    OmniLab Advanced + Ventilator Product Numbers 1111122 1111123 1111124 1111125 1111126 1111127 1111138 1111141 1111142 KR1111127 R1111122 R1111123 R1111124 U1111122 U1111124
  • Z-1974-2021 · Jun 14, 2021 · Open, Classified
    SystemOne (Q-Series), DreamStation, DreamStation Go, Dorma 400, Dorma 500, REMstar SE Auto Noncontinuous Ventilators The Philips Respironics E30 ventilator is intended to provide

Other VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices

All VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices →