REMstar System One 60 Series REMstar Plus, with Humidifier, 60 Series, CARecalls & reported problems
Respironics, Inc. · REMstar Plus, with Humidifier, System One 60 Series, Canada · Class 2
All models of REMstar System One 60 Series (92)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Respironics REMstar Plus is a non-continuous ventilator (CPAP-type device) used with a humidifier. The 42 reports mostly describe the internal sound-damping foam breaking down, releasing particles and dust, plus a few burnt power cords and connectors. One injury was reported and no deaths. There are no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 40
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 40 adverse event reports that list this exact device (DI 00606959030060), received between Dec 21, 2021 and Jul 27, 2026 — classified as 1 injury, 39 malfunctions.
The most frequently reported specific device problem is “Degraded” (38 reports).
2 other recalls cover devices cleared under the same FDA submission (K113068); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Degraded38
- Contamination3
- Activation Problem1
- Excessive Heating1
- Thermal Decomposition of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions34
- Cough2
- Dyspnea2
- Sore Throat2
- Cancer1
- Dry Mouth1
- Fatigue1
- Thrombosis/Thrombus1
Recent adverse event reports
The 25 most recent of 40 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 27, 2026Thermal Decomposition of Device
A REMstar Plus device was returned in reference to the voluntary Field Safety Notice / recall notification related to the sound abatement foam in certain CPAP, BiPAP, and mechanical ventilator devices. There was no report of serious patient harm or injury. There was no report of medical intervention. During the evaluation of the device at the third-party service center, the device was visually inspected and found a burned connector. The device was forwarded to the RIQL for further investigation. The manufacturer's investigation is ongoing. A follow-up report will be submitted when the manufacturer's investigation is complete.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50017800 ↗MalfunctionJun 9, 2026Excessive Heating
THE DEVICE WAS SENT TO A THIRD-PARTY SERVICE CENTER FOR INVESTIGATION. DURING EVALUATION, THE THIRD-PARTY SERVICE CENTER FOUND A BURNT POWER CORD. DUE TO THE ALLEGED MALFUNCTION, THE DEVICE WILL BE FORWARDED TO THE MANUFACTURER¿S PRODUCT INVESTIGATION LAB (PIL) FOR ADDITIONAL TESTING AND INVESTIGATION. A FINAL REPORT WILL BE SENT ONCE THE INVESTIGATION IS CONCLUDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25451921 ↗MalfunctionMar 25, 2026Degraded
A DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE, THE THIRD-PARTY SERVICE CENTER VISUALLY INSPECTED THE DEVICE AND FOUND EVIDENCE OF FOAM DEGRADATION AND DUST CONTAMINATION. ADDITIONALLY, THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE HAS BEEN SCRAPPED. AT THIS TIME, NO FURTHER INVESTIGATION CAN BE PERFORMED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP REPORT WILL BE FILED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24688640 ↗MalfunctionMar 10, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A REMSTAR PLUS DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF VISIBLE FOAM PARTICLES WAS FOUND INSIDE THE BLOWER KIT. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST FAIL AND FLUIDS FAIL CONTAMINATION. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24562635 ↗MalfunctionMar 10, 2026Contamination; Degraded
A DEVICE WAS RETURNED TO THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM DEGRADATION AND DUST/DIRT WAS OBSERVED. THE DEVICE POWERED ON AND AIRFLOW WAS CONFIRMED. THE DEVICE'S DOWNLOADED LOGS WERE REVIEWED BY THE THIRD-PARTY SERVICE CENTER. THERE WERE ERROR CODES FOUND. THE THIRD-PARTY SERVICE CENTER CONCLUDES THAT THEY CONFIRM THE CUSTOMER'S ALLEGATION AND THERE WAS A VISIBLE FOAM DEGRADATION AND UNIT GOT CORRECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24555398 ↗MalfunctionMar 10, 2026Contamination; Degraded
A DEVICE WAS RETURNED TO THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM DEGRADATION AND DUST CONTAMINATION WERE OBSERVED. THE DEVICE POWERED ON AND AIRFLOW WAS CONFIRMED. THE DEVICE'S DOWNLOADED LOGS WERE REVIEWED BY THE THIRD-PARTY SERVICE CENTER. THERE WERE NO ERROR CODES FOUND. THE THIRD-PARTY SERVICE CENTER CONCLUDES THAT THEY CONFIRM THE CUSTOMER'S ALLEGATION AND THERE WAS A VISIBLE FOAM DEGRADATION AND UNIT GOT CORRECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24554702 ↗MalfunctionMar 6, 2026Degraded
A REMSTAR PLUS DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR SERVICE AS PART OF THE SOUND ABATEMENT FOAM RECALL PROCESS DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES INSIDE THE BLOWER KIT AND A BLOWER FOAM DEGRADATION. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST FAIL CONTAMINATION. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24532868 ↗MalfunctionMar 4, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A REMSTAR PLUS DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER THE DEVICE WAS VISUALLY INSPECTED AND EVIDENCE OF FOAM DEGRADATION WAS FOUND INSIDE THE BLOWER KIT. MOREOVER, THE DEVICE HAD DUST CONTAMINATION. ADDITIONALLY, THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. IN ADDITION, THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24512001 ↗MalfunctionFeb 13, 2026Degraded
A REMSTAR PLUS DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR SERVICE AS PART OF THE SOUND ABATEMENT FOAM RECALL PROCESS WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES INSIDE THE BLOWER KIT AND OBJECT CODE INDICATES THE BLOWER CONTRIBUTED TO THE FOAM DEGRADATION. ADDITIONALLY, THE SERVICE TECHNICIAN NOTED DUST CONTAMINATION. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24334293 ↗MalfunctionFeb 3, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A REMSTAR PLUS, DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION/PARTICLES WAS FOUND INSIDE THE BLOWER KIT. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED FLUID CONTAMINATION. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24246956 ↗MalfunctionFeb 2, 2026Degraded
A REMSTAR PLUS DEVICE WAS RETURNED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED CONTAMINATION. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24240221 ↗MalfunctionJan 27, 2026Degraded
A REMSTAR PLUS DEVICE WAS RETURNED TO THE THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM DEGRADATION. DURING THE EVALUATION, FOAM PARTICLES WERE OBSERVED, AND DUST CONTAMINATION WAS FOUND IN THE DEVICE. ADDITIONALLY, THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24195189 ↗MalfunctionJan 21, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A REMSTAR PLUS DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION/PARTICLES WAS FOUND INSIDE THE BLOWER KIT. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED FLUIDS FAIL CONTAMINATION. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24132127 ↗MalfunctionJan 19, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A REMSTAR PLUS DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION/PARTICLES WAS FOUND INSIDE THE BLOWER KIT. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED FLUID CONTAMINATION. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24111610 ↗MalfunctionJan 13, 2026Degraded
A REMSTAR PLUS DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER AS PART OF THE UNO RECALL PROCESS WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES INSIDE THE BLOWER KIT. ADDITIONALLY, THE SERVICE TECHNICIAN NOTED DUST FAIL CONTAMINATION. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24058584 ↗MalfunctionJan 12, 2026Degraded
A REMSTAR PLUS DEVICE WAS RETURNED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED CONTAMINATION. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24039733 ↗MalfunctionJan 5, 2026Degraded
A REMSTAR PLUS DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER AS PART OF THE UNO RECALL PROCESS WITH NO INITIAL ALLEGED COMPLAINT. DURING EVALUATION NO FOAM PARTICLES WERE NOTED IN THE AIRPATH, YET FOAM DEGRADATION WAS NOTED. THE DEVICES SOUND ABATEMENT FOAM (NEEDS) OR (WAS) REPLACED TO ADDRESS THE ISSUE. ALSO, A DUST CONTAMINATION WAS PRESENT ON THE DEVICE. THERE WAS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23956094 ↗MalfunctionJan 3, 2026Degraded
A REMSTAR PLUS DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER AS PART OF THE UNO RECALL PROCESS WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED 32 ERROR CODES PRESENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23952956 ↗MalfunctionDec 31, 2025Degraded; Degraded
A REMSTAR PLUS DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER AS PART OF THE UNO RECALL PROCESS WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES INSIDE THE BLOWER KIT. ADDITIONALLY, THE SERVICE TECHNICIAN NOTED DUST CONTAMINATION. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED. THIS EVENT IS REPORTABLE UNDER FDA 21 CFR PART 803 AS IT MEETS THE CRITERIA FOR A SERIOUS INJURY AND/OR PRODUCT MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23936676 ↗MalfunctionDec 31, 2025Degraded; Degraded
A REMSTAR PLUS W/HUM DEVICE WAS RETURNED TO A THIRD PARTY SERVICE CENTER FOR SERVICE AS PART OF THE SOUND ABATEMENT FOAM RECALL PROCESS WITH NO INITIAL ALLEGED COMPLAINT. DURING EVALUATION NO FOAM PARTICLES WERE NOTED IN THE AIRPATH, YET FOAM DEGRADATION WAS NOTED. THE DEVICES SOUND ABATEMENT FOAM WAS REPLACED TO ADDRESS THE ISSUE. IN ADDITION, THE SERVICE CENTER FOUND CONTAMINATION FROM DUST FAIL, THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED. THIS EVENT IS REPORTABLE UNDER FDA 21 CFR PART 803 AS IT MEETS THE CRITERIA FOR A SERIOUS INJURY AND/OR PRODUCT MALFUNCTION
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23936078 ↗MalfunctionDec 30, 2025Degraded; Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE/RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A REMSTAR PLUS DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR EVALUATION OF THE DEVICE WITH NO INITIAL ALLEGED COMPLAINT. THERE WAS NO DEATH, SERIOUS HARM, OR INJURY REPORTED, AND NO MEDICAL INTERVENTION WAS PROVIDED. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION/PARTICLES WAS FOUND INSIDE THE BLOWER KIT. IN ADDITION, THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23927635 ↗MalfunctionSep 8, 2025Degraded; Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING VISUALIZATION OF FOAM PARTICLES INSIDE THE ASSEMBLY. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS SPECIFIED AS REQUIRED BY THE PATIENT. DURING THE EVALUATION OF THE DEVICE, AT THE MANUFACTURE'S THIRD-PARTY'S SERVICE CENTER THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM PARTICLES INSIDE ASSEMBLY. THE DEVICE HAS BEEN SCRAPPED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22993407 ↗MalfunctionSep 8, 2025Degraded; Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING VISUALIZATION OF FOAM PARTICLES INSIDE THE ASSEMBLY. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS SPECIFIED AS REQUIRED BY THE PATIENT. DURING THE EVALUATION OF THE DEVICE, AT THE MANUFACTURE'S THIRD-PARTY'S SERVICE CENTER THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM PARTICLES INSIDE ASSEMBLY. THE DEVICE HAS BEEN SCRAPPED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22992449 ↗MalfunctionSep 4, 2025Degraded; Degraded
A REMSTAR DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER AS PART OF THE UNO RECALL PROCESS WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED AN ERROR CODE PRESENT, SD CARD FOUND, AND SERVICE IS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22973034 ↗MalfunctionJul 25, 2025Degraded; Degraded
DURING THE EVALUATION OF THE DEVICE, THE THIRD-PARTY SERVICE CENTER VISUALLY INSPECTED THE DEVICE AND FOUND EVIDENCE OF FOAM DEGRADATION AND ERROR CODES 53, 62 AND 63 WERE LOGGED. THE DEVICE HAS BEEN SCRAPPED. AT THIS TIME, NO FURTHER INVESTIGATION CAN BE PERFORMED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP REPORT WILL BE FILED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22608984 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K113068. One submission often covers many models, so check the product and lot details in each record.
- Z-1817-2024 · Apr 1, 2024 · Open, ClassifiedOmniLab Advanced + Ventilator Product Numbers 1111122 1111123 1111124 1111125 1111126 1111127 1111138 1111141 1111142 KR1111127 R1111122 R1111123 R1111124 U1111122 U1111124
- Z-1974-2021 · Jun 14, 2021 · Open, ClassifiedSystemOne (Q-Series), DreamStation, DreamStation Go, Dorma 400, Dorma 500, REMstar SE Auto Noncontinuous Ventilators The Philips Respironics E30 ventilator is intended to provide
Other VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices
- REMstar System One 60 Series REMstar Auto A-Flex, w/Humid System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)117 deaths10,075 reports
- DreamStation DreamStation Auto BiPAP w/Humidifier, w/Cell, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)106 deaths5,937 reports
- DreamStation DreamStation Auto CPAP with Humidifier, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)44 deaths5,793 reports
- REMstar System One 60 Series REMstar Auto A-Flex System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)64 deaths5,138 reports
- REMstar System One 60 Series REMstar Pro C-Flex+, w/Humid System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)65 deaths4,520 reports
- REMstar System One 60 Series REMstar Auto A-Flex w/HT Humid System One 60 Series, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)43 deaths4,149 reports
- DreamStation DreamStation Auto CPAP w/Humid w/Heated Tube, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)30 deaths3,981 reports
- DreamStation DreamStation Auto CPAP w/Humidifier, w/Cell, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)86 deaths3,516 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page