Skip to content

REMstar System One 60 Series REMstar Auto A-Flex, System One 60 Series, GBRecalls & reported problems

Respironics, Inc. · REMstar Auto A-Flex, System One 60 Series, GB · Class 2

All models of REMstar System One 60 Series (92)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Respironics REMstar Auto A-Flex is a non-invasive ventilator for sleep breathing support. Reports mostly describe sound-abatement foam deteriorating or leaving particles inside the device. There are 40 reports, including 2 deaths, 1 injury and 37 malfunctions. No recalls name this device. Reports in the last 12 months (23) far exceeded the prior year (4).

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
37
Deaths reported
2
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 37 adverse event reports that list this exact device (DI 00606959030589), received between Feb 3, 2020 and Mar 26, 2026 — classified as 2 deaths, 1 injury, 34 malfunctions.

The most frequently reported specific device problem is “Degraded” (37 reports).

2 other recalls cover devices cleared under the same FDA submission (K113068); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 3 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 20 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202620
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Degraded37
  • Loose or Intermittent Connection1
  • Thermal Decomposition of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions32
  • Cough3
  • Cancer2
  • Headache1
  • No Patient Involvement1
  • Skin Inflammation/ Irritation1
  • Unspecified Respiratory Problem1

Recent adverse event reports

The 25 most recent of 37 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 26, 2026Degraded

    A REMSTAR AUTO A-FLEX DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOAM PARTICLES WERE OBSERVED INSIDE THE ASSEMBLY. IN ADDITION, INSECT INFESTATION, TOBACCO CONTAMINATION AND DUST/DIRT WAS FOUND INSIDE THE DEVICE. ADDITIONALLY, THE DEVICE WAS SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24695970 ↗
  • MalfunctionMar 25, 2026Degraded

    A REMSTAR AUTO A-FLEX DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND OBSERVED DETERIORATED FOAM. ADDITIONALLY, THE DEVICE WAS SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24688404 ↗
  • MalfunctionMar 25, 2026Degraded

    A REMSTAR AUTO A-FLEX DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND OBSERVED DETERIORATED FOAM. ADDITIONALLY, THE DEVICE WAS SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24688395 ↗
  • MalfunctionMar 25, 2026Degraded

    A REMSTAR AUTO A-FLEX DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOAM PARTICLES WERE OBSERVED INSIDE THE ASSEMBLY. ADDITIONALLY, THE DEVICE WAS SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24688383 ↗
  • MalfunctionMar 25, 2026Degraded

    A REMSTAR AUTO A-FLEX DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND OBSERVED DETERIORATED FOAM. ADDITIONALLY, DUST/DIRT WAS FOUND INSIDE THE DEVICE, AND THE DEVICE WAS SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24688376 ↗
  • MalfunctionMar 25, 2026Degraded

    A REMSTAR AUTO A-FLEX DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND OBSERVED DETERIORATED FOAM. ADDITIONALLY, THE DEVICE WAS SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24688360 ↗
  • MalfunctionMar 25, 2026Degraded

    A REMSTAR AUTO A-FLEX DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND OBSERVED DETERIORATED FOAM. ADDITIONALLY, THE DEVICE WAS SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24688359 ↗
  • MalfunctionMar 25, 2026Degraded

    A REMSTAR AUTO A-FLEX DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND OBSERVED DETERIORATED FOAM. ADDITIONALLY, DUST/DIRT WAS FOUND INSIDE THE DEVICE, AND THE DEVICE WAS SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24688353 ↗
  • MalfunctionMar 25, 2026Degraded

    A REMSTAR AUTO A-FLEX DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND OBSERVED DAMAGED FOAM THAT CANNOT BE DISASSEMBLED FROM THE CHASSIS. IN ADDITION, FOAM PARTICLES WERE OBSERVED INSIDE THE ASSEMBLY AND DUST/DIRT WAS FOUND INSIDE THE DEVICE. ADDITIONALLY, THE DEVICE WAS SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24688349 ↗
  • MalfunctionMar 23, 2026Degraded

    A REMSTAR AUTO A-FLEX DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND OBSERVED DETERIORATED FOAM. ADDITIONALLY, DUST/DIRT WAS FOUND INSIDE THE DEVICE, AND THE DEVICE WAS SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24663480 ↗
  • MalfunctionMar 23, 2026Degraded

    A REMSTAR AUTO A-FLEX DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOAM PARTICLES WERE OBSERVED INSIDE THE ASSEMBLY. ADDITIONALLY, DUST/DIRT WAS FOUND INSIDE THE DEVICE, AND THE DEVICE WAS SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24663478 ↗
  • MalfunctionMar 23, 2026Degraded

    A REMSTAR AUTO A-FLEX DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND OBSERVED DETERIORATED FOAM. ADDITIONALLY, DUST/DIRT WAS FOUND INSIDE THE DEVICE, AND THE DEVICE WAS SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24663477 ↗
  • MalfunctionMar 23, 2026Degraded

    A REMSTAR AUTO A-FLEX DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND OBSERVED DETERIORATED FOAM. ADDITIONALLY, THE DEVICE WAS SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24663475 ↗
  • MalfunctionMar 23, 2026Degraded

    A REMSTAR AUTO A-FLEX DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND OBSERVED DAMAGED FOAM THAT CANNOT BE DISASSEMBLED FROM THE CHASSIS. ADDITIONALLY, THE PATIENT¿S COMPLAINT WAS NOT CONFIRMED, AND THE DEVICE WAS SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24663474 ↗
  • MalfunctionMar 23, 2026Degraded

    A REMSTAR AUTO A-FLEX DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND OBSERVED DAMAGED FOAM THAT CANNOT BE DISASSEMBLED FROM THE CHASSIS. ADDITIONALLY, DUST/DIRT WAS FOUND INSIDE THE DEVICE, AND THE DEVICE WAS SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24663472 ↗
  • MalfunctionMar 23, 2026Degraded

    A REMSTAR AUTO A-FLEX DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND OBSERVED DETERIORATED FOAM. ADDITIONALLY, THE DEVICE WAS SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24663471 ↗
  • MalfunctionMar 23, 2026Degraded

    A REMSTAR AUTO A-FLEX DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND OBSERVED DETERIORATED FOAM. IN ADDITION, FOAM PARTICLES WERE OBSERVED INSIDE THE ASSEMBLY AND DUST/DIRT WAS FOUND INSIDE THE DEVICE. ADDITIONALLY, THE DEVICE WAS SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24663470 ↗
  • MalfunctionMar 23, 2026Degraded

    A REMSTAR AUTO A-FLEX DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND OBSERVED DAMAGED FOAM THAT CANNOT BE DISASSEMBLED FROM THE CHASSIS. ADDITIONALLY, DUST/DIRT WAS FOUND INSIDE THE DEVICE, AND THE DEVICE WAS SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24663468 ↗
  • MalfunctionMar 23, 2026Degraded

    A REMSTAR AUTO A-FLEX DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND OBSERVED DAMAGED FOAM THAT CANNOT BE DISASSEMBLED FROM THE CHASSIS. IN ADDITION, FOAM PARTICLES WERE OBSERVED INSIDE THE ASSEMBLY AND DUST/DIRT WAS FOUND INSIDE THE DEVICE. ADDITIONALLY, THE DEVICE WAS SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24663462 ↗
  • MalfunctionMar 23, 2026Degraded

    A REMSTAR AUTO A-FLEX DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOAM PARTICLES WERE OBSERVED INSIDE THE ASSEMBLY. IN ADDITION, INSECT INFESTATION, TOBACCO CONTAMINATION AND DUST/DIRT WAS FOUND INSIDE THE DEVICE. ADDITIONALLY, THE DEVICE WAS SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24663458 ↗
  • MalfunctionJul 30, 2025Degraded; Degraded

    A DEVICE WAS RETURNED TO A THIRD PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED/SCRAPPED AND FOUND EVIDENCE OF FOAM DEGRADATION. DURING THE EVALUATION, FOAM PARTICLES WERE VISIBLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22670302 ↗
  • MalfunctionJul 21, 2025Degraded; Degraded

    A DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED/SCRAPPED AND FOUND EVIDENCE OF FOAM DEGRADATION. DURING THE EVALUATION, FOAM PARTICLES WERE VISIBLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22555667 ↗
  • MalfunctionJul 21, 2025Degraded; Degraded

    A DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED/SCRAPPED AND FOUND EVIDENCE OF FOAM DEGRADATION. DURING THE EVALUATION, FOAM PARTICLES WERE VISIBLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22555375 ↗
  • DeathMay 9, 2025Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING DEATH OF THE PATIENT DUE TO CANCER. THERE WAS A REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS SPECIFIED BY THE PATIENT. THE DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. THE MANUFACTURER WAS MADE AWARE OF THIS COMPLAINT THROUGH A REPRESENTATIVE OF THE CUSTOMER. AT THIS TIME, NO FURTHER INVESTIGATION CAN BE PERFORMED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP REPORT WILL BE FILED.

    Patient problems: Cancer

    MAUDE report 21992565 ↗
  • InjurySep 26, 2022Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM BECAME DEGRADED AND CAUSED ESOPHAGEAL CANCER. THE PATIENT DID NOT REPORT TO RECEIVE MEDICAL INTERVENTION. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: Cancer

    MAUDE report 15486557 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K113068. One submission often covers many models, so check the product and lot details in each record.

  • Z-1817-2024 · Apr 1, 2024 · Open, Classified
    OmniLab Advanced + Ventilator Product Numbers 1111122 1111123 1111124 1111125 1111126 1111127 1111138 1111141 1111142 KR1111127 R1111122 R1111123 R1111124 U1111122 U1111124
  • Z-1974-2021 · Jun 14, 2021 · Open, Classified
    SystemOne (Q-Series), DreamStation, DreamStation Go, Dorma 400, Dorma 500, REMstar SE Auto Noncontinuous Ventilators The Philips Respironics E30 ventilator is intended to provide

Other VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices

All VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page