REMstar System One 60 Series REMstar Auto A-Flex W/SD Card JPRecalls & reported problems
Respironics, Inc. · REMstar Auto A-Flex with SD Card, Japan · Class 2
All models of REMstar System One 60 Series (92)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00606959030923), received between Feb 28, 2022 and Jul 9, 2025 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Degraded” (10 reports).
2 other recalls cover devices cleared under the same FDA submission (K113068); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Degraded10
- Melted1
- Overheating of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 9, 2025Degraded; Melted; Overheating of Device; Degraded; Melted; Overheating of Device
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP/ BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED ALLEGATION OF UI KNOB BROKEN AND THE OUTER CASE MELTED. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THE DEVICE WAS RETURNED TO THE MANUFACTURER'S PRODUCT INVESTIGATION LABORATORY FOR INVESTIGATION. AN EXTERNAL AND INTERNAL VISUAL INSPECTION OF THE DEVICE WAS COMPLETED BY THE MANUFACTURER AND FOUND DUST-LIKE CONTAMINANT, FOAM PARTICLES IN THE BOTTOM ENCLOSURE AND BLOWER ASSEMBLY. THE MANUFACTURER CONFIRMED EVIDENCE OF SOUND ABATEMENT FOAM DEGRADATION/BREAKDOWN IN THE RETURNED MATERIALS. THE MANUFACTURER ALSO CONFIRMED THERMAL FAILURE OF PCA. THE MANUFACTURER CONCLUDES THERE WAS EVIDENCE OF SOUND ABATEMENT FOAM DEGRADATION AND THERMAL FAILURE OF THE DEVICE. THE MANUFACTURER IS SUBMITTING A FINAL REPORT AT THIS TIME. IF PERTINENT INFORMATION BECOMES AVAILABLE TO THE MANUFACTURER AT A LATER DATE, AN ADDENDUM TO THIS FINAL REPORT WILL BE FILED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22459672 ↗MalfunctionApr 1, 2025Degraded; Degraded
THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A REMSTAR AUTO A-FLEX DEVICE'S SOUND ABATEMENT FOAM. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND VISIBLE FOAM WERE OBSERVED. ALSO FOUND LIFE RELATED SMELL AND E--94 OCCURRED. AT THIS TIME, NO FURTHER INVESTIGATION CAN BE PERFORMED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP REPORT WILL BE FILED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21741332 ↗MalfunctionJul 6, 2022Degraded; Degraded
THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE DEVICE WAS RETURNED AND OBSERVE BREAKAGE TO THE UI KNOB DURING DEVICE EVALUATION . THERE WAS NO ALLEGATION OF HARM OR INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14935371 ↗MalfunctionJul 5, 2022Degraded; Degraded
THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE DEVICE WAS RETURNED AND OBSERVE BREAKAGE TO THE UI KNOB DURING DEVICE EVALUATION. THERE WAS NO ALLEGATION OF HARM OR INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14918575 ↗MalfunctionMar 31, 2022Degraded; Degraded
THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO REMSTAR DEVICE'S SOUND ABATEMENT FOAM. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13969066 ↗MalfunctionMar 15, 2022Degraded; Degraded
THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. DURING DEVICE EVALUATION CONTAMINATION IN THE BLOWER FOAM AND BLOWER BOX WAS FOUND. NO ALLEGATION OF PATIENT HARM OR INJURY WAS REPORTED. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13773123 ↗MalfunctionMar 2, 2022Degraded; Degraded
THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO THE CPAP DEVICE'S SOUND ABATEMENT FOAM. DURING THE EVALUATION AT THE MANUFACTURER'S SERVICE CENTER, THE SOUND ABATEMENT FOAM WAS REPLACED FOR REPAIR. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13653841 ↗MalfunctionMar 2, 2022Degraded; Degraded
THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO THE CPAP DEVICE'S SOUND ABATEMENT FOAM. DURING THE EVALUATION AT THE MANUFACTURER'S SERVICE CENTER, THE BLOWER WAS REPLACED FOR REPAIR. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13653795 ↗MalfunctionFeb 28, 2022Degraded; Degraded
THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM.THERE WAS NO REPORT OF PATIENT HARM OR INJURY.THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13625147 ↗MalfunctionFeb 28, 2022Degraded; Degraded
THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13625053 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K113068. One submission often covers many models, so check the product and lot details in each record.
- Z-1817-2024 · Apr 1, 2024 · Open, ClassifiedOmniLab Advanced + Ventilator Product Numbers 1111122 1111123 1111124 1111125 1111126 1111127 1111138 1111141 1111142 KR1111127 R1111122 R1111123 R1111124 U1111122 U1111124
- Z-1974-2021 · Jun 14, 2021 · Open, ClassifiedSystemOne (Q-Series), DreamStation, DreamStation Go, Dorma 400, Dorma 500, REMstar SE Auto Noncontinuous Ventilators The Philips Respironics E30 ventilator is intended to provide
Other VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices
- REMstar System One 60 Series REMstar Auto A-Flex, w/Humid System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)117 deaths10,075 reports
- DreamStation DreamStation Auto BiPAP w/Humidifier, w/Cell, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)106 deaths5,937 reports
- DreamStation DreamStation Auto CPAP with Humidifier, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)44 deaths5,793 reports
- REMstar System One 60 Series REMstar Auto A-Flex System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)64 deaths5,138 reports
- REMstar System One 60 Series REMstar Pro C-Flex+, w/Humid System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)65 deaths4,520 reports
- REMstar System One 60 Series REMstar Auto A-Flex w/HT Humid System One 60 Series, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)43 deaths4,149 reports
- DreamStation DreamStation Auto CPAP w/Humid w/Heated Tube, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)30 deaths3,981 reports
- DreamStation DreamStation Auto CPAP w/Humidifier, w/Cell, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)86 deaths3,516 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page