UltraFill ULTRAFILL, USRecalls & reported problems
Respironics, Inc. · ULTRAFILL, US · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00606959032477), received between Sep 27, 2017 and Feb 14, 2018 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Device Operational Issue” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Operational Issue1
- Disconnection1
- Loose or Intermittent Connection1
Patient problems
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
- No Patient Involvement1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 14, 2018Disconnection; Loose or Intermittent Connection; Loose or Intermittent Connection
THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN OXYGEN CYLINDER BECAME DISCONNECTED FROM THE ULTRAFILL DEVICE. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THE INVESTIGATION IS STILL ONGOING. A FOLLOW UP REPORT WILL BE FILED WHEN THE MANUFACTURER HAS COMPLETED THE INVESTIGATION.
Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement
MAUDE report 7270684 ↗MalfunctionSep 27, 2017Device Operational Issue; Device Operational Issue
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A CYLINDER SEPARATED FROM THE ULTRAFILL DEVICE. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THE UTLRAFILL DEVICE AND CYLINDER WERE RETURNED TO THE MANUFACTURER FOR EVALUATION. THE ULTRAFILL DEVICE WAS TESTED AND WAS FOUND TO OPERATE TO DESIGN SPECIFICATIONS. THE CYLINDER WAS TESTED AND THE FILL FITTING HAS SEPARATED FROM THE POST VALVE. THIS OCCURS WHEN EXCESSIVE FORCE HAS BEEN APPLIED TO TIGHTEN THE FITTING TO THE POINT THE BRASS HAS FRACTURED BUT NOT FULLY SEPARATED FROM THE POST VALVE. VISUAL INSPECTION OF THE CYLINDER WAS COMPLETED BY THE MANUFACTURER. THE MANUFACTURER FOUND EVIDENCE OF MARKINGS CAUSED BY A TOOL USED FOR OVERTIGHTENING OF THE FITTING. THIS OCCURRED IN THE FIELD, NOT DURING MANUFACTURING.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6898008 ↗
Other Generator, oxygen, portable devices
- DeVilbiss 525DSDevilbiss Healthcare LLC · Generator, Oxygen, Portable2 recalls14 deaths287 reports
- Inogen One G5 Oxygen Concentrator - Bluetooth 1260 BluetoothInogen, Inc. · Generator, Oxygen, Portable3 deaths235 reports
- EverFlo EverFlo w/O2 Percentage Indicator, 230V, EURespironics, Inc. · Generator, oxygen, portable5 deaths112 reports
- Inogen One G3 Oxygen Concentrator 1050Inogen, Inc. · Generator, Oxygen, Portable2 deaths111 reports
- DeVilbiss 125DDevilbiss Healthcare LLC · Generator, Oxygen, Portable2 recalls1 death3 reports
- EverFlo EVERFLO NON-O2 Percentage Indicator, 120V, US,CARespironics, Inc. · Generator, oxygen, portable19 deaths80 reports
- EverFlo EVERFLO w/Oxygen Percentage Indicator, 120V US, CARespironics, Inc. · Generator, oxygen, portable2 deaths71 reports
- SimplyGo Mini SimplyGo Mini, USRespironics, Inc. · Generator, oxygen, portable1 recall1 death5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page