Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00606959032590), received between Nov 15, 2023 and Apr 25, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Gas Output Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Gas Output Problem1
  • Insufficient Device Problem Information1
  • Leak/Splash1
  • Overheating of Device1
  • Pressure Problem1
  • Smoking1
  • Thermal Decomposition of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 25, 2025Smoking; Smoking

    THE MANUFACTURER WAS CONTACTED ON EVERFLO DEVICE THAT CAME BACK FROM REPAIRS A WEEK AGO, BUT IT SMELLS LIKE SMOKE WHEN IT IS IN OPERATION. NO MEDICAL INTERVENTION WAS SPECIFIED AS REQUIRED BY THE PATIENT. THE DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A SUPPLEMENTAL FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21908266 ↗
  • MalfunctionJan 21, 2025Thermal Decomposition of Device; Gas Output Problem; Leak/Splash; Pressure Problem; Insufficient Device Problem Information; Thermal Decomposition of Device; Gas Output Problem; Leak/Splash; Pressure Problem; Insufficient Device Problem Information

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING NOT COMPLIANCE PRESSURE. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND COMPRESSOR BURN. ADDITIONALLY, FLOWMETER IS CRACK, SIEVES IS CLOGGED, O2 BODY AND VALVES IS DEFECTIVE. THE CUSTOMER COMPLAINT WAS REPRODUCED. THERE WAS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS SPECIFIED BY THE PATIENT. THE DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. THE MANUFACTURER WAS MADE AWARE OF THIS COMPLAINT THROUGH A REPRESENTATIVE OF THE CUSTOMER. AT THIS TIME, NO FURTHER INVESTIGATION CAN BE PERFORMED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP REPORT WILL BE FILED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21201026 ↗
  • MalfunctionNov 15, 2023Overheating of Device; Overheating of Device

    THE MANUFACTURER RECEIVED AN ALLEGATION THAT THE BODY WORK OF THE EVERFLO DEVICE IS DEFORMED DUE TO OVERHEATING. NO ALLEGATIONS OF PATIENT HARM, SERIOUS INJURY, OR MEDICAL INTERVENTION ARE SPECIFIED. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18145883 ↗

Other Generator, oxygen, portable devices

All Generator, oxygen, portable devices →

Sources and updates