EverFlo EverFlo w/Oxygen Percentage Indicator, 230V, IT, CLRecalls & reported problems
Respironics, Inc. · EverFlo w/Oxygen Percentage Indicator, 230V Italy Chile · Class 2
All models of EverFlo (20)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00606959032590), received between Nov 15, 2023 and Apr 25, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Gas Output Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Gas Output Problem1
- Insufficient Device Problem Information1
- Leak/Splash1
- Overheating of Device1
- Pressure Problem1
- Smoking1
- Thermal Decomposition of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 25, 2025Smoking; Smoking
THE MANUFACTURER WAS CONTACTED ON EVERFLO DEVICE THAT CAME BACK FROM REPAIRS A WEEK AGO, BUT IT SMELLS LIKE SMOKE WHEN IT IS IN OPERATION. NO MEDICAL INTERVENTION WAS SPECIFIED AS REQUIRED BY THE PATIENT. THE DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A SUPPLEMENTAL FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21908266 ↗MalfunctionJan 21, 2025Thermal Decomposition of Device; Gas Output Problem; Leak/Splash; Pressure Problem; Insufficient Device Problem Information; Thermal Decomposition of Device; Gas Output Problem; Leak/Splash; Pressure Problem; Insufficient Device Problem Information
THE MANUFACTURER RECEIVED INFORMATION ALLEGING NOT COMPLIANCE PRESSURE. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND COMPRESSOR BURN. ADDITIONALLY, FLOWMETER IS CRACK, SIEVES IS CLOGGED, O2 BODY AND VALVES IS DEFECTIVE. THE CUSTOMER COMPLAINT WAS REPRODUCED. THERE WAS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS SPECIFIED BY THE PATIENT. THE DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. THE MANUFACTURER WAS MADE AWARE OF THIS COMPLAINT THROUGH A REPRESENTATIVE OF THE CUSTOMER. AT THIS TIME, NO FURTHER INVESTIGATION CAN BE PERFORMED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP REPORT WILL BE FILED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21201026 ↗MalfunctionNov 15, 2023Overheating of Device; Overheating of Device
THE MANUFACTURER RECEIVED AN ALLEGATION THAT THE BODY WORK OF THE EVERFLO DEVICE IS DEFORMED DUE TO OVERHEATING. NO ALLEGATIONS OF PATIENT HARM, SERIOUS INJURY, OR MEDICAL INTERVENTION ARE SPECIFIED. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18145883 ↗
Other Generator, oxygen, portable devices
- DeVilbiss 525DSDevilbiss Healthcare LLC · Generator, Oxygen, Portable2 recalls14 deaths287 reports
- Inogen One G5 Oxygen Concentrator - Bluetooth 1260 BluetoothInogen, Inc. · Generator, Oxygen, Portable3 deaths235 reports
- EverFlo EverFlo w/O2 Percentage Indicator, 230V, EURespironics, Inc. · Generator, oxygen, portable5 deaths112 reports
- Inogen One G3 Oxygen Concentrator 1050Inogen, Inc. · Generator, Oxygen, Portable2 deaths111 reports
- DeVilbiss 125DDevilbiss Healthcare LLC · Generator, Oxygen, Portable2 recalls1 death3 reports
- EverFlo EVERFLO NON-O2 Percentage Indicator, 120V, US,CARespironics, Inc. · Generator, oxygen, portable19 deaths80 reports
- EverFlo EVERFLO w/Oxygen Percentage Indicator, 120V US, CARespironics, Inc. · Generator, oxygen, portable2 deaths71 reports
- SimplyGo Mini SimplyGo Mini, USRespironics, Inc. · Generator, oxygen, portable1 recall1 death5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗