EverFlo EVERFLO QUIET w/O2 Percent Indicator, 120V, US, CARecalls & reported problems
Respironics, Inc. · EVERFLO QUIET w/O2 Percent Indicator, 120V, US, Canada · Class 2
All models of EverFlo (20)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00606959032620), received between Nov 21, 2017 and Mar 3, 2026 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Failure to Deliver” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Deliver2
- Mechanical Problem2
- Adverse Event Without Identified Device or Use Problem1
- Crack1
- Device Alarm System1
- Electrical /Electronic Property Problem1
- Fire1
- Gas/Air Leak1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Dyspnea1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryMar 3, 2026No Audible Alarm; Failure to Deliver
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A PATIENT EXPERIENCED SEVERE SHORTNESS OF BREATH WHILE USING AN EVERFLO DEVICE. THE PATIENT'S SPOUSE CALLED THE EMT WHO CHECKED THE DEVICE. THE DEVICE HAD NO FLOW COMING FROM THE OUTLET. THE DEVICE SETTINGS DURING THE EVENT WERE REPORTED TO BE 4.5-5L AND A GREEN LIGHT WAS SHOWING. THERE WERE NO AUDIBLE OR VISUAL ALARMS AT THE TIME. THE PATIENT IS CURRENTLY HOSPITALIZED. THE DEVICE WAS EVALUATED BY THE SERVICE TECHNICIAN. DURING THE EVALUATION IT WAS FOUND THAT THE NIPPLE ON THE FINAL BACTERIAL FILTER (BETWEEN THE FLOW METER AND PATIENT OUTLET) HAD SNAPPED IN HALF, THUS DELIVERING THE OXYGEN TO THE INSIDE OF THE MACHINE AND NOT THE PATIENT, HOWEVER, NO ALARM CONDITIONS OCCURRED. THE DEVICE HAS BEEN FORWARDED TO THE MANUFACTURER FOR FURTHER INVESTIGATION. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP REPORT WILL BE FILED.
Patient problems: Dyspnea
MAUDE report 24503294 ↗MalfunctionJun 6, 2024Electrical /Electronic Property Problem; Mechanical Problem; Electrical /Electronic Property Problem; Mechanical Problem
THE MANUFACTURER BECAME AWARE OF AN ALLEGATION OF EVERFLO QUIET THAT THE PATIENT ALLEGING FAULTY PARTS. A DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER. DURING THE EVALUATION OF THE DEVICE, THEY FOUND OUT THAT THE COMPLAINT WAS CONFIRMED, THE SIEVE WAS LEAKING, THE FLOW METER WAS CRACKED, CONCENTRATOR TUBING WAS BRITTLE, COMPRESSOR WAS FAILING, POWER CORD WAS SPLIT OPEN AND FAULTY PARTS WERE REPLACED. AT THIS TIME, NO FURTHER INVESTIGATION CAN BE PERFORMED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP REPORT WILL BE FILED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19476349 ↗MalfunctionJan 19, 2024Device Alarm System; Crack; Failure to Deliver; Gas/Air Leak; Intermittent Loss of Power; Device Alarm System; Crack; Failure to Deliver; Gas/Air Leak; Intermittent Loss of Power
THE MANUFACTURER WAS CONTACTED ABOUT AN EVERFLO OXYGEN CONCENTRATOR. PER THE WORK ORDER RECEIVED BY THE FASC THE SIEVES BEDS WERE BLOWN, SOLENOID INTERMITTENT, THE REGULATOR WAS LEAKING, THE COMPRESSOR RATTLING, THE PCA BOARD HAD LIGHTS, AND THE POWER CORD WAS CUT OFF. THERE WAS NO ALLEGATION OF PATIENT HARM OR SERIOUS INJURY. NO CLINICAL INFORMATION OR MEDICAL INFORMATION WAS SPECIFIED. THE FASC PERFORMED A COMPREHENSIVE INSPECTION AND REPLACED THE NECESSARY PARTS. AT THIS TIME, NO FURTHER INVESTIGATION CAN BE PERFORMED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP REPORT WILL BE FILED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18546156 ↗MalfunctionJan 17, 2024Mechanical Problem; Power Problem; Mechanical Problem; Power Problem
THE MANUFACTURER BECAME AWARE OF AN ALLEGATION OF EVERFLO QUIET THAT THE PATIENT ALLEGES THAT THE DEVICE IS ALARMING. A DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER. DURING THE EVALUATION OF THE DEVICE, THEY FOUND OUT THAT THE COMPLAINT WAS CONFIRMED, THE SIEVE BEDS BLOWN, THE SOLENOID WAS INTERMITTENT, THE REGULATOR WAS DAMAGED, PCA BOARD HAS NON-STOP ALARMING, FLOW METER WAS CRACKED, AND POWER CORD WAS SMASHED. AT THIS TIME, NO FURTHER INVESTIGATION CAN BE PERFORMED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP REPORT WILL BE FILED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18530827 ↗MalfunctionJan 2, 2024Leak/Splash; Leak/Splash
THE MANUFACTURER RECEIVED INFORMATION IN RELATION TO A EVERFLO QUIET DEVICE. THE DEVICE WAS SERVICED BY A FIELD SERVICE TECHNICIAN. DURING THE EVALUATION OF THE DEVICE HAD A SIEVE BED SMELL THAT WAS REPLACED FOR PREVENTATIVE MAINTENANCE. THE INLET FILTER WAS DIRTY, THE MANIFOLD ASSEMBLY WAS LEAKING AND THE SHIPPING BOX WAS DAMAGED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18431183 ↗MalfunctionNov 21, 2017Fire; Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN EVERFLO OXYGEN CONCENTRATOR WAS INVOLVED IN A HOUSE FIRE. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THE INVESTIGATION IS STILL ONGOING. A FOLLOW UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER HAS COMPLETED THE INVESTIGATION.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7049010 ↗
Other Generator, oxygen, portable devices
- DeVilbiss 525DSDevilbiss Healthcare LLC · Generator, Oxygen, Portable2 recalls14 deaths287 reports
- Inogen One G5 Oxygen Concentrator - Bluetooth 1260 BluetoothInogen, Inc. · Generator, Oxygen, Portable3 deaths235 reports
- EverFlo EverFlo w/O2 Percentage Indicator, 230V, EURespironics, Inc. · Generator, oxygen, portable5 deaths112 reports
- Inogen One G3 Oxygen Concentrator 1050Inogen, Inc. · Generator, Oxygen, Portable2 deaths111 reports
- DeVilbiss 125DDevilbiss Healthcare LLC · Generator, Oxygen, Portable2 recalls1 death3 reports
- EverFlo EVERFLO NON-O2 Percentage Indicator, 120V, US,CARespironics, Inc. · Generator, oxygen, portable19 deaths80 reports
- EverFlo EVERFLO w/Oxygen Percentage Indicator, 120V US, CARespironics, Inc. · Generator, oxygen, portable2 deaths71 reports
- SimplyGo Mini SimplyGo Mini, USRespironics, Inc. · Generator, oxygen, portable1 recall1 death5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026