EverFlo EVERFLO 120V w/O2 Percent Indicator, UltraFill, USRecalls & reported problems
Respironics, Inc. · EVERFLO 120V w/O2 Percent Indicator, UltraFill, USA · Class 2
All models of EverFlo (20)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00606959032675), received between Nov 17, 2016 and Feb 14, 2018 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Device Contamination with Chemical or Other Material” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Device Contamination with Chemical or Other Material1
- Device Operational Issue1
- Fire1
- Improper or Incorrect Procedure or Method1
Patient problems
- Burn(s)1
- Burn, Thermal1
- Dyspnea1
- Increased Respiratory Rate1
- No Consequences Or Impact To Patient1
- Respiratory Distress1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryFeb 14, 2018Improper or Incorrect Procedure or Method; Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A PATIENT LIT A CIGARETTE WHILE USING AN EVERFLO OXYGEN CONCENTRATOR. THE PATIENT WAS ADMITTED TO THE HOSPITAL WITH FACIAL BURNS AND SMOKE INHALATION. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THE DEVICE HAS YET TO BE RETURNED TO THE MANUFACTURER FOR EVALUATION. A FOLLOW UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER HAS COMPLETED THE INVESTIGATION.
Patient problems: Burn(s); Burn, Thermal; Respiratory Distress
MAUDE report 7270678 ↗MalfunctionJun 21, 2017Fire; Fire
THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN EVERFLO OXYGEN CONCENTRATOR CAUGHT ON FIRE. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THE DEVICE WAS RETURNED TO THE MANUFACTURER'S SERVICE CENTER FOR EVALUATION. THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE HAD EVIDENCE OF THERMAL DAMAGE TO THE OUTLET PORT. THE THERMAL DAMAGE WAS CAUSED BY AN EXTERNAL SOURCE. THE DEVICE WAS TESTED AND WAS FOUND TO OPERATE AND ALARM TO DESIGN SPECIFICATIONS. PRODUCT LABELING STATES, "OXYGEN VIGOROUSLY ACCELERATES COMBUSTION AND SHOULD BE KEPT AWAY FROM HEAT OR OPEN FLAME. NOT SUITABLE FOR USE IN THE PRESENCE OF A FLAMMABLE ANESTHETIC MIXTURE WITH AIR OR WITH OXYGEN OR NITROUS OXIDE. DO NOT SMOKE, ALLOW OTHERS TO SMOKE OR HAVE OPEN FLAMES NEAR THE CONCENTRATOR WHEN IT IS IN USE."
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6657410 ↗MalfunctionNov 17, 2016Device Operational Issue; Device Operational Issue; Device Contamination with Chemical or Other Material
THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN EVERFLO OXYGEN CONCENTRATOR EMITTED A WHITE POWDER. THE PATIENT BECAME SHORT OF BREATH AND WAS HOSPITALIZED FOR ONE DAY AND WAS RELEASED. THE DEVICE HAS YET TO BE RETURNED TO THE MANUFACTURER'S SERVICE CENTER FOR EVALUATION. A FOLLOW UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER HAS COMPLETED THE INVESTIGATION.
Patient problems: Dyspnea; Increased Respiratory Rate
MAUDE report 6109167 ↗
Other Generator, oxygen, portable devices
- DeVilbiss 525DSDevilbiss Healthcare LLC · Generator, Oxygen, Portable2 recalls14 deaths287 reports
- Inogen One G5 Oxygen Concentrator - Bluetooth 1260 BluetoothInogen, Inc. · Generator, Oxygen, Portable3 deaths235 reports
- EverFlo EverFlo w/O2 Percentage Indicator, 230V, EURespironics, Inc. · Generator, oxygen, portable5 deaths112 reports
- Inogen One G3 Oxygen Concentrator 1050Inogen, Inc. · Generator, Oxygen, Portable2 deaths111 reports
- DeVilbiss 125DDevilbiss Healthcare LLC · Generator, Oxygen, Portable2 recalls1 death3 reports
- EverFlo EVERFLO NON-O2 Percentage Indicator, 120V, US,CARespironics, Inc. · Generator, oxygen, portable19 deaths80 reports
- EverFlo EVERFLO w/Oxygen Percentage Indicator, 120V US, CARespironics, Inc. · Generator, oxygen, portable2 deaths71 reports
- SimplyGo Mini SimplyGo Mini, USRespironics, Inc. · Generator, oxygen, portable1 recall1 death5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page