SimplyGo Mini SimplyGo Mini with Standard Battery, IKKRecalls & reported problems
Respironics, Inc. · SimplyGo Mini with Standard Battery, IKK. · Class 2
All models of SimplyGo Mini (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00606959033825), received between May 5, 2025 and Feb 3, 2026 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Overheating of Device3
- Thermal Decomposition of Device2
- Smoking1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 3, 2026Thermal Decomposition of Device
A SIMPLYGO MINI DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR SERVICE. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. THERE WAS NO REPORT OF MEDICAL INTERVENTION. DURING THE EVALUATION OF THE DEVICE AT THIRD-PARTY SERVICE CENTER, THE SERVICE OBSERVED THE LOW O2 LEVELS WERE LOW BECAUSE THE MOLECULAR SIEVES WERE CLOGGED. ADDITIONALLY, LEAKS WERE IDENTIFIED ON BOTH THE OXYGEN AND A CIRCUITS. THE COMPRESSOR IS NOT OPERATING WITHIN NORMAL PARAMETERS, AND ITS TUBING IS CONTAMINATED. FURTHERMORE, THE PCA FAILED T START BECAUSE IT IS BURNT OUT, AND THE PATIENT FILTER IS CONTAMINATED. THE DEVICE RETURNED UNREPAIRED. FURTHER THE DEVICE WAS SENT TO PIL FOR FURTHER INVESTIGATION. THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION AT THIS TIME. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP REPORT WILL BE FILED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24250771 ↗MalfunctionDec 22, 2025Overheating of Device; Overheating of Device
A SIMPLYGO MINI DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION AT THE SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND TO HAVE A BURNED PCA BOARD. ADDITIONALLY, THE SIEVES WERE CLOGGED AND REQUIRED REPLACEMENT. BOTH THE AIR SIDE AND OXYGEN SIDE WERE CRACKED AND NEEDED REPLACEMENT, AND THE INLET FILTER REQUIRED REPLACEMENT AS PART OF PREVENTIVE MAINTENANCE. THE THIRD-PARTY SERVICE CENTER WAS UNABLE TO COMPLETE THE INVESTIGATION, AND THE DEVICE WILL BE FORWARDED TO THE MANUFACTURER¿S PRODUCT INVESTIGATION LABORATORY FOR FURTHER EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23871288 ↗MalfunctionNov 14, 2025Overheating of Device; Overheating of Device
A SIMPLYGO MINI DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR EVALUATION, AND THE CUSTOMER COMPLAINT WAS CONFIRMED. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED THAT THE MAIN PCA WAS BURNED AND NEEDED REPLACEMENT. DUE TO THE ALLEGED MALFUNCTION, THE DEVICE WILL BE RETURNED TO THE MANUFACTURER'S PRODUCT INVESTIGATION LAB FOR ADDITIONAL TESTING. THE ROOT CAUSE IS NOT CONFIRMED AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23553786 ↗MalfunctionJul 16, 2025Overheating of Device; Overheating of Device
A DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR SERVICE. THERE WAS NO REPORT OF SERIOUS PATIENT HARM OR INJURY. THERE WAS NO REPORT OF MEDICAL INTERVENTION. DURING THE EVALUATION OF THE DEVICE, THEY FOUND OUT THAT PCA IS BURNT, THE UNIT DOES NOT START, SIEVES ARE CLOGGED, THE AIR SIDE AND O2 SIDE AND COMPRESSOR ARE CONTAMINATED AND NEED TO BE REPLACED. PASS ALL TEST. DEVICE IS RETURNED, REPAIRED. REPORTABLE SINCE DEVICE ISSUE/EVENT HAS OR MAY HAVE CAUSED OR CONTRIBUTED TO A DEATH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22523683 ↗MalfunctionMay 7, 2025Smoking; Smoking
THE MANUFACTURER WAS CONTACTED ABOUT A SIMPLYGO MINI STAND BATTERY IKK DEVICE DUE TO REPORTS OF IT SMOKING. THE SERVICE CENTER HAS RECEIVED THE DEVICE FOR EVALUATION. IMPORTANTLY, THERE ARE NO CLAIMS OF PATIENT HARM OR SERIOUS INJURY, AND NO MEDICAL INTERVENTION HAS BEEN SPECIFIED. THE MANUFACTURER'S INVESTIGATION IS STILL IN PROGRESS. IF ANY ADDITIONAL RELEVANT INFORMATION COMES IN, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21982161 ↗MalfunctionMay 5, 2025Thermal Decomposition of Device; Thermal Decomposition of Device
A SIMPLY GO MINI DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO A REQUEST FOR SERVICING THE SIEVE FILTER. DURING THE INITIAL EVALUATION OF THE DEVICE, THE THIRD-PARTY SERVICE CENTER VISUALLY INSPECTED THE UNIT AND FOUND THE TOP OF THE ENCLOSURE DENTED AND BURNED. THERE WAS NO ALLEGATION OF PATIENT HARM OR INJURY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21964476 ↗
Other Generator, oxygen, portable devices
- DeVilbiss 525DSDevilbiss Healthcare LLC · Generator, Oxygen, Portable2 recalls14 deaths287 reports
- Inogen One G5 Oxygen Concentrator - Bluetooth 1260 BluetoothInogen, Inc. · Generator, Oxygen, Portable3 deaths235 reports
- EverFlo EverFlo w/O2 Percentage Indicator, 230V, EURespironics, Inc. · Generator, oxygen, portable5 deaths112 reports
- Inogen One G3 Oxygen Concentrator 1050Inogen, Inc. · Generator, Oxygen, Portable2 deaths111 reports
- DeVilbiss 125DDevilbiss Healthcare LLC · Generator, Oxygen, Portable2 recalls1 death3 reports
- EverFlo EVERFLO NON-O2 Percentage Indicator, 120V, US,CARespironics, Inc. · Generator, oxygen, portable19 deaths80 reports
- EverFlo EVERFLO w/Oxygen Percentage Indicator, 120V US, CARespironics, Inc. · Generator, oxygen, portable2 deaths71 reports
- SimplyGo Mini SimplyGo Mini, USRespironics, Inc. · Generator, oxygen, portable1 recall1 death5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page