SimplyGo Mini SimplyGo Mini with Extended Battery, IKKRecalls & reported problems
Respironics, Inc. · SimplyGo Mini with Extended Battery, IKK. · Class 2
All models of SimplyGo Mini (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00606959033832), received between Apr 25, 2024 and May 27, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Overheating of Device2
- Excessive Heating1
- Thermal Decomposition of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 27, 2026Overheating of Device
A SIMPLYGO MINI DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR EVALUATION, AND THE CUSTOMER COMPLAINT WAS CONFIRMED. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED THAT THE PCA WAS BURNED AND REQUIRED REPLACEMENT. DUE TO THE ALLEGED MALFUNCTION, THE DEVICE WILL BE RETURNED TO THE MANUFACTURER'S PRODUCT INVESTIGATION LAB FOR ADDITIONAL TESTING. THE ROOT CAUSE IS NOT CONFIRMED AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25301050 ↗MalfunctionApr 30, 2026Overheating of Device
A SIMPLYGO MINI DEVICE WAS RETURNED TO THE THIRD-PARTY SERVICE CENTER FOR EVALUATION. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN NOTED A THERMAL EVENT WITH BURN MARKS AND BURNT PCA. THE DEVICE WILL BE RETURNED TO THE PRODUCT INVESTIGATION LAB (PIL) FOR FURTHER INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25049942 ↗MalfunctionJul 14, 2025Excessive Heating; Excessive Heating
THE MANUFACTURER RECEIVED INFORMATION ON A SIMPLYGO MINI DEVICE. ACCORDING TO CUSTOMER, THE DEVICE INDICATED OVERHEATING. THE DEVICE HAS BEEN RETRIEVED FROM THE PATIENT. DURING EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF BURNT PCA. MAIN PCA, CLOGGED, SIEVES, AND DEFECTIVE O2 SIDE WAS REPLACED THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE DEVICE WAS REPAIRED AND PASSED ALL FINAL TESTING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22496569 ↗MalfunctionApr 25, 2024Thermal Decomposition of Device; Thermal Decomposition of Device
AN OXYGEN CONCENTRATOR WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR REPAIRS OF A PRODUCT MALFUNCTION. THE PATIENT REPORTED ALLEGATIONS OF A BURNING SMELL COMING FROM THE DEVICE. THERE WERE NO ALLEGATIONS OF HARM OR SERIOUS INJURY. MEDICAL INTERVENTION WAS NOT SPECIFIED. DURING THE EVALUATION OF THE DEVICE AT THE THIRD PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF A BURNT PCA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19178047 ↗
Other Generator, oxygen, portable devices
- DeVilbiss 525DSDevilbiss Healthcare LLC · Generator, Oxygen, Portable2 recalls14 deaths287 reports
- Inogen One G5 Oxygen Concentrator - Bluetooth 1260 BluetoothInogen, Inc. · Generator, Oxygen, Portable3 deaths235 reports
- EverFlo EverFlo w/O2 Percentage Indicator, 230V, EURespironics, Inc. · Generator, oxygen, portable5 deaths112 reports
- Inogen One G3 Oxygen Concentrator 1050Inogen, Inc. · Generator, Oxygen, Portable2 deaths111 reports
- DeVilbiss 125DDevilbiss Healthcare LLC · Generator, Oxygen, Portable2 recalls1 death3 reports
- EverFlo EVERFLO NON-O2 Percentage Indicator, 120V, US,CARespironics, Inc. · Generator, oxygen, portable19 deaths80 reports
- EverFlo EVERFLO w/Oxygen Percentage Indicator, 120V US, CARespironics, Inc. · Generator, oxygen, portable2 deaths71 reports
- SimplyGo Mini SimplyGo Mini, USRespironics, Inc. · Generator, oxygen, portable1 recall1 death5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page