REMstar System One 60 Series REMstar Pro C-Flex+, BRRecalls & reported problems
Respironics, Inc. · REMstar Pro C-Flex Plus, Brazil · Class 2
All models of REMstar System One 60 Series (92)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Respironics REMstar Pro C-Flex+ is a breathing support device. The 24 reports mostly describe sound-dampening foam breaking down, sometimes with particles found inside the device. They report no deaths and no injuries, and no recalls name this device. Reports rose from 4 in the previous 12 months to 20 in the most recent 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 24
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 00606959037885), received between Oct 29, 2025 and Mar 9, 2026 — classified as 24 malfunctions.
The most frequently reported specific device problem is “Degraded” (24 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Degraded24
Patient problems
- No Clinical Signs, Symptoms or Conditions24
Recent adverse event reports
The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 9, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A REMSTAR PRO C-FLEX+ DEVICE WAS RETURNED TO A MANUFACTURER SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION/PARTICLES WAS FOUND. ADDITIONALLY, THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24543149 ↗MalfunctionMar 4, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A REMSTAR PRO CFLEX+ DEVICE WAS RETURNED TO A SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING EVALUATION, NO FOAM PARTICLES WERE NOTED IN THE AIRPATH, YET FOAM DEGRADATION WAS FOUND. ADDITIONALLY, THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT, THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24510992 ↗MalfunctionMar 4, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A REMSTAR PRO CFLEX+ DEVICE WAS RETURNED TO A MANUFACTURER SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION WAS FOUND. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24507920 ↗MalfunctionMar 3, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A REMSTAR PRO C-FLEX+ DEVICE WAS RETURNED TO A MANUFACTURER SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION/PARTICLES WAS FOUND. ADDITIONALLY, THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24500538 ↗MalfunctionMar 3, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A REMSTAR PRO C-FLEX+ DEVICE WAS RETURNED TO A MANUFACTURER SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION/PARTICLES WAS FOUND. ADDITIONALLY, THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24500343 ↗MalfunctionMar 3, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A REMSTAR PRO C FLEX+ DEVICE WAS RETURNED TO A MANUFACTURER SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION WAS FOUND. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DURING THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24500140 ↗MalfunctionMar 3, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A REMSTAR PRO CFLEX+ DEVICE WAS RETURNED TO A MANUFACTURER SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION/PARTICLES WAS FOUND. ADDITIONALLY, THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24499939 ↗MalfunctionMar 3, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A REMSTAR PRO C-FLEX+, DEVICE WAS RETURNED TO A SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION/PARTICLES WAS FOUND. ADDITIONALLY, DURING THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24495716 ↗MalfunctionMar 3, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A REMSTAR PRO CFLEX + DEVICE WAS RETURNED TO MANUFACTURER SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE VISIBLE FOAM PARTICLES/DEGRADATION. THE DEVICE WAS SCRAPPED AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24494705 ↗MalfunctionMar 2, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A REMSTAR PRO CFLEX+ DEVICE WAS RETURNED TO A MANUFACTURER SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT MANUFACTURER SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION/PARTICLES WAS FOUND. ADDITIONALLY, THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24482614 ↗MalfunctionFeb 27, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A REMSTAR PRO C-FLEX+ DEVICE WAS RETURNED TO A MANUFACTURER SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION/PARTICLES WAS FOUND. LASTLY, THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24476458 ↗MalfunctionFeb 27, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A REMSTAR PRO CFLEX+ DEVICE WAS RETURNED TO MANUFACTURER SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE VISIBLE FOAM PARTICLES/DEGRADATION. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24472505 ↗MalfunctionFeb 25, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A REMSTAR PRO CFLEX+ DEVICE WAS RETURNED TO A MANUFACTURER SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND CONFIRMED EVIDENCE OF FOAM. ADDITIONALLY, THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24448374 ↗MalfunctionFeb 25, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A REMSTAR PRO CFLEX+ DEVICE WAS RETURNED TO A MANUFACTURER SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION/PARTICLES WAS FOUND. ADDITIONALLY, THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24446409 ↗MalfunctionFeb 19, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A REMSTAR PRO CFLEX+ DEVICE WAS RETURNED TO THE SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION/PARTICLES WAS FOUND. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO FOUND THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24394250 ↗MalfunctionFeb 19, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A REMSTAR PRO CFLEX+ DEVICE WAS RETURNED TO MANUFACTURER SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT.DURING THE EVALUATION OF THE DEVICE AT SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM WAS CONFIRMED. ADDITIONALLY, THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24393952 ↗MalfunctionFeb 18, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A REMSTAR PRO C-FLEX+ DEVICE WAS RETURNED TO A SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION/PARTICLES WAS FOUND. LASTLY, THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24380620 ↗MalfunctionFeb 17, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE/RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A REMSTAR PRO CFLEX+ ASSY DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND VISIBLE FOAM PARTICLES WERE OBSERVED. THE DEVICE WAS SCRAPPED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24364658 ↗MalfunctionNov 27, 2025Degraded; Degraded
A DEVICE WAS RETURNED TO A MANUFACTURER'S SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM DEGRADATION. DURING THE EVALUATION, THE DEVICE POWERED ON AND AIRFLOW WAS CONFIRMED. THE DEVICE'S DOWNLOADED LOGS WERE REVIEWED BY THE MANUFACTURER'S SERVICE CENTER. THERE WAS THREE ERROR CODES FOUND. THE MANUFACTURER'S SERVICE CENTER CONCLUDES THAT THERE WAS VISIBLE FOAM DEGRADATION AND UNIT GOT SCRAPPED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23665275 ↗MalfunctionNov 27, 2025Degraded; Degraded
A DEVICE WAS RETURNED TO A MANUFACTURER'S SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM DEGRADATION. DURING THE EVALUATION, THE DEVICE POWERED ON AND AIRFLOW WAS CONFIRMED. THE DEVICE'S DOWNLOADED LOGS WERE REVIEWED BY THE MANUFACTURER'S SERVICE CENTER. THERE WAS NO ERROR CODES FOUND. THE MANUFACTURER'S SERVICE CENTER CONCLUDES THAT THERE WAS VISIBLE FOAM DEGRADATION AND UNIT GOT SCRAPPED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23665267 ↗MalfunctionOct 30, 2025Degraded; Degraded
A REMSTAR PRO CFLEX+ DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER AS PART OF THE UNO RECALL PROCESS WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE BLOWER FOAM DEGRADATION. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED ERROR CODE 30: ERR_MOTOR_SPINUP_FLUX_HIGH AND ERROR CODE 53: ERR_COMP_LOG_SEM_TIMEOUT WERE PRESENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23432859 ↗MalfunctionOct 29, 2025Degraded; Degraded
A REMSTAR PRO C-FLEX+ DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER AS PART OF THE UNO RECALL PROCESS WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED. AT THIS TIME, NO FURTHER INVESTIGATION CAN BE PERFORMED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP REPORT WILL BE FILED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23418413 ↗MalfunctionOct 29, 2025Degraded; Degraded
A REMSTAR PRO C FLEX DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER AS PART OF THE UNO RECALL PROCESS WITH NO INITIAL ALLEGED COMPLAINT. THERE WAS NO DEATH, SERIOUS HARM OR INJURY WERE REPORTED, AND NO MEDICAL INTERVENTION WAS PROVIDED. DURING THE EVALUATION OF THE DEVICE, THE THIRD-PARTY SERVICE CENTER VISUALLY INSPECTED THE DEVICE AND EVIDENCE OF FOAM DEGRADATION WAS FOUND. IN ADDITION, THE DEVICE DISPLAYED MULTIPLE OCCURRENCES OF ERROR CODE E-53: ERR_COMP_LOG_SEM_TIMEOUT AS RECORDED IN ERROR LOG. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23416612 ↗MalfunctionOct 29, 2025Degraded; Degraded
A REMSTAR PRO C-FLEX DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER AS PART OF THE UNO RECALL PROCESS WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED AN ERROR CODE PRESENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23415025 ↗
Other VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices
- REMstar System One 60 Series REMstar Auto A-Flex, w/Humid System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)117 deaths10,075 reports
- DreamStation DreamStation Auto BiPAP w/Humidifier, w/Cell, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)106 deaths5,937 reports
- DreamStation DreamStation Auto CPAP with Humidifier, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)44 deaths5,793 reports
- REMstar System One 60 Series REMstar Auto A-Flex System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)64 deaths5,138 reports
- REMstar System One 60 Series REMstar Pro C-Flex+, w/Humid System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)65 deaths4,520 reports
- REMstar System One 60 Series REMstar Auto A-Flex w/HT Humid System One 60 Series, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)43 deaths4,149 reports
- DreamStation DreamStation Auto CPAP w/Humid w/Heated Tube, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)30 deaths3,981 reports
- DreamStation DreamStation Auto CPAP w/Humidifier, w/Cell, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)86 deaths3,516 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page