REMstar System One 50 Series REMstar Auto A-Flex, CNRecalls & reported problems

Respironics, Inc. · REMstar Auto A-Flex, China · Class 2

All models of REMstar System One 50 Series (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This is a Respironics REMstar Auto A-Flex home CPAP ventilator. The 27 reports mostly describe black or sponge-like particles breaking off the sound-dampening foam inside the air path and appearing in the air the user breathes, along with cough, chest tightness, nasal discomfort and lung pain. There are 3 reports of injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
27
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 27 adverse event reports that list this exact device (DI 00606959039612), received between Sep 10, 2021 and Oct 18, 2025 — classified as 3 injuries, 24 malfunctions.

The most frequently reported specific device problem is “Degraded” (27 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 5 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Degraded27

Patient problems

  • No Clinical Signs, Symptoms or Conditions18
  • Nodule4
  • Discomfort2
  • Cancer1
  • Ischemia Stroke1
  • Pain1
  • Pneumonia1
  • Sleep Dysfunction1

Recent adverse event reports

The 25 most recent of 27 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 18, 2025Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THIS REPORT WAS RELATED TO A REMSTAR AUTO A-FLEX. THE MANUFACTURER RECEIVED INFORMATION FROM THE PATIENT ALLEGING SPONGE LIKE PARTICLES IN THE AIRPATH CAUSING DISCOMFORT IN THEIR NASAL CAVITY. MEDICAL INTERVENTION WAS NOT SPECIFIED. THE DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP REPORT WILL BE FILED.

    Patient problems: Discomfort

    MAUDE report 23332033 ↗
  • InjuryMay 7, 2025Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION OF PATIENT ALLEGING LUNG PAIN, CEREBRAL INFARCTION. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE. REPORTABLE SINCE DEVICE ISSUE/EVENT HAS OR MAY HAVE CAUSED OR CONTRIBUTED TO A SERIOUS INJURY. THIS COMPLAINT IS INCLUDED IN THE VOLUNTARY FIELD SAFETY NOTICE/RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES.

    Patient problems: Ischemia Stroke; Pain

    MAUDE report 21981006 ↗
  • MalfunctionMar 18, 2025Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING A DRY COUGH AND CHEST TIGHTNESS RECENTLY, WHICH WAS SUSPECTED TO BE CAUSED BY A VENTILATOR. CUSTOMER ALSO REPORTED THAT HE WOULD GO TO THE HOSPITAL FOR EXAMINATION LATER, AND REPORTED THAT HE FOUND A LARGE NUMBER OF BLACK PARTICLES IN THE VENTILATOR TUBING IN THE PAST TWO MONTHS, SUSPECTED THAT HE HAD INHALED THE PARTICLES, AND WAS WORRIED ABOUT THE IMPACT ON HIS BODY. THERE WAS NO REPORT OF MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION CAN BE REQUESTED AT THIS TIME. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21633568 ↗
  • InjurySep 27, 2024Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE CUSTOMER ALLEGED THAT THE USER HAS CANCER IN RELATION TO THE DEVICE. BASED ON AVAILABLE INFORMATION AT THIS TIME, THE ALLEGED CANCER IS ASSESSED AS NOT RELATED TO THE DEVICE IN THIS CASE. ADDITIONAL INFORMATION HAS BEEN REQUESTED. THE MANUFACTURER IS IN THE PROCESS OF OBTAINING ADDITIONAL INFORMATION CONCERNING THIS EVENT AND THE COMPLAINT IS STILL UNDER INVESTIGATION. A FINAL REPORT WILL BE SUBMITTED ONCE THE INVESTIGATION IS COMPLETE.

    Patient problems: Cancer

    MAUDE report 20325706 ↗
  • MalfunctionSep 6, 2024Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING VISUALIZATION OF BLACK PARTICLES. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS SPECIFIED BY THE PATIENT. THE DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER FOR INVESTIGATION. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20162730 ↗
  • MalfunctionFeb 15, 2024Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE/RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT HAS ALLEGED VISUALIZATION OF PARTICLES. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. THE DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER FOR INVESTIGATION. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18715531 ↗
  • MalfunctionFeb 15, 2024Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE/RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT HAS ALLEGED VISUALIZATION OF PARTICLES. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. THE DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER FOR INVESTIGATION. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18715114 ↗
  • MalfunctionFeb 15, 2024Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE/RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT HAS ALLEGED VISUALIZATION OF PARTICLES. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. THE DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER FOR INVESTIGATION. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18714835 ↗
  • MalfunctionFeb 15, 2024Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE/RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT HAS ALLEGED VISUALIZATION OF PARTICLES. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. THE DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER FOR INVESTIGATION. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18714736 ↗
  • MalfunctionFeb 14, 2024Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE/RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A REMSTAR AUTO AFLEX DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT ALLEGES UNCOMFORTABLE SYMPTOMS OF PNEUMONIA AND FEELS THESE EVENTS ARE RELATED TO THE DEVICE. MEDICAL INTERVENTION WAS NOT SPECIFIED. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. THE DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER FOR INVESTIGATION. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: Pneumonia

    MAUDE report 18705822 ↗
  • InjuryNov 29, 2022Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM BECAME DEGRADED AND CAUSED TINY NODULES IN THE LEFT UPPER LOBE OF THE LUNG. THE PATIENT DID REPORT TO RECEIVE MEDICAL INTERVENTION AND WAS HOSPITALIZED AND HAD DIAGNOSTIC TESTING AND MEDICATION. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: Nodule

    MAUDE report 15883271 ↗
  • MalfunctionJul 15, 2022Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT HAS ALLEGED TO SEEING BLACK PARTICLES IN THE DEVICE. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15031356 ↗
  • MalfunctionJun 30, 2022Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT HAS ALLEGED TO SEEING BLACK DROPLETS IN THE DEVICE. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14877184 ↗
  • MalfunctionJun 21, 2022Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT HAS ALLEGED TO BEING WORRIED ABOUT THE BLACK MATERIALS IN THE DEVICE NOT COMING CLEAN. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14755765 ↗
  • MalfunctionMay 7, 2022Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT ALLEGED VISUALIZATION OF A BLACK UNKNOWN SUBSTANCE BLOWING FROM THE DEVICE. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14317858 ↗
  • MalfunctionApr 1, 2022Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE'S SOUND ABATEMENT FOAM. THERE WAS NO REPORT OF PATIENT HARM/INJURY. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.

    Patient problems: Discomfort; No Clinical Signs, Symptoms or Conditions

    MAUDE report 13978575 ↗
  • MalfunctionMar 25, 2022Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT VISUALIZED BLACK PARTICLES. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13902724 ↗
  • MalfunctionDec 27, 2021Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE'S SOUND ABATEMENT FOAM. THERE WAS NO REPORT OF PATIENT HARM/INJURY. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13088301 ↗
  • MalfunctionDec 7, 2021Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE'S SOUND ABATEMENT FOAM. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12955930 ↗
  • MalfunctionOct 13, 2021Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THERE WAS NO REPORT OF PATIENT HARM OR INJURY THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12625490 ↗
  • MalfunctionOct 12, 2021Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING A CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE'S SOUND ABATEMENT FOAM BECAME DEGRADED AND CAUSED A PATIENT TO DEVELOP DIFFICULTY SLEEPING. THE PATIENT WAS PRESCRIBED MEDICATION IN RESPONSE TO THE REPORTED EVENT. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.

    Patient problems: Sleep Dysfunction

    MAUDE report 12616743 ↗
  • MalfunctionOct 6, 2021Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING A CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE'S SOUND ABATEMENT FOAM BECAME DEGRADED AND CAUSED THE PATIENT TO DEVELOP LUNG NODULES. THE MEDICAL INTERVENTION THAT THE PATIENT RECEIVED IN RESPONSE TO THE EVENT IS CURRENTLY UNKNOWN. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.

    Patient problems: Nodule

    MAUDE report 12583538 ↗
  • MalfunctionOct 5, 2021Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING A CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE'S SOUND ABATEMENT FOAM BECAME DEGRADED AND CAUSED A PATIENT TO DEVELOP NODULES IN THEIR LUNGS. THERE IS NO REPORT OF THE MEDICAL INTERVENTION THAT THE PATIENT HAS RECEIVED AT THIS TIME. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.

    Patient problems: Nodule

    MAUDE report 12579032 ↗
  • MalfunctionSep 30, 2021Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE'S SOUND ABATEMENT FOAM. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12558514 ↗
  • MalfunctionSep 22, 2021Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE'S SOUND ABATEMENT FOAM. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12516349 ↗

Other VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices

All VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices →