Skip to content

DreamWear Full DreamWear Full Mask Small, w/Medium Frame w/Headgear, GlobalRecalls & reported problems

Respironics, Inc. · DreamWear Full Mask, Small, with Medium Frame with Headgear, Global · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00606959046030), received between Jul 15, 2022 and Jul 15, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Patient-Device Incompatibility” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Patient-Device Incompatibility1

Patient problems

  • Unspecified Heart Problem1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 15, 2022Patient-Device Incompatibility; Patient-Device Incompatibility

    THE MANUFACTURER BECAME AWARE OF AN ALLEGATION FROM A USER WHO REPORTED WHILE USING HIS DREAMWEAR FULL FACE MASK WITH MAGNETS, HE HAD ISSUES WITH HIS IMPLANTED DEFIBRILLATOR SHUTTING OFF PERIODICALLY DURING THE NIGHT. THE USER WAS ADVISED OF THIS AFTER HE HAD VISITED HIS CARDIOLOGIST AND THEY WERE ATTRIBUTING THIS PROBLEM TO THE MASK. THE USER WAS SWITCHED TO A DIFFERENT MASK. THE USER STATED NO HARM OR INJURY OCCURRED. NO PRODUCT WAS RETURNED FOR INVESTIGATION. THE DREAMWEAR INSTRUCTIONS FOR USE INCLUDES THE FOLLOWING WARNINGS: THE HEADGEAR CLIPS CONTAIN MAGNETS. CONTACT YOUR HEALTHCARE PROFESSIONAL BEFORE YOU USE THIS MASK. SOME MEDICAL DEVICES MAY BE AFFECTED BY MAGNETIC FIELDS. THE MAGNETIC CLIPS IN THIS MASK SHOULD BE KEPT AT LEAST 2 IN. (50 MM) AWAY FROM ANY ACTIVE MEDICAL DEVICE WITH SPECIAL ATTENTION TO IMPLANTED DEVICES SUCH AS PACEMAKERS, DEFIBRILLATORS AND COCHLEAR IMPLANTS. DO NOT USE IN OR NEAR MAGNETIC RESONANCE IMAGING (MRI) EQUIPMENT. THIS REPORT WILL BE FILED AS AN INITIAL-FINAL REPORT. WE WILL CONTINUE TO MONITOR THESE COMPLAINTS REGARDING MAGNETS. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP REPORT WILL BE FILED.

    Patient problems: Unspecified Heart Problem

    MAUDE report 15033181 ↗