Skip to content

DreamWear Full DreamWear Full Mask Medium, w/Medium Frame w/Headgear, GlobalRecalls & reported problems

Respironics, Inc. · DreamWear Full Mask, Medium, with Medium Frame with Headgear, Global · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00606959046047), received between Nov 9, 2022 and Apr 2, 2024 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient Device Interaction Problem3
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Discomfort1
  • Dry Eye(s)1
  • Dry Mouth1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 2, 2024Leak/Splash; Leak/Splash

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING THE PATIENT STATES HIS MASK HAS AN AIR LEAKAGE FROM THE LIPS WHEN HE WEARS IT. PATIENT ALSO ALLEGES THAT HE WAKES UP WITH DRY MOUTH, BLOATING AND DISCOMFORT IN THE STOMACH WITH DRYNESS IN HIS EYES. THERE WAS NO HARM OR INJURY REPORTED. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: Discomfort; Dry Eye(s); Dry Mouth

    MAUDE report 19017235 ↗
  • InjuryNov 10, 2022Patient Device Interaction Problem; Patient Device Interaction Problem

    THE MANUFACTURER BECAME AWARE OF AN ALLEGATION THAT A USER WITH ORTHODONTIC BRACES IN HIS MOUTH EXPERIENCED LOWER JAW GROWTH FROM USING THE DREAMWEAR FULL FACE MASK WITH MAGNETIC CLIPS. THE USER HAS SINCE HAD A TONSILLECTOMY, THE BRACES WERE REMOVED  AND NO LONGER USES A CPAP. THE USER'S MOTHER STATES HE NOW NEEDS A SECOND SET OF ORTHODONTIC BRACES AND JAW SURGERY DUE TO HIS LOWER JAW GROWING OUT ALLEGEDLY DUE TO THE MAGNETS. REQUESTED THE RETURN OF THE MASK FOR INVESTIGATION. TO DATE, NO MASK HAS BEEN RETURNED. THE INVESTIGATION IS ON-GOING. A FOLLOW UP FINAL WILL BE SUBMITTED ONCE THE INVESTIGATION IS COMPLETE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 15770548 ↗
  • MalfunctionNov 9, 2022Patient Device Interaction Problem; Patient Device Interaction Problem

    THE MANUFACTURER RECEIVED INFORMATION THAT AN END USER WAS ASKING IF THERE WAS ANY DOCUMENTATION OF TESTING WITH THE DREAMWEAR FULL FACE MASK AND A PATIENT WHO HAS A PACEMAKER. THE USER WAS INFORMED THAT THE MANUAL STATES TO KEEP THE MASK 2 INCHES AWAY FROM THE PACEMAKER AND IT WAS SUGGESTED THAT THE USER CONTACT HIS CARDIOLOGIST WITH ANY CONCERNS AND TO STOP USING THE MASK UNTIL HE DOES SO. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THERE WAS NO REPORT OF MEDICAL INTERVENTION. NO PRODUCT WILL BE RETURNED FOR INVESTIGATION. LABELING INCLUDED WITH THE MASK WARNS THE USER "THE HEADGEAR CLIPS AND MASK CUSHION CONTAIN MAGNETS. CONTACT YOUR HEALTHCARE PROVIDER BEFORE YOU USE THIS MASK. SOME MEDICAL DEVICES MAY BE AFFECTED BY MAGNETIC FIELDS. THE MAGNETIC CLIPS IN THIS MASK SHOULD BE KEPT AT LEAST 2 INCHES AWAY FROM ANY ACTIVE MEDICAL DEVICE WITH SPECIAL ATTENTION TO IMPLANTED DEVICES SUCH AS PACEMAKERS, DEFIBRILLATORS, AND COCHLEAR IMPLANTS. THE MANUFACTURER WILL CONTINUE TO MONITOR COMPLAINTS FOR SIMILAR ISSUES. THE MANUFACTURER CONCLUDES NO FURTHER ACTION IS NEEDED AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15760843 ↗
  • MalfunctionNov 9, 2022Patient Device Interaction Problem; Patient Device Interaction Problem

    THE MANUFACTURER RECEIVED INFORMATION THAT THE DREAMWEAR FULL FACE MASK WITH MAGNETIC CLIPS WHILE USING WITH A PACEMAKER. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THE MANUFACTURER HAS NOT RECEIVED THE PRODUCT FOR INVESTIGATION. THE DREAMWEAR INSTRUCTIONS FOR USE INCLUDES THE FOLLOWING WARNINGS: THE HEADGEAR CLIPS CONTAIN MAGNETS. CONTACT YOUR HEALTHCARE PROFESSIONAL BEFORE YOU USE THIS MASK. SOME MEDICAL DEVICES MAY BE AFFECTED BY MAGNETIC FIELDS. THE MAGNETIC CLIPS IN THIS MASK SHOULD BE KEPT AT LEAST 2 IN. (50 MM) AWAY FROM ANY ACTIVE MEDICAL DEVICE WITH SPECIAL ATTENTION TO IMPLANTED DEVICES SUCH AS PACEMAKERS, DEFIBRILLATORS AND COCHLEAR IMPLANTS. DO NOT USE IN OR NEAR MAGNETIC RESONANCE IMAGING (MRI) EQUIPMENT. THIS REPORT WILL BE FILED AS AN INITIAL-FINAL REPORT. WE WILL CONTINUE TO MONITOR THESE COMPLAINTS REGARDING MAGNETS. IF ANY ADDITIONAL INFORMATION IS RECEIVED, AN ADDITIONAL REPORT WILL BE FILED. THE MANUFACTURER CONCLUDES NO FURTHER ACTION IS NEEDED AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15760073 ↗

Other VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices

All VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page