Skip to content

SIMPLYGO SimplyGo, INTL, ITRecalls & reported problems

Respironics, Inc. · SIMPLYGO,INTERNATIONAL, IT · Class 2

All models of SimplyGo (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00606959048812), received between May 20, 2026 and May 20, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Overheating of Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Overheating of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 20, 2026Overheating of Device

    A SIMPLYGO, INTERNATIONAL, IT DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR EVALUATION. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. THERE WAS NO REPORT OF MEDICAL INTERVENTION. DURING THE EVALUATION OF THE DEVICE, THE THIRD-PARTY SERVICE CENTER VISUALLY INSPECTED THE DEVICE AND FOUND THAT THE PCA IS BURNT. SIEVES ARE CLOGGED; THEY HAVE TO BE REPLACED. INLET FILTER WILL BE REPLACED DUE TO PREVENTIVE MAINTENANCE. COMPRESSOR IS DEFECTIVE, IT HAS TO BE REPLACED. TUBING IS CONTAMINATED; IT HAS TO BE REPLACED. CHECK VALVE COVER IS CRACKED, IT HAS TO BE REPLACED. THERE ARE MISSING SCREWS ON THE REAR ENCLOSURE; THEY HAVE TO BE ADDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25237595 ↗

Other Generator, oxygen, portable devices

All Generator, oxygen, portable devices →

Sources and updates