The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00606959053168), received between Aug 13, 2021 and Jun 1, 2026 — classified as 2 injuries, 9 malfunctions.

The most frequently reported specific device problem is “Degraded” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Degraded11
  • Contamination5

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Asthma1
  • Chest Pain1
  • Dizziness1
  • Dyspnea1
  • Headache1
  • Pleural Effusion1
  • Pulmonary Edema1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 1, 2026Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A DREAMSTATION EXPERT DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION WAS FOUND. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST CONTAMINATION. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25351956 ↗
  • MalfunctionMay 28, 2026Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A DREAMSTATION EXPERT DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION WAS FOUND. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST CONTAMINATION. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25323875 ↗
  • MalfunctionMay 25, 2026Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A DREAMSTATION EXPERT DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION WAS FOUND. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST CONTAMINATION. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25272651 ↗
  • MalfunctionMar 3, 2025Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE/RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A REPLACEMENT DREAMSTATION EXPERT DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT HAS ALLEGED FOAM IN THE TUBING, WHICH WAS USED ON THE ORIGINAL DEVICE, AND THE WATER IS DIRTY AFTER USE. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21502896 ↗
  • MalfunctionSep 5, 2024Contamination; Degraded; Contamination; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO REP DREAMSTATION AUTO CPAP. THE PATIENT HAS ALLEGED VISUALIZATION OF BLACK SPECS. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THE DEVICE WAS RETURNED TO MANUFACTURER SERVICE CENTER. AN INTERNAL AND EXTERNAL VISUAL INSPECTION WAS COMPLETED AND FOUND WHITE FOAM PARTICLES IN THE AIRPATH. THERE WERE SECONDARY FINDINGS LIKE DUST PARTICLES. THE MANUFACTURER IS SUBMITTING A FINAL REPORT AT THIS TIME. IF PERTINENT INFORMATION BECOMES AVAILABLE TO THE MANUFACTURER AT A LATER DATE, AN ADDENDUM TO THIS FINAL REPORT WILL BE FILED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20152661 ↗
  • MalfunctionAug 15, 2023Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING HEADACHES WHILE USING THE DEVICE. MEDICAL INTERVENTION WAS NOT SPECIFIED. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: Headache

    MAUDE report 17550347 ↗
  • MalfunctionJan 20, 2022Contamination; Degraded; Contamination; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING A CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE'S SOUND ABATEMENT FOAM BECAME DEGRADED AND CAUSED A PATIENT TO DEVELOP A RESPIRATORY INFECTION. THE PATIENT WAS HOSPITALIZED IN RESPONSE TO THE REPORTED EVENT. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.

    Patient problems: Unspecified Respiratory Problem

    MAUDE report 13305783 ↗
  • MalfunctionDec 28, 2021Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE'S SOUND ABATEMENT FOAM. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13095786 ↗
  • InjuryDec 21, 2021Contamination; Degraded; Contamination; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING A CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE'S SOUND ABATEMENT FOAM BECAME DEGRADED AND CAUSED A PATIENT TO DEVELOP WATER IN THEIR LUNGS. THE PATIENT WAS HOSPITALIZED IN RESPONSE TO THE REPORTED EVENT. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.

    Patient problems: Dyspnea; Pleural Effusion; Pulmonary Edema

    MAUDE report 13051890 ↗
  • MalfunctionAug 24, 2021Contamination; Degraded; Contamination; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE'S SOUND ABATEMENT FOAM. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: Chest Pain; Dizziness; No Clinical Signs, Symptoms or Conditions; Respiratory Tract Infection

    MAUDE report 12360143 ↗
  • InjuryAug 13, 2021Contamination; Degraded; Contamination; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING A CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE'S SOUND ABATEMENT FOAM BECAME DEGRADED AND CAUSED A PATIENT TO DEVELOP ASTHMA. THERE IS NO REPORT OF THE MEDICAL INTERVENTION THAT THE PATIENT HAS RECEIVED AT THIS TIME. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: Asthma