Skip to content

DreamStation DreamStation CPAP Pro, USRecalls & reported problems

Respironics, Inc. · DreamStation CPAP Pro, USA · Class 2

All models of Philips DreamStation (190)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Respironics DreamStation CPAP Pro is a breathing machine. Its 92 reports, with 5 injuries and 1 death, mostly describe the sound-reducing foam inside the device breaking down, and one describes contamination. No recalls name this device. Reports for the last 12 months (6) are fewer than in the previous 12 months (77).

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
92
Deaths reported
1
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 92 adverse event reports that list this exact device (DI 00606959424128), received between Sep 12, 2024 and Feb 24, 2026 — classified as 1 death, 5 injuries, 86 malfunctions.

The most frequently reported specific device problem is “Degraded” (92 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 5 reportsOct 2024: 4 reportsNov 2024: 1 reportsDec 2024: 5 reportsJan 2025: 1 reportsFeb 2025: 31 reportsMar 2025: 27 reportsApr 2025: 5 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 5 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202631
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Degraded92
  • Contamination1
  • Smoking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions84
  • Cancer5
  • Insufficient Information3
  • Unspecified Respiratory Problem3
  • Dizziness2
  • Headache2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Cardiovascular Dysfunction/Insufficiency1

Recent adverse event reports

The 25 most recent of 92 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 24, 2026Degraded

    A DEVICE WAS RETURNED TO THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM DEGRADATION. THE DEVICE POWERED ON AND AIRFLOW WAS CONFIRMED. THE DEVICE'S DOWNLOADED LOGS WERE REVIEWED BY THE THIRD-PARTY SERVICE CENTER. THERE WERE NO ERROR CODES FOUND. THE THIRD-PARTY SERVICE CENTER CONCLUDES THAT VISIBLE FOAM DEGRADATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427300 ↗
  • MalfunctionFeb 19, 2026Degraded

    A DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN RELATION TO THE VOLUNTARY FIELD SAFETY NOTICE AND RECALL NOTIFICATION INVOLVING THE SOUND ABATEMENT FOAM USED IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION, THE SERVICE CENTER PERFORMED A VISUAL INSPECTION AND OBSERVED EVIDENCE OF FOAM DEGRADATION. THE DEVICE POWERED ON SUCCESSFULLY, AND AIRFLOW WAS VERIFIED. THE SERVICE CENTER CONCLUDED THAT THEY COULD CONFIRM THE CUSTOMER REPORTED ALLEGATION. HOWEVER, VISIBLE FOAM DEGRADATION WAS IDENTIFIED, AND THE DEVICE WAS SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24385950 ↗
  • MalfunctionFeb 18, 2026Degraded

    A DEVICE WAS RETURNED TO THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM DEGRADATION. THE DEVICE POWERED ON AND AIRFLOW WAS CONFIRMED. THE DEVICE'S DOWNLOADED LOGS WERE REVIEWED BY THE THIRD-PARTY SERVICE CENTER. THERE WERE NO ERROR CODES FOUND. THE THIRD-PARTY SERVICE CENTER CONCLUDES THAT THEY COULDN'T CONFIRM THE CUSTOMER'S ALLEGATION AND THERE WAS A VISIBLE FOAM DEGRADATION AND UNIT GOT SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24376014 ↗
  • MalfunctionFeb 9, 2026Degraded

    A DEVICE WAS RETURNED TO THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM DEGRADATION. THE DEVICE POWERED ON AND AIRFLOW WAS CONFIRMED. THE DEVICE'S DOWNLOADED LOGS WERE REVIEWED BY THE THIRD-PARTY SERVICE CENTER. THERE WERE ZERO CODE FOUND. THE THIRD-PARTY SERVICE CENTER CONCLUDES THAT THEY CONFIRM THE CUSTOMER'S ALLEGATION AND THERE WAS A VISIBLE FOAM DEGRADATION AND UNIT GOT SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24289695 ↗
  • MalfunctionFeb 4, 2026Degraded

    A DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND VISIBLE FOAM PARTICLES WERE OBSERVED. THE DEVICE WAS SCRAPPED. AT THIS TIME, NO FURTHER INVESTIGATION CAN BE PERFORMED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP REPORT WILL BE FILED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24258157 ↗
  • MalfunctionDec 11, 2025Degraded; Degraded

    A DREAMSTATION CPAP PRO DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER AS PART OF THE UNO RECALL PROCESS WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED ERROR CODES E022 (ERR_MOTOR_FLUX_MAGNITUDE), E020 (ERR_MOTOR_SPINUP_FLUX_LOW), E053 (ERR_COMP_LOG_SEM_TIMEOUT) AND E200 (ERR_RTOS_ABORT_ERROR). THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23778412 ↗
  • MalfunctionOct 22, 2025Degraded; Degraded

    A DREAMSTATION CPAP PRO DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER AS PART OF THE UNO RECALL PROCESS WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23363287 ↗
  • MalfunctionOct 15, 2025Degraded; Degraded

    A DREAMSTATION CPAP PRO DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER AS PART OF THE UNO RECALL PROCESS WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23295339 ↗
  • MalfunctionSep 16, 2025Degraded; Degraded

    A DREAMSTATION DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER AS PART OF THE UNO RECALL PROCESS WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23066560 ↗
  • MalfunctionSep 8, 2025Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING VISUALIZATION OF FOAM DEGRADATION INSIDE THE ASSEMBLY. MEDICAL INTERVENTION WAS NOT SPECIFIED. THE MANUFACTURER WAS MADE AWARE OF THIS COMPLAINT THROUGH A REPRESENTATIVE OF THE CUSTOMER. DURING THE EVALUATION OF THE DEVICE, AT THE MANUFACTURE'S THIRD-PARTY'S SERVICE CENTER THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM DEGRADATION INSIDE ASSEMBLY. THE DEVICE HAS BEEN SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22991370 ↗
  • MalfunctionJul 14, 2025Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING VISIBLE FOAM DEGRADATION THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS SPECIFIED AS REQUIRED BY THE PATIENT. DURING THE EVALUATION OF THE DEVICE, AT THE MANUFACTURE'S SERVICE CENTER THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM DEGRADATION. THE DEVICE HAS BEEN SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22498219 ↗
  • MalfunctionJun 13, 2025Degraded; Degraded

    A DEVICE WAS RETURNED TO A THIRD PARTY SERVICE CENTER/MANUFACTURER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM DEGRADATION. DURING THE EVALUATION, FOAM PARTICLES WERE OBSERVED. THE DEVICE WAS SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22206541 ↗
  • MalfunctionMay 27, 2025Degraded; Degraded

    THE DREAMSTATION CPAP PRO DEVICE WAS RETURNED TO A SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE/RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE DEVICE WAS VISUALLY INSPECTED DURING THE EVALUATION, AND EVIDENCE OF FOAM PARTICLES/DEGRADATION WAS FOUND. THE DEVICE WAS SCRAPPED. AT THIS TIME, NO FURTHER INVESTIGATION CAN BE PERFORMED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22088629 ↗
  • MalfunctionApr 28, 2025Degraded; Smoking; Degraded; Smoking

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT ALLEGES VISUALIZATION OF PARTICLES IN DEVICE. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21917647 ↗
  • MalfunctionApr 22, 2025Degraded; Degraded

    A DEVICE WAS RETURNED TO A THIRD PARTY SERVICE CENTER/MANUFACTURER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM DEGRADATION. DURING THE EVALUATION, FOAM PARTICLES WERE OBSERVED. THE DEVICE WAS SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21882969 ↗
  • MalfunctionApr 21, 2025Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT ALLEGES VISUALIZATION OF PARTICLES IN DEVICE. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21874923 ↗
  • MalfunctionApr 18, 2025Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE/RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT ALLEGES VISUALIZATION OF PARTICLES IN DEVICE. THERE WAS NO REPORT OF MEDICAL INTERVENTION. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21864729 ↗
  • MalfunctionApr 17, 2025Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE/RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT HAS ALLEGED VISUALIZATION OF PARTICLES. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. THE DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER. THE THIRD-PARTY SERVICE CENTER CONCLUDES THAT FOUND NO EVIDENCE OF FOAM DEGRADATION, THEY COULDN'T CONFIRM THE CUSTOMER'S ALLEGATION AND UNIT SCRAPPED. THE MANUFACTURER IS SUBMITTING A FINAL REPORT AT THIS TIME. IF PERTINENT INFORMATION BECOMES AVAILABLE TO THE MANUFACTURER AT A LATER DATE, AN ADDENDUM TO THIS FINAL REPORT WILL BE FILED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21856395 ↗
  • MalfunctionMar 21, 2025Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT ALLEGES VISUALIZATION OF PARTICLES IN DEVICE. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21661073 ↗
  • MalfunctionMar 20, 2025Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING VISUALIZATION OF PARTICLES RELATED TO CPAP DEVICE'S SOUND ABATEMENT FOAM. THERE WAS NO REPORT OF MEDICAL INTERVENTION. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. THE DEVICE WAS RETURNED TO THE THIRD-PARTY SERVICE CENTRE FOR FURTHER EVALUATION. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTRE, THE DEVICE WAS VISUALLY INSPECTED AND FOUND NO EVIDENCE OF VISIBLE FOAM PARTICLES. THE THIRD-PARTY SERVICE CENTRE CONCLUDES THAT THEY COULDN'T CONFIRM THE CUSTOMER'S ALLEGATION AND UNIT SCRAPPED. THE MANUFACTURER IS SUBMITTING A FINAL REPORT AT THIS TIME. IF PERTINENT INFORMATION BECOMES AVAILABLE TO THE MANUFACTURER AT A LATER DATE, AN ADDENDUM TO THIS FINAL REPORT WILL BE FILED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21652173 ↗
  • MalfunctionMar 19, 2025Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE/RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT HAS ALLEGED VISUALIZATION OF PARTICLES THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. THE DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER. THE TECHNICIAN CONFIRMED NO PARTICLES IN THE AIR PATH DURING THE DEVICE EVALUATION. THE THIRD-PARTY SERVICE CENTER CONCLUDES THAT THEY COULDN'T CONFIRM THE CUSTOMER'S ALLEGATION AND THE DEVICE WAS SCRAPPED. THE MANUFACTURER IS SUBMITTING A FINAL REPORT AT THIS TIME. IF PERTINENT INFORMATION BECOMES AVAILABLE TO THE MANUFACTURER AT A LATER DATE, AN ADDENDUM TO THIS FINAL REPORT WILL BE FILED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21639454 ↗
  • MalfunctionMar 17, 2025Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE/RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT HAS ALLEGED VISUALIZATION OF PARTICLES IN DEVICE. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21621447 ↗
  • MalfunctionMar 17, 2025Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING VISUALIZATION OF PARTICLES RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THERE WAS NO REPORT OF MEDICAL INTERVENTION. THE DEVICE WAS RETURNED TO THE THIRD-PARTY SERVICE CENTRE FOR FURTHER EVALUATION. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTRE, THE DEVICE WAS VISUALLY INSPECTED AND FOUND NO EVIDENCE OF VISIBLE FOAM PARTICLES. THE TECHNICIAN CONFIRMED NO PARTICLES IN THE AIR PATH. THE THIRD-PARTY SERVICE CENTRE CONCLUDES THAT THEY COULDN'T CONFIRM THE CUSTOMER'S ALLEGATION AND UNIT SCRAPPED. THE MANUFACTURER IS SUBMITTING A FINAL REPORT AT THIS TIME. IF PERTINENT INFORMATION BECOMES AVAILABLE TO THE MANUFACTURER AT A LATER DATE, AN ADDENDUM TO THIS FINAL REPORT WILL BE FILED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21617473 ↗
  • MalfunctionMar 17, 2025Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT ALLEGES VISUALIZATION OF PARTICLES IN DEVICE. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21617091 ↗
  • MalfunctionMar 17, 2025Degraded; Degraded

    THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT ALLEGES VISUALIZATION OF PARTICLES IN DEVICE. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21616908 ↗

Other VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices

All VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page