N/A 1600-7Recalls & reported problems
Salter Labs · Nasal Cannula (adult) Salter Style with 7’ (2.1 m) supply tube · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00607411100253), received between Oct 20, 2023 and Nov 6, 2025 — classified as 3 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Fire” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Fire2
- Appropriate Device Problem Term/Code Not Available1
- Material Split, Cut or Torn1
- Obstruction of Flow1
- Off-Label Use1
- Partial Blockage1
- Patient Device Interaction Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Burn(s)1
- Partial thickness (Second Degree) Burn1
- Skin Inflammation/ Irritation1
- Unspecified Respiratory Problem1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryNov 6, 2025Partial Blockage; Obstruction of Flow; Partial Blockage; Obstruction of Flow
IT WAS REPORTED THAT THE TUBING OF WHAT THEY RECEIVED IS EXPIRED BECAUSE IT'S DRIED OUT AND IT RESTRICTED THE OXYGEN FLOW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23493536 ↗InjuryJan 22, 2025Fire; Patient Device Interaction Problem; Fire; Patient Device Interaction Problem
IT WAS REPORTED THAT A PATIENT WAS SMOKING DURING INHALATION AND BURNED HIMSELF. ONE INCIDENT WAS REPORTED FOR THIS DEVICES. IT WAS STATEED THAT " ON (B)(6) 2024 AROUND 9 PM PATIENT INHALED OXYGEN PRODUCED BY PHILIPS -EVERFLO. PATIENT SMOKED DURING INHALATION AND BURNED HIMSELF. ACCORDING TO THE INFORMATION HE HAS 2ND DEGREE BURNS IN THE HEAD AND UPPER BODY. HIS CONDITION WAS STABILIZED, AND HE WAS HOSPITALIZED IN THE HOSPITAL IN PILSEN. ACCORDING TO THE INFORMATION, THE FIRE BRIGADE WAS PRESENT AT THE SCENE. CHIEF PHYSICIAN REFERRED THE PATIENT AND AGREED WITH HOMECARE SERVICE MANAGER TO STOP THE OXYGEN TREATMENT."
Patient problems: Burn(s); Partial thickness (Second Degree) Burn
MAUDE report 21211811 ↗MalfunctionSep 4, 2024Fire; Off-Label Use; Fire; Off-Label Use
IT WAS REPORTED, THE PATIENT LIT A CANDLE NEXT TO HER OXYGEN GOGGLES, DESPITE THE TRAINING SHE RECEIVED WHEN THE DEVICE WAS INSTALLED. AS A RESULT, THE OXYGEN CONCENTRATOR AND ALL THE HOSES LEADING TO THE DEVICE WERE BURNED AND DAMAGED. THE PATIENT WAS UNHARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20141066 ↗InjuryOct 20, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
WHEN HIS OXYGEN WAS INSTALLED, HE WAS PROVIDED WITH GLASSES WHICH CAUSED HIM AN ALLERGY, WITH ITCHY ECZEMA-LIKE LESIONS AT THE LEVEL OF THE TUBING (UNDER THE CHIN, AND IN THE NECK). NO CHANGE IN COSMETIC PRODUCT HABITS THAT COULD EXPLAIN AN ALLERGY. NO MISUSE OF THE GLASSES. THE PATIENT SAW THE LESIONS DISAPPEAR AS SOON AS SHE CHANGED HER GLASSES MODEL.
Patient problems: Skin Inflammation/ Irritation
MAUDE report 17978299 ↗MalfunctionOct 20, 2023Material Split, Cut or Torn; Material Split, Cut or Torn
"THE NASAL TIP BECAME SEPARATED FROM THE TUBES DURING THE NIGHT. THE TELESCOPE WAS INSTALLED NEW ON SUNDAY EVENING (B)(6) BEFORE BEDTIME AND THE INCIDENT OCCURRED ON THE NIGHT OF JUNE (B)(6). THE BATCH NUMBER HAS NOT BEEN PRESERVED. NO CLINICAL CONSEQUENCES."
Patient problems: Unspecified Respiratory Problem
MAUDE report 17978264 ↗
FDA clearance and classification
The FDA regulates product code CAT (cannula, nasal, oxygen) under 21 CFR 868.5340 as a Class I device. 21 CFR 868.5340 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the N/A 1600-7
Has the N/A 1600-7 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00607411100253), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the N/A 1600-7?
The FDA's MAUDE database lists 5 adverse event reports for this device, received between Oct 20, 2023 and Nov 6, 2025. In those reports, the most common specific device problems are “Fire” (2), “Material Split, Cut or Torn” (1) and “Obstruction of Flow” (1); the most common patient problems are “Burn(s)” (1), “Partial thickness (Second Degree) Burn” (1) and “Skin Inflammation/ Irritation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the N/A 1600-7?
None of the reports are classified as deaths, 3 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the N/A 1600-7 FDA approved?
Devices of this type (product code CAT) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Salter Labs.
Who makes the N/A 1600-7, and is it still sold?
The FDA's device database lists Salter Labs as the company responsible for it (brand N/A, model 1600-7). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CANNULA, NASAL, OXYGEN devices
- Fisher & Paykel Healthcare OPT844Fisher & Paykel Healthcare Limited · Cannula, Nasal, Oxygen39 reports
- Fisher and Paykel Healthcare OPT312Fisher & Paykel Healthcare Limited · Cannula, Nasal, Oxygen21 reports
- Fisher and Paykel Healthcare OPT314Fisher & Paykel Healthcare Limited · Cannula, Nasal, Oxygen17 reports
- O2 Cannula 001366Salter Labs · Cannula, Nasal, Oxygen9 reports
- N/A 16SOFT-4Salter Labs · CANNULA, NASAL, OXYGEN8 reports
- N/A 1053-7Salter Labs · CANNULA, NASAL, OXYGEN8 reports
- Fisher & Paykel Healthcare OPT842Fisher & Paykel Healthcare Limited · Cannula, Nasal, Oxygen7 reports
- O2 Cannula 001365Salter Labs · Cannula, Nasal, Oxygen7 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page