Parker H-PFST-70Recalls & reported problems
Salter Labs · Parker Flex-Tip Tracheal Tube with Stylet, High Volume Low Pressure, Cuffed (PFST), 7.0 mm Parker Original DID: 10749065801708 · Class 2
All models of Parker (13)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00607411958014), received between Jul 10, 2018 and Nov 14, 2025 — classified as 1 death, 4 malfunctions.
The most frequently reported specific device problem is “Gas/Air Leak” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K984528); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Gas/Air Leak2
- Material Integrity Problem2
- Deflation Problem1
- Incomplete or Inadequate Connection1
- Loose or Intermittent Connection1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Death1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 14, 2025Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT IMMEDIATELY AFTER INTUBATION, AN AIR LEAK OCCURRED. THE TUBE WAS REMOVED, AND THE PATIENT WAS REINTUBATED WITH A NEW TUBE. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23560518 ↗MalfunctionSep 25, 2025Deflation Problem; Deflation Problem
IT WAS REPORTED, THE PILOT BALLOON DEFLATED IMMEDIATELY AFTER INTUBATION, (THERE WAS NO PROBLEM WITH THE CUFF TEST PRIOR TO INTUBATION.) THE PATIENT WAS REINTUBATED, THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23151445 ↗MalfunctionSep 27, 2024Loose or Intermittent Connection; Material Integrity Problem; Incomplete or Inadequate Connection; Loose or Intermittent Connection; Material Integrity Problem; Incomplete or Inadequate Connection
CONNECTION BETWEEN THE PILOT BALLOON AND THE INFLATING TUBE WAS RIPPED OFF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20326480 ↗MalfunctionSep 6, 2024Gas/Air Leak; Gas/Air Leak
THERE IS A LEAK FROM THE CONNECTION BETWEEN THE INFLATING TUBE AND THE PILOT BALLOON.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20163915 ↗DeathJul 10, 2018Material Integrity Problem; Material Integrity Problem
PATIENT UNDERWENT SURGERY AND SUBSEQUENTLY DEVELOPED DECREASED OXYGEN SATURATION. X-RAY REVEALED PNEUMOMEDIASTINUM. PATIENT WAS INTUBATED AND PLACED ON VENTILATOR. PATIENT WAS DETERMINED TO HAVE A SUBACUTE RIGHT ACA INFARCT AND CHRONIC RIGHT MCA INFARCT. PATIENT EXPIRED ON (B)(6) 2018.
Patient problems: Death
MAUDE report 7674163 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K984528. One submission often covers many models, so check the product and lot details in each record.
- Z-1399-2017 · Dec 16, 2016 · TerminatedParker Left Double Lumen Endobronchial Tube with Parker Flex-Tip (Parker Endo-Bronch with Flex Tip). Model Numbers: H-PFTDL-28-1, H-PFTDL-32-1, H-PFTDL-35-1, H-PFTDL-37-1, H-PFTDL-
Other Tube, Tracheal (W/Wo Connector) devices
- Shiley 18875Covidien LP · Tube, tracheal (w/wo connector)4 deaths570 reports
- RUSCH IPN041476Teleflex Incorporated · Tube, tracheal (w/wo connector)10 recalls0 reports
- Shiley 18775Covidien LP · Tube, tracheal (w/wo connector)489 reports
- Shiley 18770Covidien LP · Tube, tracheal (w/wo connector)2 deaths434 reports
- Shiley 18880Covidien LP · Tube, tracheal (w/wo connector)2 deaths370 reports
- Shiley 18780Covidien LP · Tube, tracheal (w/wo connector)3 deaths286 reports
- Shiley 18770SCovidien LP · Tube, tracheal (w/wo connector)273 reports
- Shiley 18780SCovidien LP · Tube, tracheal (w/wo connector)251 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page