Skip to content

Epix Laparoscopic Grasper C4130Recalls & reported problems

Applied Medical Resources Corporation · 5mm x 35cm Disposable Grasper · Class 1

All models of Epix Laparoscopic Grasper (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Applied Medical Epix laparoscopic grasper is a disposable 5mm instrument used to hold and pull tissue during minimally invasive surgery. The 35 reports mostly describe the grasper fraying, its jaw breaking off, locking up or sticking to tissue, and pads with sharp edges that tore or injured the stomach and bowel. Two reports involve injuries and no deaths are described. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
34
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 00607915110147), received between Feb 24, 2016 and May 1, 2026 — classified as 2 injuries, 32 malfunctions.

The most frequently reported specific device problem is “Material Frayed” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 6 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Frayed7
  • Detachment of Device or Device Component5
  • Mechanical Jam5
  • Detachment Of Device Component3
  • Material Fragmentation3
  • Break2
  • Entrapment of Device2
  • Insufficient Device Problem Information2

Patient problems

  • No Clinical Signs, Symptoms or Conditions16
  • No Consequences Or Impact To Patient10
  • Insufficient Information2
  • No Known Impact Or Consequence To Patient2
  • Physical Entrapment2
  • Bowel Perforation1
  • Hemorrhage/Blood Loss/Bleeding1
  • Inflammation1

Recent adverse event reports

The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 1, 2026Material Fragmentation

    PROCEDURE PERFORMED: LAP CHOL EVENT DESCRIPTION: THE DEVICE WAS BEING USED DURING PULLING TISSUE. PULLING AWAY TISSUE WITH FORCE THE JAW BROKE OFF. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA EMAIL ON 09APR26: THE JAWS FULLY DETACHED AND ALL THE MATERIALS WERE RETRIEVED FROM THE PATIENT. NO INFORMATION ON THE FOLLOWING, WHAT WAS THE SURGEON GRASPING WHEN THE BREAKAGE OCCURRED. INTERVENTION: DEVICE REPLACEMENT PATIENT STATUS: PATIENT IS OK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25062232 ↗
  • MalfunctionMar 10, 2026Sharp Edges

    PROCEDURE PERFORMED: NISSEN FUNDOPLICATION. EVENT DESCRIPTION: THE DEVICE WAS USED FOR GRASPING THE STOMACH DURING THE STOMACH MOBILIZATION. AS SOON AS THE DEVICE GRASPED THE STOMACH THERE WAS INJURY AND BLEEDING OF THE STOMACH WHICH WAS CAUSED, ACCORDING TO THE SURGEON'S OPINION, BY THE TRAUMATIC LATIS PADS. THERE WAS NOT ANY CLINICAL IMPACT TO THE PATIENT OTHER THAN BLEEDING, WHICH WAS IMMEDIATELY CONTROLLED BY THE SURGEON, BY STITCHING THE TRAUMA. A DEVICE OF ANOTHER MANUFACTURER WAS USED TO CONTINUE THE SURGICAL OPERATION. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA EMAIL ON 20FEB26: THE CORRECT LOT NR. IS 1563295. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA EMAIL ON 27FEB26: THE BLEEDING OF THE STOMACH LINING OCCURRED AS A RESULT OF USING THE FORCEPS ON THE STOMACH. THEY STITCHED UP THE STOMACH AND CONTINUED THE SURGERY. THE PATIENT¿S CONDITION IS FINE. NO OTHER PRODUCT WAS INVOLVED. THE INCIDENT OCCURRED IMMEDIATELY AFTER THE FORCEPS GRASPED THE STOMACH WALL. RIPS, TEARS OR FRAYS WEREN'T OBSERVED ON THE PAD BEFORE GRASPING. INTERVENTION: A DEVICE OF ANOTHER MANUFACTURER WAS USED TO CONTINUE THE SURGICAL OPERATION PATIENT

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 24565310 ↗
  • MalfunctionFeb 7, 2025Material Frayed; Sharp Edges; Material Frayed; Sharp Edges

    PROCEDURE PERFORMED: APPENDIX. EVENT DESCRIPTION: THE GRASPER REPEATEDLY GET CAUGHT ON THE BOWEL DURING SURGERY AND THUS CAUSE INJURIES. ADDITIONAL INFORMATION RECEIVED FROM COMPANY REP. VIA EMAIL ON 04FEB25: THE GRASPER GOT CAUGHT AT THE PROXIMAL END OF THE PADDING. NO INJURY. THE INTESTINE HAS BECOME STUCK AT THE PROXIMAL END OF THE PADDING (SEVERAL TIMES DURING SURGERY). INTERVENTION: INTESTINE SHAKEN OFF THE GRASPING. PATIENT STATUS: NO INJURY. PATIENT IS GOOD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21337184 ↗
  • MalfunctionDec 13, 2024Difficult to Open or Close; Difficult to Open or Close

    PROCEDURE PERFORMED: EVENT DESCRIPTION: TWO OF TWO COMPLAINTS FOR THIS REPORT: (B)(4) (COMPLAINT #1) (B)(4) (COMPLAINT #2) GRASPER WAS BEING USED TO REMOVE THE GALL BLADDER. SURGEON HEARD A CRACK WHEN CLOSING THE HANDLE AND THE GRASPER LOCKED UP. SURGEON GRABBED A SECOND GRASPER AND THE JAWS WOULD NOT CLOSE ALL THE WAY. THE THIRD GRASPER WORKED AND WAS USED TO COMPLETE THE CASE. ACCOUNT MANAGER WAS ONSITE AND CONFIRMED THE FINDINGS OF THE SURGEON FOR THE FIRST TWO GRASPERS. NO PATIENT INJURY. PRODUCT IS AVAILABLE FOR RETURN. ADDITIONAL INFORMATION PROVIDED BY [ACCOUNT MANAGER] ON 28FEB2024 VIA EMAIL: I SPOKE WITH THE CUSTOMER AGAIN LAST NIGHT AND SHE INFORMED ME THAT THE INCIDENT OCCURRED ON MONDAY 2/26, NOT YESTERDAY. THANK YOU FOR YOU ASSISTANCE. PLEASE LET ME KNOW IF THERE IS ANYTHING ELSE I CAN ASSIST WITH. ADDITIONAL INFORMATION PROVIDED BY [ACCOUNT MANAGER] ON 28FEB2024 VIA EMAIL: THE JAWS DID NOT DETACH OR BREAK OFF. NO MATERIALS WERE LOST IN THE PATIENT. THE SURGEON WAS GRASPING THE GALLBLADDER WHEN HE HEARD A SNAP SOUND COME FROM THE HANDLE. HE WAS ABLE TO RELEASE THE GALLBLADDER AT WHICH POINT A SECOND GRASPER WAS OPENED. THE SECOND GRASPER LOCKED UP WITHOUT TISSUE IN THE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20932112 ↗
  • MalfunctionDec 13, 2024Material Frayed; Material Frayed

    PROCEDURE PERFORMED: ROBOTIC LAR. EVENT DESCRIPTION: COMPLAINT 1 OF 2: (B)(4), COMPLAINT 2 OF 2: (B)(4). [HOSPITAL] AFTER USING THIS INSTRUMENT FOR A PERIOD OF TIME, THE FRONT TIP WILL BECOME ADHESIVE TO TISSUES. CHANGING TO ANOTHER INSTRUMENT FOR USE WILL STILL RESULT IN ADHESION AT THE FRONT TIP, UNTIL THE THIRD GRASPER IS USED TO COMPLETE THE SURGERY. THERE IS NO IMPACT TO PATIENT. USER REPLACE WITH NEW GRASPER TO COMPLETE THIS SURGERY. THERE IS NO OTHER INSTRUMENTS TO AFFECT THIS EVENT. PRODUCT IS AVAILABLE FOR RETURN. ADDITIONAL INFORMATION RECEIVED ON 15AUG2023 VIA EMAIL FROM [DISTRIBUTOR]: INTESTINE TISSUE WAS BEING GRASPED AT THE TIME OF THE INCIDENT. INTERVENTION: USER REPLACE WITH NEW GRASPER TO COMPLETE THIS SURGERY. PATIENT STATUS: FINE

    Patient problems: Physical Entrapment

    MAUDE report 20931951 ↗
  • MalfunctionDec 13, 2024Material Frayed; Material Frayed

    PROCEDURE PERFORMED: ROBOTIC LAR. EVENT DESCRIPTION: COMPLAINT 1 OF 2: #: (B)(4). COMPLAINT 2 OF 2: #: (B)(4). [HOSPITAL]. AFTER USING THIS INSTRUMENT FOR A PERIOD OF TIME, THE FRONT TIP WILL BECOME ADHESIVE TO TISSUES. CHANGING TO ANOTHER INSTRUMENT FOR USE WILL STILL RESULT IN ADHESION AT THE FRONT TIP, UNTIL THE THIRD GRASPER IS USED TO COMPLETE THE SURGERY. THERE IS NO IMPACT TO PATIENT. USER REPLACE WITH NEW GRASPER TO COMPLETE THIS SURGERY. THERE IS NO OTHER INSTRUMENTS TO EFFECT THIS EVENT. PRODUCT IS AVAILABLE FOR RETURN. ADDITIONAL INFORMATION RECEIVED ON 15AUG2023 VIA EMAIL FROM [DISTRIBUTOR]: INTESTINE TISSUE WAS BEING GRASPED AT THE TIME OF THE INCIDENT. INTERVENTION: USER REPLACE WITH NEW GRASPER TO COMPLETE THIS SURGERY. PATIENT STATUS: FINE.

    Patient problems: Physical Entrapment

    MAUDE report 20931925 ↗
  • MalfunctionDec 13, 2024Material Frayed; Material Frayed

    PROCEDURE PERFORMED: UNKNOWN EVENT DESCRIPTION: SCRUB NURSE NOTICED LOOSE THREAD(S) ON LATIS PAD WHEN HANDED. THE DEVICE WAS SET ASIDE AND NOT USED FOR THE PROCEDURE. NO CLINICAL IMPACT TO THE PATIENT. OPENED A NEW GRASPER. INTERVENTION: DEVICE REPLACEMENT PATIENT STATUS: NA (BEFORE USE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20931825 ↗
  • MalfunctionDec 3, 2024No Apparent Adverse Event; Insufficient Device Problem Information; No Apparent Adverse Event; Insufficient Device Problem Information

    PROCEDURE PERFORMED: NA EVENT DESCRIPTION: COMPLAINT (B)(4), COMPLAINT (B)(4). THE INFORMATION WAS COLLECTED AS FEEDBACK IN A MEETING WITH THE SURGEON. THE FEEDBACK SUBMISSION NUMBER IS 7139. THE DOCTOR SAID THAT THE LATIS PADS HAVE CHANGED SINCE THE MIM UPDATE. THEY ARE NOW 'LIKE SANDPAPER' AND HAS STOPPED USING THEM ALTOGETHER NOW. DR [NAME] WAS A HIGH-VOLUME USER PREVIOUSLY. ADDITIONAL INFORMATION WAS PROVIDED BY [NAME], AMA TERRITORY MANAGER VIA EMAIL ON 16-DEC-2024. "IT WAS JUST FEAR THE SURGEON HAD WHICH KEPT HIM FROM USING THE DEVICES." ADDITIONAL INFORMATION WAS PROVIDED BY [NAME], AMA TERRITORY MANAGER VIA EMAIL ON 23-DEC-2024. "YES, I REPLIED TO THIS ONE WITH THE SECOND COMPLAINT FOLLOW-UP. BOTH CASES DID NOT INVOLVE A PATIENT, THESE WERE JUST THE THOUGHTS OF BOTH SEPARATE SURGEONS ON EACH REPORT." PATIENT STATUS: NA INTERVENTION: NA

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20838524 ↗
  • MalfunctionDec 3, 2024No Apparent Adverse Event; Insufficient Device Problem Information; No Apparent Adverse Event; Insufficient Device Problem Information

    PROCEDURE PERFORMED: NA EVENT DESCRIPTION: COMPLAINT (B)(4), COMPLAINT (B)(4). THE INFORMATION WAS COLLECTED AS FEEDBACK IN A MEETING WITH THE SURGEON. THE FEEDBACK SUBMISSION NUMBER IS 7138. SURGEON FELT THAT THE LATIS PADS ARE EXTREMELY ROUGH AND STICKY AND CAUSES 'HOLES IN THE BOWEL'. WILL NOT USE THEM AND CONTRAINDICATES THEIR USE IN COLORECTAL SURGERY. THE DEVICE WAS BEING USED FOR GRASPING BOWEL. POTENTIAL HOLES IN BOWEL. THIS IS A GENERIC COMPLAINT, REGARDING THE NEW ALTERATIONS TO THE GRASPER SINCE THE MIM UPGRADE WAS IMPLEMENTED. DOCTOR STATED THAT THIS WAS ABSOLUTELY NOT A DEVICE TO BE USED ON THE BOWEL DUE TO RISK OF TEARING. IT IS UNKNOWN IF THERE WAS ANY CLINICAL IMPACT TO THE PATIENT AS A RESULT OF THE EVENT. IT IS UNKNOWN HOW THE USER RESOLVED THE SITUATION. ADDITIONAL INFORMATION WAS PROVIDED BY [NAME], (B)(4) TERRITORY MANAGER VIA EMAIL ON 16-DEC-2024. "IT WAS JUST A FEAR THE SURGEON HAD WHICH KEPT HIM FROM USING THE DEVICES." ADDITIONAL INFORMATION WAS PROVIDED BY [NAME], (B)(4) TERRITORY MANAGER VIA EMAIL ON 23-DEC-2024. "YES, I REPLIED TO THIS ONE WITH THE SECOND COMPLAINT FOLLOW-UP. BOTH CASES DID NOT INVOLVE A PATIENT, THESE WERE JUST THE THOUGHTS OF BOTH SEPA

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20838179 ↗
  • MalfunctionOct 3, 2024Material Separation; Material Separation

    PROCEDURE PERFORMED: LOWER ANTERIOR RESECTION WITH DA VINCI ROBOTIC SURGERY. EVENT DESCRIPTION: [HOSPITAL] "THE SURGEON DISCOVERED THAT THE TIP OF THE INSTRUMENT WAS BROKEN AND UNUSABLE UPON INITIAL USE." ADDITIONAL INFORMATION RECEIVED FROM [DISTRIBUTOR] VIA EMAIL ON 10SEP2024 "JAW DETACHED. MALFUNCTION NOTICED OR HAPPENED BEGINNING OF THE SURGERY." ADDITIONAL INFORMATION OBTAINED FROM [DISTRIBUTOR] VIA EMAIL ON 22NOV2024: AS THE INCIDENT OCCURRED LONG AGO, THE USER NO LONGER RECALLS THE DETAILS, AND THE UNIT HAS BEEN RETURNED TO APPLIED, MAKING IT IMPOSSIBLE FOR US TO VERIFY THE ISSUE. INTERVENTION: THE CASE WAS COMPLETED WITH THE NEW ONE. PATIENT STATUS: NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20359178 ↗
  • InjuryAug 21, 2024Material Frayed; Material Frayed

    PROCEDURE PERFORMED: APPENDECTOMY EVENT DESCRIPTION: COMPLAINT CREATED BASED OFF OF MEDWATCH UF/IMPORTER REPORT # (B)(4). [HOSPITAL] "UPON USING EPIX LAPAROSCOPIC GRASPER (REF# C4130, LOT #1 512170), A THREAD ON THE PAD, AT THE TIP OF THE GRASPER BECAME CAUGHT UPON THE INTESTINES WHEN USING THE GRASPER, DURING REMOVAL OF THE APPENDIX. IRRITATION AND REDNESS SEEN WHERE THE THREAD WAS, UPON USAGE OF THE GRASPER. SURGEON ASSESSED THE AREA OF THE INTESTINES AND DETERMINED THAT THERE WAS NO ACTIVE BLEEDING AND PROCESSED WITH APPENDECTOMY." ADDITIONAL INFORMATION RECEIVED FROM [CUSTOMER] VIA EMAIL ON 30JUL2024: [HOSPITAL] DOES PURCHASE DEVICES DIRECTLY FROM AMR. THE DEPARTMENT NO LONGER HAS THE DEVICE, DEVICE WAS DISCHARGED. THE PRODUCT IS NOT AVAILABLE FOR RETURN. ADDITIONAL INFORMATION RECEIVED FROM [CUSTOMER] VIA EMAIL ON 30JUL2024: THE GRASPERS THAT THE HOSPITAL PURCHASES DO NOT COME AS PART OF A KIT. THE HOSPITAL PURCHASES APPLIED MEDICAL GRASPERS BY THE BOX, QTY (B)(4). ADDITIONAL INFORMATION RECEIVED FROM [CUSTOMER] VIA EMAIL ON 1AUG2024: THE COMPLAINT OCCURRED AT [HOSPITAL] WHICH IS UNDER [HOSPITAL]'S LICENSE. [HOSPITAL] IS [HOSPITAL]'S CAMPUS IN [LOCATION]. INTERVENTION: SURGEON

    Patient problems: Inflammation

    MAUDE report 20019033 ↗
  • MalfunctionMay 7, 2024Break; Break

    PROCEDURE PERFORMED: LAPAROSCOPIC SALPINGO OOPHORECTOMY. EVENT DESCRIPTION: DURING THE PROCEDURE, THE SHAFT OF THE DEVICE WAS BENT. IT WAS NOTED AS A VISIBLE BREAK BUT WITH ALL PARTS INTACT. NO REMNANTS OF THE DEVICE FELL INTO THE PATIENT. THE CASE WAS COMPLETED WITH THE SAME DEVICE AND HAD NO OTHER PROBLEMS. THERE WAS NO PATIENT INJURY AND THE PATIENT IS DOING FINE. PRODUCT IS AVAILABLE FOR RETURN. PRODUCT UPDATED TO NPR: ADDITIONAL INFORMATION RECEIVED VIA MEDWATCH #MW5147682 ON 29DEC2023: TODAY WE HAD A DISPOSABLE EPIX LAPAROSCOPIC GRASPER BREAK DURING USE. ALL PARTS OF THE INSTRUMENT ARE INTACT (VISIBLE BREAK), AND NO REMNANTS OF THE INSTRUMENT WERE FOUND OR LEFT IN THE PATIENT. PATIENT STATUS: NO PATIENT INJURY AND THE PATIENT IS DOING FINE. INTERVENTION: THE CASE WAS COMPLETED WITH THE SAME DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19265509 ↗
  • MalfunctionApr 10, 2024Mechanical Jam; Mechanical Jam

    PROCEDURE PERFORMED: LAP CHOLECYSTECTOMY EVENT DESCRIPTION: GRASPER WAS BEING USED TO REMOVE THE GALL BLADDER. SURGEON HEARD A CRACK WHEN CLOSING THE HANDLE AND THE GRASPER LOCKED UP. SURGEON GRABBED A SECOND GRASPER AND THE JAWS WOULD NOT CLOSE ALL THE WAY. THE THIRD GRASPER WORKED AND WAS USED TO COMPLETE THE CASE. ACCOUNT MANAGER WAS ONSITE AND CONFIRMED THE FINDINGS OF THE SURGEON FOR THE FIRST TWO GRASPERS. NO PATIENT INJURY. PRODUCT IS AVAILABLE FOR RETURN. ADDITIONAL INFORMATION PROVIDED BY [NAME] ON 28FEB2024 VIA EMAIL: I SPOKE WITH THE CUSTOMER AGAIN LAST NIGHT AND SHE INFORMED ME THAT THE INCIDENT OCCURRED ON MONDAY 2/26, NOT YESTERDAY. THANK YOU FOR YOU ASSISTANCE. PLEASE LET ME KNOW IF THERE IS ANYTHING ELSE I CAN ASSIST WITH. ADDITIONAL INFORMATION PROVIDED BY [NAME] ON 28FEB2024 VIA EMAIL: THE JAWS DID NOT DETACH OR BREAK OFF. NO MATERIALS WERE LOST IN THE PATIENT. THE SURGEON WAS GRASPING THE GALLBLADDER WHEN HE HEARD A SNAP SOUND COME FROM THE HANDLE. HE WAS ABLE TO RELEASE THE GALLBLADDER AT WHICH POINT A SECOND GRASPER WAS OPENED. THE SECOND GRASPER LOCKED UP WITHOUT TISSUE IN THE JAWS AND WAS REMOVED. CASE WAS COMPLETED USING A THIRD GRASPER. THEY HAD ONE GRASPER

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19084022 ↗
  • MalfunctionNov 9, 2023Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    NAME OF PROCEDURE BEING PERFORMED: NA (BEFORE USE). DETAILED DESCRIPTION OF EVENT 13-OCT-2023. COMPLAINT PRODUCT FAMILY: GRASPER, PRODUCT NAME: EPIX GRASPER, MODEL NUMBER: C4130, LOT NUMBER: 1468046. THE EVENT OCCURRED BEFORE USE OF THE DEVICE, FOR AN UNKNOWN PROCEDURE. THE NURSE HAD IDENTIFIED WHILST DOING SET-UP. IT WAS REPORTED THAT HAIR HAD BEEN FOUND INSIDE THE SEALED PACKAGE OF THE DEVICE. THERE WAS NO CLINICAL IMPACT TO THE PATIENT AS A RESULT OF BEFORE USE OF THE PROCEDURE. THE USER RESOLVED THE SITUATION BY USING ANOTHER DEVICE FROM THE SHELF. NO OTHER INSTRUMENTS WERE USED WHEN THE COMPLAINT EVENT OCCURRED. PATIENT STATUS: NA (NO PATIENT INVOLVEMENT). TYPE OF INTERVENTION: NA (BEFORE USE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18107291 ↗
  • MalfunctionMar 22, 2023Material Frayed; Detachment of Device or Device Component; Material Frayed; Detachment of Device or Device Component

    PROCEDURE PERFORMED: LAP GEN SURG EVENT DESCRIPTION: EVENT 1 OF 2: 2023-000613 [MFR #2027111-2023-00392] EVENT 2 OF 2: 2023-000614 [MFR #2027111-2023-00393] IN BOTH PROCEDURES THE DEVICES WERE BEING USED TO MOVE AND MANIPULATE INTERNAL ORGANS. BOTH PROCEDURES WHERE THE DEVICES FAILED WERE BEING USED BY DIFFERENT SURGEON. ONE INCIDENT WAS 6/8 WEEKS AGO AND THE OTHER WAS 2 WEEKS AGO. NO CHANGES IN HEALTH RESULTING FROM THE EVENT. ONE PROCEDURE ONE OF THE PADS FELL OFF AND WAS RETRIEVED BY THE DEVICE ITSELF. THE OTHER PROCEDURE THE DEVICE PAD/S STARTED TO FRAY AND AGAIN WAS RETRIEVED BY THE DEVICE. THE FRAYED PART IS INCLUDED IN THE GRASPER PACKAGING. NO DETAILS AVAILABLE ON WHAT OTHER INSTRUMENTS WERE USED WHEN THE COMPLAINT EVENT OCCURRED. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA EMAIL 16MAR23: IT IS THE REPRESENTATIVE'S UNDERSTANDING THAT A WHOLE PAD CAME OFF AND WAS RETRIEVED. IT WAS NOT WITH THE DEVICE WHEN IT WAS GIVEN TO THE REPRESENTATIVE. ADDITIONAL INFORMATION RECEIVED FROM IRIC IN SCOTLAND 23MAR23: THE EVENT HAPPENED ON MARCH 1, 2023. APPLIED MEDICAL 'PAD' FRAYED DURING USE ON PATIENT. POTENTIAL FOR FRAYED THREADS TO GET LOST INSIDE THE PATIEN

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16594365 ↗
  • MalfunctionMar 22, 2023Loss of or Failure to Bond; Detachment of Device or Device Component; Loss of or Failure to Bond; Detachment of Device or Device Component

    PROCEDURE PERFORMED: LAP GEN SURG EVENT DESCRIPTION: EVENT 1 OF 2: 2023-000613 [MFR #2027111-2023-00392] EVENT 2 OF 2: 2023-000614 [MFR #2027111-2023-00393] IN BOTH PROCEDURES THE DEVICES WERE BEING USED TO MOVE AND MANIPULATE INTERNAL ORGANS. BOTH PROCEDURES WHERE THE DEVICES FAILED WERE BEING USED BY DIFFERENT SURGEON. ONE INCIDENT WAS 6/8 WEEKS AGO AND THE OTHER WAS 2 WEEKS AGO. NO CHANGES IN HEALTH RESULTING FROM THE EVENT. ONE PROCEDURE ONE OF THE PADS FELL OFF AND WAS RETRIEVED BY THE DEVICE ITSELF. THE OTHER PROCEDURE THE DEVICE PAD/S STARTED TO FRAY AND AGAIN WAS RETRIEVED BY THE DEVICE. THE FRAYED PART IS INCLUDED IN THE GRASPER PACKAGING. NO DETAILS AVAILABLE ON WHAT OTHER INSTRUMENTS WERE USED WHEN THE COMPLAINT EVENT OCCURRED. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA EMAIL 16MAR23: IT IS THE REPRESENTATIVE'S UNDERSTANDING THAT A WHOLE PAD CAME OFF AND WAS RETRIEVED. IT WAS NOT WITH THE DEVICE WHEN IT WAS GIVEN TO THE REPRESENTATIVE. INTERVENTION: ZERO ISSUES PATIENT STATUS: RETRIEVED BY THE DEVICE ITSELF

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16593787 ↗
  • MalfunctionFeb 16, 2023Entrapment of Device; Material Frayed; Material Split, Cut or Torn; Physical Resistance/Sticking; Entrapment of Device; Material Frayed; Material Split, Cut or Torn; Physical Resistance/Sticking

    PROCEDURE PERFORMED: APPENDECTOMY. EVENT DESCRIPTION: TODAY (ON (B)(6) 2023) I GOT ANOTHER CER FROM THE SAME HOSPITAL. THE JAWS OF THE GRASPING FORCEPS STUCK TO THE TISSUE SEVERAL TIMES DURING THE PROCEDURE. ONE SUSPECTS AN UNCLEANLY MANUFACTURED JAWS. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE CONFIRMED IN PERSON ON (B)(6) 2023: THE CUSTOMER IS RETURNING THE DEVICE. INCIDENTS (B)(4) ARE IDENTICAL IN TERMS OF HOW THE JAWS WERE STUCK TO THE TISSUE. STITCHED OVER THE TISSUE FOR SECURITY. THE TISSUE CAME LOOSE AFTER SEVERAL ATTEMPTS TO PULL ON IT. INTERVENTION: STITCHED OVER THE TISSUE FOR SECURITY. THE TISSUE CAME LOOSE AFTER SEVERAL ATTEMPTS TO PULL ON IT. PATIENT STATUS: NO PATIENT HARM HAS BEEN REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16387990 ↗
  • MalfunctionDec 3, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    NAME OF PROCEDURE BEING PERFORMED: UNSURE. DETAILED DESCRIPTION OF EVENT: AS PER ATTACHED FORM FILLED OUT BY NURSE ATTENDING, PART OF THE DEVICE CAME OFF IN PATIENT. QUALITY REPORT FORM FROM HOSPITAL STATES THAT SMALL PIECE DETACHED FROM GRASPER TIP. PATIENT STATUS: NO ADVERSE EVENT FOR PATIENT. TYPE OF INTERVENTION: SURGEON HAD TO LOCATE THE LOST PIECE FROM ABDOMEN - NO COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12929608 ↗
  • MalfunctionNov 7, 2021Flaked; Flaked

    NAME OF PROCEDURE BEING PERFORMED: LAP CHOLE DETAILED DESCRIPTION OF EVENT: I WAS ASKED BY A SURGEON TO CONTACT YOU ABOUT ONE OF YOUR PRODUCTS. WE WERE USING THE "EPIX LAPAROSCOPIC GRASPER" REF# (B)(4) TODAY FOR A LAP CHOLE. THE SURGEON STATED THAT THE GREEN SHAFT OF THE INSTRUMENT WAS FLAKING MATERIAL OFF INTO THE PATIENT. SMALL HAIR LIKE GREEN FIBERS OR PAINT OR SOMETHING WAS FLAKING OR COMING OFF OF THE SHAFT OF THE INSTRUMENT AND WAS ENDING UP INSIDE OF HER PATIENT." PRODUCT IS AVAILABLE FOR RETURN. ADDITIONAL INFORMATION RECEIVED VIA EMAIL ON 11OCT2021 FROM [NAME], [FACILITY] SERVICE COORDINATOR: THE STRANDS WERE REMOVED FROM THE PATIENT. THE FOLLOWING INFORMATION IS FROM THE SURGEON OF THE CASE. THE PATIENT WAS A THIN, YOUNG GIRL GOING IN FOR AN ELECTIVE CASE. [...] I HAVE USED THEIR INSTRUMENTS A LOT BEFORE AND THERE WAS NO TORQUE, EXCESS FORCE ETC WITH HER CASE. ADDITIONAL INFORMATION RECEIVED VIA EMAIL ON 11OCT2021 FROM [NAME], [FACILITY] SERVICE COORDINATOR: PHOTO OF THE DEVICE HAS BEEN PROVIDED. ADDITIONAL INFORMATION RECEIVED VIA EMAIL ON 22OCT2021 FROM (B)(6). (ENTERED BY (B)(4)) THERE IS NO PATIENT INJURY AT THIS TIME. ADDITIONAL INFORMATION RECEIVED VIA EMAIL ON 02NO

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12763640 ↗
  • MalfunctionAug 4, 2021Break; Break

    NAME OF PROCEDURE BEING PERFORMED: ROBOTIC LAPAROSCOPIC RECTOPEXY DETAILED DESCRIPTION OF EVENT: THE REP WAS NOT PRESENT DURING THE CASE. DURING THE CASE THE ASSISTANT WAS MOVING THE GRASPER UP THE SIGMOID COLON WHILE THE ROBOT ARM WAS MOVING DOWN THE SIGMOID COLON. DURING THE COURSE OF THIS MOVEMENT, THE GRASPER JAWS AND A PORTION OF THE GREEN SHAFT BROKE AND FELL INTO THE PATIENT. THE ENTIRE DEVICE, INCLUDING THE PORTION THAT BROKE OFF, WAS REMOVED FROM THE PATIENT AND THE CASE WAS COMPLETED WITH THE OTHER GRASPER ALREADY IN USE DURING THE CASE. NO HARM TO PATIENT. PRODUCT IS AVAILABLE FOR RETURN. ADDITIONAL INFORMATION RECEIVED VIA EMAIL ON 13JUL2021 FROM [NAME]: CONTACT INFORMATION FOR FACILITY CONTACT HAS BEEN PROVIDED. PHOTOS OF THE DEVICE HAVE BEEN PROVIDED. PATIENT STATUS: NO HARM TO PATIENT. TYPE OF INTERVENTION: USED OTHER GRASPER IN THE CASE TO COMPLETE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12278392 ↗
  • MalfunctionOct 22, 2020Detachment of Device or Device Component; Detachment of Device or Device Component

    TYPE OF PROCEDURE PERFORMED: VNOTES HYSTERECTOMY+BILATERAL SALPINGECTOMY. [NAME] WAS DOING A VNOTES CASE, AND TOWARDS THE END OF THE CASE ON THE SCREEN WE NOTICED THE GRASPER WAS SCISSORING. THE GRASPER WAS REMOVED AND SHE CONTINUED THE CASE WITH A REUSABLE GRASPER FROM THE TRAY. THE PATIENT WAS NO HARMED. [NAME] WAS NOT CONCERNED DURING OR AFTER THE CASE. IT WAS THE FIRST TIME SHE HAD SEEN THIS WITH THE APPLIED MEDICAL GRASPER. PATIENT STATUS: PATIENT IS FINE AND COMPLETELY UNAFFECTED. TYPE OF INTERVENTION: PRODUCT WAS REPLACED WITH ANOTHER.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10722955 ↗
  • MalfunctionOct 20, 2020Mechanical Jam; Mechanical Jam

    PROCEDURE PERFORMED: LAP CHOLE. CER 1 OF 2: (B)(4). CER 2 OF 2: (B)(4). A SECOND EPIX GRASPER WITH SAME LOTNUMBER WAS USED AFTER ISSUES WITH THE FIRST ONE THAT WAS USED (SEE PRECIOUS CER OF SAME DATE). EPIX GRASPER JAWS WERE LOCKED ON THE GALLBLADDER, EXPERIENCED DIFFICULTY TO UNLOCK THE JAWS. AFTER SEVERAL GRIP ACTIONS THEY SUCCEEDED TO UNLOCK THE JAWS. INTERVENTION: UNKNOWN. PATIENT STATUS: NO PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10708579 ↗
  • MalfunctionOct 15, 2020Mechanical Jam; Mechanical Jam

    PROCEDURE PERFORMED: LAP CHOLE. CER 1 OF 2: (B)(4). CER 2 OF 2: (B)(4). GRASPERS JAWS WERE LOCKED ON GALLBLADDER BUT THE GRASPER WOULD NOT UNLOCK. JAWS WERE UNLOCKED AFTER SEVERAL GRIP ACTIONS. GRASPER WAS DISCARDED & NEW EPIX GRASPER WAS USED WITH WHICH THEY HAD A VERY SIMILAR EXPERIENCE (SEPERATE CER). THIS EPIX GRASPER HAD THE SAME LOTNUMBER. INTERVENTION: CHANGE OF DEVICE. PATIENT STATUS: NO PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10687290 ↗
  • InjuryMar 6, 2020Patient Device Interaction Problem; Patient Device Interaction Problem

    PROCEDURE PERFORMED: LAP SUBTOTAL HYSTERECTOMY AND BSO. PATIENT DETAILS: FEMALE; DOB (B)(6). DR [NAME] WAS ASSISTING DR [NAME] CASE IN A LAP SUBTOTAL HYSTERECTOMY AND BSO; WHILST USING THE GRASPER A SMALL AMOUNT OF SMALL BOWEL WAS CAUGHT IN-BETWEEN THE HINGE OF THE EPIX GRASPER TIPS WHEN THE JAWS WERE OPEN AND SURGEON FOUND IT DIFFICULT TO DISLODGE ENTRAPPED TISSUE AND SOME BLEEDING OCCURRED SUGGESTING SMALL INJURY OCCURRED TO THE SMALL BOWEL. ADDITIONAL INFORMATION WAS RECEIVED VIA EMAIL FROM ADVANCED PROCEDURE SPECIALIST, [NAME] ON 14FEB2020: I HAVE SPOKEN TO [NAME] ABOUT THIS AND UNFORTUNATELY CANNOT CONFIRM THE BELOW. INTERVENTION: NI. PATIENT STATUS: STABLE.

    Patient problems: Bowel Perforation

    MAUDE report 9803203 ↗
  • MalfunctionOct 18, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    NAME OF PROCEDURE: LAP ABDOMINAL PERINEAL RESECTION. APPLIED MEDICAL TM WAS NOT PRESENT BUT THE FOLLOWING DESCRIPTION WAS GIVEN BY THE SCRUB NURSE [NAME]: "THE PADDED GRASPER HAD BEEN IN USE FOR APPROXIMATELY 4-5 HOURS. AFTER COMPLETING THE 'BOTTOM' END OF THE CASE, THE 'TOP' END WAS RESUMED. I HANDED OVER THE GRASPER AND AFTER A FEW SECONDS (30) I WAS INFORMED THAT THE GRASPER 'WASN'T WORKING'. THE SURGEON CONTINUED TO ATTEMPT TO USE THE GRASPER FOR A FEW MORE MINUTES UNTIL I OFFERED AN ALTERNATIVE. WHEN THE GRASPER HAD BEEN HANDED BACK OVER, I HAD A LOOK TO TRY AND SEE WHAT THE PROBLEM COULD BE, THAT'S WHEN I NOTICED, AT THE TIP, THE OUTSIDE LAYER HAD SPLIT AWAY. I CHECKED TO SEE THAT THE SPLIT MATCHED UP AND THERE WERE NO MISSING SEGMENTS THAT MAY HAVE BEEN LEFT INSIDE THE PATIENT - THERE WAS NONE AS IT MATCHED UP PERFECTLY. I INFORMED THE SURGEON, THE FLOOR NURSE AND ASKED FOR THE CHARGE NURSE (CN [NAME]) TO COME IN SO I COULD INFORM HER AND SHOW HER WHAT HAD HAPPENED. THE GRASPER WAS HANDED OFF MY TABLE." INTERVENTION: NI. PATIENT STATUS: PATIENT NOT AFFECTED BY INCIDENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9211503 ↗

FDA clearance and classification

The FDA regulates product code NWV (laparoscopic accessories, gynecologic) under 21 CFR 884.1720 as a Class I device. 21 CFR 884.1720 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

FDA definition of this device type: Exempt laparoscope accessories are non-powered, non-inflatable, simple manual mechanical devices, e.g., cannula, graspers, forceps, dissectors, scissors, etc. A gynecologic laparoscope is a device used to permit direct viewing of the organs within the peritoneum by a telescopic system introduced through the abdominal wall. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include: Trocar and cannula, instruments used through an operating channel, scope preheater, light source and cables, and component parts.

Common questions about the Epix Laparoscopic Grasper C4130

Has the Epix Laparoscopic Grasper C4130 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00607915110147), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Epix Laparoscopic Grasper C4130?

The FDA's MAUDE database lists 34 adverse event reports for this device, received between Feb 24, 2016 and May 1, 2026. In those reports, the most common specific device problems are “Material Frayed” (7), “Detachment of Device or Device Component” (5) and “Mechanical Jam” (5); the most common patient problems are “Physical Entrapment” (2), “Bowel Perforation” (1) and “Hemorrhage/Blood Loss/Bleeding” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Epix Laparoscopic Grasper C4130?

None of the reports are classified as deaths, 2 as injuries and 32 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Epix Laparoscopic Grasper C4130 FDA approved?

Devices of this type (product code NWV) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Applied Medical Resources Corporation.

Who makes the Epix Laparoscopic Grasper C4130, and is it still sold?

The FDA's device database lists Applied Medical Resources Corporation as the company responsible for it (brand Epix Laparoscopic Grasper, model C4130). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Laparoscopic Accessories, Gynecologic devices

All Laparoscopic Accessories, Gynecologic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page