Microcuff* Endotracheal Tube for Adults 35216Recalls & reported problems
Avanos Medical, Inc. · MICROCUFF* Adult Endotracheal Tube · Class 2
All models of Microcuff* Endotracheal Tube for Adults (8)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Microcuff endotracheal tube has 39 adverse event reports naming its identifier, including 4 injuries and no deaths. Reports mostly describe cuff air leaks and balloons that cannot hold air, sometimes causing device movement or disconnection. No recalls name this device. Reports fell from 9 in the previous 12 months to 2 in the most recent 12.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 39
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 39 adverse event reports that list this exact device (DI 00609038352169), received between May 28, 2020 and Jun 10, 2026 — classified as 4 injuries, 35 malfunctions.
The most frequently reported specific device problem is “Gas/Air Leak” (15 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Gas/Air Leak15
- Pressure Problem9
- Migration or Expulsion of Device7
- Burst Container or Vessel5
- Disconnection5
- Defective Device2
- Therapeutic or Diagnostic Output Failure2
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions18
- No Known Impact Or Consequence To Patient9
- Insufficient Information6
- Cardiac Arrest2
- Low Oxygen Saturation2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Bradycardia1
- Low Blood Pressure/ Hypotension1
Recent adverse event reports
The 25 most recent of 39 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 10, 2026Defective Device; Gas/Air Leak; Therapeutic or Diagnostic Output Failure
IT WAS REPORTED: THAT DURING BEDSIDE INTUBATION, THREE DIFFERENT ET (ENDOTRACHEAL TUBE) TUBES WERE FOUND TO BE DEFECTIVE. THE BALLOON ON THE END OF THE TUBE WAS UNABLE TO HOLD AIR ON ALL THREE TUBES. ON THE FOURTH TRY, THE FOURTH ET TUBE WAS SUCCESSFUL.PT: 4582. DEVICE: 2588, 2946, 3023. REFERENCE: MW5187406, MW5187408.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25465935 ↗MalfunctionJun 10, 2026Defective Device; Gas/Air Leak; Therapeutic or Diagnostic Output Failure
IT WAS REPORTED: THAT DURING BEDSIDE INTUBATION, THREE DIFFERENT ET (ENDOTRACHEAL TUBE) TUBES WERE FOUND TO BE DEFECTIVE. THE BALLOON ON THE END OF THE TUBE WAS UNABLE TO HOLD AIR ON ALL THREE TUBES. ON THE FOURTH TRY, THE FOURTH ET TUBE WAS SUCCESSFUL. PT: 4582. DEVICE: 2588, 2946, 3023. REFERENCE: MW5187406, MW5187407.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25465930 ↗MalfunctionOct 29, 2025Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED: THE CUFF "BLEW" POST INTUBATION ON MULTIPLE (FIVE (5)) PATIENTS, ACROSS THREE (3) UNITS AND IN THE EMERGENCY ROOM (ER) IN LESS THAN A TWO-MONTH PERIOD. ALL THE PATIENTS WERE RE-INTUBATED WITHOUT INCIDENT. 5 DIFFERENT REPORTS WILL BE SUBMITTED FOR EACH PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23411319 ↗MalfunctionOct 29, 2025Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED: THE CUFF "BLEW" POST INTUBATION ON MULTIPLE (FIVE (5) PATIENTS, ACROSS THREE (3) UNITS AND IN THE EMERGENCY ROOM (ER) IN LESS THAN A TWO-MONTH PERIOD. ALL THE PATIENTS WERE RE-INTUBATED WITHOUT INCIDENT. 5 DIFFERENT REPORTS WILL BE SUBMITTED FOR EACH PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23411316 ↗MalfunctionOct 29, 2025Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED: THE CUFF "BLEW" POST INTUBATION ON MULTIPLE (FIVE (5) PATIENTS, ACROSS THREE (3) UNITS AND IN THE EMERGENCY ROOM (ER) IN LESS THAN A TWO-MONTH PERIOD. ALL THE PATIENTS WERE RE-INTUBATED WITHOUT INCIDENT. 5 DIFFERENT REPORTS WILL BE SUBMITTED FOR EACH PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23410667 ↗MalfunctionOct 29, 2025Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED: THE CUFF "BLEW" POST INTUBATION ON MULTIPLE (FIVE (5)) PATIENTS, ACROSS THREE (3) UNITS AND IN THE EMERGENCY ROOM (ER) IN LESS THAN A TWO-MONTH PERIOD. ALL THE PATIENTS WERE RE-INTUBATED WITHOUT INCIDENT. 5 DIFFERENT REPORTS WILL BE SUBMITTED FOR EACH PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23410654 ↗MalfunctionOct 29, 2025Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED: THE CUFF "BLEW" POST INTUBATION ON MULTIPLE (FIVE (5) PATIENTS, ACROSS THREE (3) UNITS AND IN THE EMERGENCY ROOM (ER) IN LESS THAN A TWO-MONTH PERIOD. ALL THE PATIENTS WERE RE-INTUBATED WITHOUT INCIDENT. 5 DIFFERENT REPORTS WILL BE SUBMITTED FOR EACH PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23410644 ↗MalfunctionAug 29, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED LEAK IN THE CUFF, DISCOVERED DURING PATIENT USE (MULTIPLE OCCURRENCES). THE ETT TUBE WAS REPLACED. THERE WAS NO REPORTED INJURY. REPORTABLE-WHILE NO HARM OR INJURY WAS REPORTED, ANY COMPONENT THAT IS PART OF THE VENT CIRCUIT (FLEX CONNECTOR, DOUBLE SWIVEL ELBOW, Y-ADAPTER, T-PIECE, ETT ADAPTER, ETC.) BREAKS, CRACKS, OR LEAKS AFTER THE VENT CIRCUIT HAS BEEN ESTABLISHED ON THE PATIENT AND WHICH REQUIRES A REPLACEMENT ETT COULD CAUSE OR LEAD TO A SERIOUS INJURY AND/OR HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22929346 ↗MalfunctionDec 3, 2024Break; Material Twisted/Bent; Break; Material Twisted/Bent
IT WAS REPORTED, THERE WAS NO SUCTION (DURING USE), IT WAS DISCOVERED KINKING HAD OCCURRED IN [THE] ORAL PHARYNX; THE PATIENT WAS REINTUBATED. IT WAS ADDITIONALLY REPORTED, A ¿NON-SERIOUS¿ OTHER EVENT(S) OCCURRED; THERE WAS NO REPORTED INJURY. ADDITIONAL INFORMATION RECEIVED 07NOV2024 REPORTED, IT IS UNKNOWN HOW LONG THE ENDOTRACHEAL TUBE (ETT) WAS IN PLACE PRIOR TO THE TUBE KINKING.
Patient problems: Insufficient Information
MAUDE report 20835742 ↗InjuryNov 12, 2024Gas/Air Leak; Gas/Air Leak
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME PATIENT. THIS IS THE SECOND OF TWO REPORTS. REFER TO 3011270181-2024-00117 FOR THE FIRST REPORT FDA USER FACILITY MW REPORT MW5159593 REPORTED, ¿[THE] PATIENT REQUIRED EMERGENT INTUBATION... [THE] TUBE WAS REPLACED DUE TO CUFF LEAKING AIR. PATIENT TO OPERATING ROOM FOR TRACHEOSTOMY¿¿ NO ADDITIONAL INFORMATION WAS PROVIDED CONCERNING THE PATIENT¿S CARE OR CURRENT CONDITION.
Patient problems: Insufficient Information
MAUDE report 20667749 ↗InjuryNov 12, 2024Gas/Air Leak; Gas/Air Leak
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME PATIENT. THIS IS THE FIRST OF TWO REPORTS. REFER TO 3011270181-2024-00118 FOR THE SECOND REPORT. FDA USER FACILITY MW REPORT MW5159592 REPORTED, ¿[THE] PATIENT REQUIRED EMERGENT INTUBATION... [THE] TUBE WAS REPLACED DUE TO CUFF LEAKING AIR. PATIENT TO OPERATING ROOM FOR TRACHEOSTOMY¿¿ NO ADDITIONAL INFORMATION WAS PROVIDED CONCERNING THE PATIENT¿S CARE OR CURRENT CONDITION.
Patient problems: Insufficient Information
MAUDE report 20666184 ↗InjuryJan 17, 2024Disconnection; Disconnection
FDA USER FACILITY MED WATCH REPORT FDA MW (B)(4) REPORTED, "THE 15MM CONNECTOR, WHERE THE VENTILATOR OR O2 [OXYGEN] SOURCE CONNECTS, DISCONNECTS FROM THE NECK OF THE ETT [ENDOTRACHEAL TUBE]. THE 15MM CONNECTOR APPEARS TO BECOME LOOSE FROM THE NECK OF THE ETT WHEN WARMED IN THE AIRWAY. WHEN THIS OCCURS, THE PATIENT IS NOT EFFECTIVELY BEING VENTILATED. AT LEAST ONE PATIENT HAS HAD TO BE REINTUBATED IN ORDER TO CONTINUE VENTILATOR TREATMENT. OVERNIGHT ON [REDACTED DATE], THERE WAS A PATIENT IN THE MEDICAL ICU [INTENSIVE CARE UNIT] WHO CODED POST INTUBATION. WHILE TAKING THE COLORMETRIC OFF AND PLACING THE EMMA [EMERGENCY CAPNOMETER], THE BLUE PIECE JUST POPPED OFF. THE RESPIRATORY THERAPIST WAS ABLE TO GET THE CONNECTOR BACK ON QUICKLY." IT WAS ADDITIONALLY REPORTED, THE EVENT OCCURRED IN THE PATIENT¿S HOSPITAL ROOM AND THE DEVICE HAD BEEN IN PLACE FOR APPROXIMATELY 1 DAY; THE PATIENT'S CURRENT CONDITION WAS NOT REPORTED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 18530546 ↗MalfunctionDec 28, 2023
RESPIRATORY THERAPY REPORTS PROBLEMS WITH THE AVANOS MICROCUFF ENDOTRACHEAL TUBES, IN ADULT SIZES 6.0MM, 7.5MM, AND 8.0MM. THE 15MM CONNECTOR, WHERE THE VENTILATOR OR O2 SOURCE CONNECTS, DISCONNECTS FROM THE NECK OF THE ETT. THE 15MM CONNECTOR APPEARS TO BECOME LOOSE FROM THE NECK OF THE ETT WHEN WARMED IN THE AIRWAY. WHEN THIS OCCURS, THE PATIENT IS NOT EFFECTIVELY BEING VENTILATED. AT LEAST ONE PATIENT HAS HAD TO BE RE-INTUBATED IN ORDER TO CONTINUE VENTILATOR TREATMENT. OVERNIGHT ON [REDACTED DATE], THERE WAS A PATIENT IN THE MEDICAL ICU WHO CODED POST INTUBATION. WHILE TAKING THE COLORMETRIC OFF AND PLACING THE EMMA, THE BLUE PIECE JUST POPPED OFF. THE RESPIRATORY THERAPIST WAS ABLE TO GET THE CONNECTOR BACK ON QUICKLY.
Patient problems: Cardiac Arrest
MAUDE report 18408532 ↗MalfunctionMay 3, 2023Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH THE SAME UNIT, INVOLVING THE SAME PATIENT. THIS IS THE FIRST OF THREE REPORTS. REFER TO 3011270181-2023-00038 FOR THE SECOND REPORT. REFER TO 3011270181-2023-00039 FOR THE THIRD REPORT. FDA MEDWATCH / FDA USER FACILITY REPORT MW REPORT 4100070000-2023-8013 REPORTED, ¿PATIENT INTUBATED. MULTIPLE TIMES DURING THE DAY, THE PATIENT BECAME DISCONNECTED FROM THE VENT. THE VENTILATOR TUBING WOULD SPONTANEOUSLY COME APART FROM THE ENDOTRACHEAL TUBE. UPON FURTHER INVESTIGATION WITH RESPIRATORY THERAPY, IT WAS FOUND THAT THE END OF THE 8.0 TUBE WAS STRETCHED OUT FROM HEAT.¿ THERE WAS NO REPORTED INJURY.
Patient problems: Insufficient Information
MAUDE report 16860698 ↗MalfunctionMay 3, 2023Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH THE SAME UNIT, INVOLVING THE SAME PATIENT. THIS IS THE SECOND OF THREE REPORTS. REFER TO 3011270181-2023-00037 FOR THE FIRST REPORT. REFER TO 3011270181-2023-00039 FOR THE THIRD REPORT. FDA MEDWATCH / FDA USER FACILITY REPORT MW REPORT 4100070000-2023-8013 REPORTED, ¿PATIENT INTUBATED. MULTIPLE TIMES DURING THE DAY, THE PATIENT BECAME DISCONNECTED FROM THE VENT. THE VENTILATOR TUBING WOULD SPONTANEOUSLY COME APART FROM THE ENDOTRACHEAL TUBE. UPON FURTHER INVESTIGATION WITH RESPIRATORY THERAPY, IT WAS FOUND THAT THE END OF THE 8.0 TUBE WAS STRETCHED OUT FROM HEAT.¿ THERE WAS NO REPORTED INJURY
Patient problems: Insufficient Information
MAUDE report 16860697 ↗MalfunctionMay 2, 2023Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH THE SAME UNIT, INVOLVING THE SAME PATIENT. THIS IS THE THIRD OF THREE REPORTS. REFER TO 3011270181-2023-00037 FOR THE FIRST REPORT REFER TO 3011270181-2023-00038 FOR THE SECOND REPORT FDA MEDWATCH / FDA USER FACILITY REPORT MW REPORT 4100070000-2023-8013 REPORTED, ¿PATIENT INTUBATED. MULTIPLE TIMES DURING THE DAY, THE PATIENT BECAME DISCONNECTED FROM THE VENT. THE VENTILATOR TUBING WOULD SPONTANEOUSLY COME APART FROM THE ENDOTRACHEAL TUBE. UPON FURTHER INVESTIGATION WITH RESPIRATORY THERAPY, IT WAS FOUND THAT THE END OF THE 8.0 TUBE WAS STRETCHED OUT FROM HEAT.¿ THERE WAS NO REPORTED INJURY
Patient problems: Insufficient Information
MAUDE report 16857836 ↗MalfunctionApr 26, 2023Separation Problem; Separation Problem
IT WAS REPORTED, "WE TURNED THE PATIENT ON HIS SIDE FOR BODY CARE. THE TUBE TWISTED AND SNAPPED OFF. ACTIVE HUMIDIFICATION WAS CONNECTED" [AT THE TIME THE INCIDENT OCCURRED]. THE PRODUCT WAS IN USE FOR LESS THAN FIVE DAYS. NO INJURY OR MEDICAL INTERVENTIONS REPORTED. PER ADDITIONAL INFORMATION RECEIVED ON 5APR2023, REINTUBATION WAS NOT REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16820263 ↗MalfunctionApr 6, 2023Disconnection; Stretched
PATIENT INTUBATED. MULTIPLE TIMES DURING THE DAY, THE PATIENT BECAME DISCONNECTED FROM THE VENT. THE VENTILATOR TUBING WOULD SPONTANEOUSLY COME APART FROM THE ENDOTRACHEAL TUBE. UPON FURTHER INVESTIGATION WITH RESPIRATORY THERAPY, IT WAS FOUND THAT THE END OF THE 8.0 TUBE WAS STRETCHED OUT FROM HEAT.
Patient problems: Ventilator Dependent
MAUDE report 16688932 ↗MalfunctionJul 26, 2022Material Integrity Problem; Material Integrity Problem
THE RESPIRATORY THERAPIST (RT) REPORTED, THE ENDOTRACHEAL TUBE (ETT) CUFF WAS NOTED TO BE ¿BLOWN¿, THE CUFF WAS NOT ABLE TO BE FILLED; THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15105272 ↗MalfunctionMay 13, 2022Migration or Expulsion of Device; Migration or Expulsion of Device
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED SIX DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS. THIS IS THE SECOND OF SIX REPORTS. REFER TO 3011270181-2022-00028 FOR THE FIRST REPORT. REFER TO 3011270181-2022-00030 FOR THE THIRD REPORT. REFER TO 3011270181-2022-00031 FOR THE FOURTH REPORT. REFER TO 3011270181-2022-00032 FOR THE FIFTH REPORT. REFER TO 3011270181-2022-00033 FOR THE SIXTH REPORT. IT WAS REPORTED THAT THE TUBE POPPED OFF AND THEY HAD TO TAPE THE CONNECTOR TO THE TUBE. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14401145 ↗MalfunctionMay 7, 2022Migration or Expulsion of Device; Migration or Expulsion of Device
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED SIX DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS. THIS IS THE SIXTH OF SIX REPORTS. REFER TO 3011270181-2022-00028 FOR THE FIRST REPORT. REFER TO 3011270181-2022-00029 FOR THE SECOND REPORT. REFER TO 3011270181-2022-00030 FOR THE THIRD REPORT. REFER TO 3011270181-2022-00031 FOR THE FOURTH REPORT. REFER TO 3011270181-2022-00032 FOR THE FIFTH REPORT. IT WAS REPORTED THAT THE TUBE POPPED OFF; THE THERAPIST STATED THAT THEY HAD TO REPLACE TUBE. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14331967 ↗MalfunctionMay 5, 2022Migration or Expulsion of Device; Migration or Expulsion of Device
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED SIX DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS. THIS IS THE FOURTH OF SIX REPORTS. REFER TO 3011270181-2022-00028 FOR THE FIRST REPORT. REFER TO 3011270181-2022-00029 FOR THE SECOND REPORT . REFER TO 3011270181-2022-00030 FOR THE THIRD REPORT . REFER TO 3011270181-2022-00032 FOR THE FIFTH REPORT. REFER TO 3011270181-2022-00033 FOR THE SIXTH REPORT. IT WAS REPORTED THAT THE TUBE POPPED OFF; THE THERAPIST STATED THAT THEY HAD TO REPLACE TUBE. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14291934 ↗MalfunctionMay 5, 2022Migration or Expulsion of Device; Migration or Expulsion of Device
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED SIX DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS. THIS IS THE FIFTH OF SIX REPORTS. REFER TO 3011270181-2022-00028 FOR THE FIRST REPORT. REFER TO 3011270181-2022-00029 FOR THE SECOND REPORT. REFER TO 3011270181-2022-00030 FOR THE THIRD REPORT. REFER TO 3011270181-2022-00031 FOR THE FOURTH REPORT. REFER TO 3011270181-2022-00033 FOR THE SIXTH REPORT. IT WAS REPORTED THAT THE TUBE POPPED OFF; THE THERAPIST STATED THAT THEY HAD TO REPLACE TUBE. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14291881 ↗MalfunctionMay 5, 2022Migration or Expulsion of Device; Migration or Expulsion of Device
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED SIX DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS. THIS IS THE SECOND OF SIX REPORTS. REFER TO 3011270181-2022-00028 FOR THE FIRST REPORT. REFER TO 3011270181-2022-00030 FOR THE THIRD REPORT . REFER TO 3011270181-2022-00031 FOR THE FOURTH REPORT. REFER TO 3011270181-2022-00032 FOR THE FIFTH REPORT. REFER TO 3011270181-2022-00033 FOR THE SIXTH REPORT. IT WAS REPORTED THAT THE TUBE POPPED OFF AND THEY HAD TO TAPE THE CONNECTOR TO THE TUBE. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14291827 ↗MalfunctionMay 5, 2022Migration or Expulsion of Device; Migration or Expulsion of Device
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED SIX DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS. THIS IS THE FIRST OF SIX REPORTS. REFER TO 3011270181-2022-00029 FOR THE SECOND REPORT. REFER TO 3011270181-2022-00030 FOR THE THIRD REPORT . REFER TO 3011270181-2022-00031 FOR THE FOURTH REPORT . REFER TO 3011270181-2022-00032 FOR THE FIFTH REPORT. REFER TO 3011270181-2022-00033 FOR THE SIXTH REPORT. IT WAS REPORTED THAT THE CODE PATIENT WAS INTUBATED WITH THE TUBE IN THE EMERGENCY ROOM (ER) AND THE TOP POPPED OFF AS SOON AS THE TUBE WAS PLACED; THE PATIENT WAS BAGGED AND REINTUBATED WITH AN 8.5 ENDOTRACHEAL (ETT). THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14291705 ↗
Other Tube, tracheal (w/wo connector) devices
- Shiley 18875Covidien LP · Tube, tracheal (w/wo connector)4 deaths570 reports
- RUSCH IPN041476Teleflex Incorporated · Tube, tracheal (w/wo connector)10 recalls0 reports
- Shiley 18775Covidien LP · Tube, tracheal (w/wo connector)489 reports
- Shiley 18770Covidien LP · Tube, tracheal (w/wo connector)2 deaths434 reports
- Shiley 18880Covidien LP · Tube, tracheal (w/wo connector)2 deaths370 reports
- Shiley 18780Covidien LP · Tube, tracheal (w/wo connector)3 deaths286 reports
- Shiley 18770SCovidien LP · Tube, tracheal (w/wo connector)273 reports
- Shiley 18780SCovidien LP · Tube, tracheal (w/wo connector)251 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page