Skip to content

VaPro 72124Recalls & reported problems

Hollister Incorporated · No Touch Urethral Intermittent Catheter, Nelaton · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00610075092635), received between Nov 22, 2024 and Nov 22, 2024 — classified as 1 injury.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available1

Patient problems

  • Hematuria1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 22, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE END USER HAD BEEN CATHETERIZING FOR AT LEAST THE LAST 18 MONTHS AND HAD RECENTLY NOTICED A DIFFERENCE WITH THE LUBRICANT AND SLEEVE OF THE VAPRO CATHETER. IT WAS FURTHER REPORTED THAT THE END USER EXPERIENCED BLEEDING WHEN WITHDRAWING THE CATHETERS FROM HIS URETHRA. IT WAS REPORTED THAT HE WENT TO THE ED AND THEY USED A CLAMP TO CREATE A CLOT. IT WAS ALSO REPORTED THAT HE WAS REFERRED TO A UROLOGIST WHO HAD THE END USER UNDERGO A CT SCAN TO INVESTIGATE HIS REPORT, AND THE CT SCAN SHOWED THE URINE WAS CLEAR AND NO SIGNS OF CANCER. THE UROLOGIST SUGGESTED THE END USER TRY A TIEMAN TIP OR OTHER TYPES OF CATHETERS.

    Patient problems: Hematuria

    MAUDE report 20758781 ↗

FDA clearance and classification

Cleared through 510(k) K233524 on May 3, 2024 (decision: Substantially Equivalent). Submitted by Hollister Incorporated as “Sleeved IC 2 Family; Sleeved IC 2 SWT (Name not Finalized); Sleeved IC 2 Plus (Name not Finalized); Sleeved IC 2 Pocket (Name not Finalized); Sleeved IC 2 Plus Pocket (Name not Finalized)”. The FDA review took 184 days. FDA record ↗

The FDA regulates product code GBM (catheter, urethral) under 21 CFR 876.5130 as a Class II device. 21 CFR 876.5130 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the VaPro 72124

Has the VaPro 72124 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00610075092635), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the VaPro 72124?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Nov 22, 2024. That report lists patient problem “Hematuria” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the VaPro 72124?

None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the VaPro 72124 FDA approved?

It was cleared through 510(k) K233524 on May 3, 2024. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the VaPro 72124, and is it still sold?

The FDA's device database lists Hollister Incorporated as the company responsible for it (brand VaPro, model 72124). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CATHETER, URETHRAL devices

All CATHETER, URETHRAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page